EUVIMEDEUVIMED entdecken
Crossref

Simultaneous quantification of endogenous biomarkers and detection of testosterone esters in human serum using gas chromatography–high‐resolution mass spectrometry

Dayamín Martínez Brito iD, Patrizia Leogrande iD, Daniel Jardines iD, Xavier de la Torre iD, Francesco Botrè iD

Rapid Communications in Mass Spectrometry · 2024

DOI: 10.1002/rcm.9685 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, 2024. Verfasst von Dayamín Martínez Brito und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratorio Antidoping FMSI, Federazione Medico Sportiva Italiana, Rome, Italy. Hauptthemen laut MEDLINE: Testosterone, Esters. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: World Anti-Doping Agency.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

High‐resolution mass spectrometry (HRMS) has been demonstrated to be an alternative platform for quantitative analyses, identifying unknown compounds and gathering information for the elucidation of chemical structures. This work describes a method to detect 13 esters of testosterone (T) and 5 biomarkers in 0.1 mL of human serum using gas chromatography (GC) coupled to HRMS.

Methods

Analytes were extracted from serum after deproteinization and liquid–liquid extraction. The trimethylsilyl derivatives were analyzed using a gas chromatograph coupled to HRMS at low electron energy to minimize molecule fragmentation. The acquisition in profiling full‐scan mode was applied with a resolving power of 30 000 at m/z 400. Linearity, lower limit of quantitation, and measurement uncertainty were assessed. Precision and accuracy were assessed at 0.5 and 2 ng/mL, respectively. Mass accuracy (MA) and mass extraction window (MEW) were also evaluated.

Results

T esters showed a linear response between 0.25 and 10 ng/mL (except for undecanoate, enanthate, and propionate that showed lineal responses between 0.5 and 10 ng/mL and isocaproate between 2 and 10 ng/mL); detection limits remained between 0.1 and 0.5 ng/mL and accuracy between 81% and 119%. The MA (MEW = 10 ppm) was maintained between −2.4 and 4.8 ppm. The biomarkers (T, androstenedione, dehydroepiandrosterone [DHEA], estradiol, and 17‐OH‐progesterone) showed a linear response within the evaluated range; quantification limits remained between 0.1 and 0.5 ng/mL (except for DHEA), the accuracy between 88% and 99%, and precision between 3.5% and 10.8%. Measurement uncertainties were found between 5.6% and 17.2%. MA (MEW = 3 ppm) was maintained between −0.47 and 0.12 ppm.

Conclusions

The method to detect T esters and five endogenous biomarkers in serum using GC coupled to HRMS showed linear responses up to 10 ng/mL with adequate precision, accuracy, and uncertainties. It was possible to distinguish cholesterol from T‐isocaproate based on the MEW of 10 ppm, preventing false positives. In addition, this method allows searching for other biomarkers and/or unknown metabolites and other ester forms not included here but at a later stage if necessary.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TestosteroneEstersHumansGas Chromatography-Mass SpectrometryMass SpectrometryDehydroepiandrosterone

Substanzen

TestosteroneEstersDehydroepiandrosterone

Autor:innen und Institutionen

  • Laboratorio Antidoping FMSI, Federazione Medico Sportiva Italiana, Rome, Italy.
  • REDs-Research and Expertise on Anti-Doping Sciences, Institute of Sport Science, University of Lausanne, Lausanne, Switzerland.

ORCID-Profile: Dayamín Martínez Brito, Patrizia Leogrande, Daniel Jardines, Xavier de la Torre, Francesco Botrè

Förderung

  • World Anti-Doping Agency

Zitieren

APA

Martínez Brito, D., Leogrande, P., Jardines, D., de la Torre, X., & Botrè, F. (2024). Simultaneous quantification of endogenous biomarkers and detection of testosterone esters in human serum using gas chromatography–high‐resolution mass spectrometry. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.9685

Vancouver

Martínez Brito D, Leogrande P, Jardines D, de la Torre X, Botrè F. Simultaneous quantification of endogenous biomarkers and detection of testosterone esters in human serum using gas chromatography–high‐resolution mass spectrometry. Rapid Communications in Mass Spectrometry. 2024. doi:10.1002/rcm.9685

BibTeX
@article{MartnezBrito2024,
  author = {Martínez Brito, Dayamín and Leogrande, Patrizia and Jardines, Daniel and de la Torre, Xavier and Botrè, Francesco},
  title = {Simultaneous quantification of endogenous biomarkers and detection of testosterone esters in human serum using gas chromatography–high‐resolution mass spectrometry},
  journal = {Rapid Communications in Mass Spectrometry},
  year = {2024},
  doi = {10.1002/rcm.9685}
}
Kostenlos starten →