Evaluating the optimal tissue thickness for mass spectrometry imaging using infrared matrix‐assisted laser desorption electrospray ionization

Alena N Joignant, Kevan T Knizner, Ying Xi, David C Muddiman

Rapid Communications in Mass Spectrometry · 2023

DOI: 10.1002/rcm.9638 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, 2023. Verfasst von Alena N Joignant und 3 weiteren Autor:innen. Beteiligte Einrichtung: FTMS Laboratory for Human Health Research, Department of Chemistry, North Carolina State University, Raleigh, North Carolina, USA. Hauptthemen laut MEDLINE: Spectrometry, Mass, Electrospray Ionization, Histological Techniques. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIGMS NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

Infrared matrix‐assisted laser desorption electrospray ionization (IR‐MALDESI) utilizes a 2970 nm mid‐IR laser to desorb samples with depth resolutions ( Z ) on the order of micrometers. Conventionally, 5–20 μm thick tissue sections are used to characterize different applications of the IR‐MALDESI source, but an optimal thickness has not been systematically investigated.

Methods

Mouse liver was sectioned to various thicknesses and analyzed using IR‐MALDESI mass spectrometry imaging (MSI). Height profiles of tissue sections of various cryosectioned thicknesses were acquired to affirm tissue thickness. Tissue sections of each thickness were measured using a Keyence microscope. Paraffin wax was cryosectioned, mounted on microscope slides, and measured using a chromatic confocal sensor system to determine the cryostat sectioning accuracy.

Results

Analyzing sectioned tissues at higher thickness (>10 μm) leads to lower ion abundance, a decrease in signal over long analysis times, and more frequent instrument cleaning. Additionally, increasing tissue thickness above the optimum (7 μm) does not result in a significant increase in lipid annotations.

Conclusions

This work defines an optimal sample thickness for IR‐MALDESI‐MSI and demonstrates the utility of optimizing tissue thickness for MSI platforms of comparable Z resolution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spectrometry, Mass, Electrospray IonizationHistological TechniquesMiceAnimalsSpectrometry, Mass, Matrix-Assisted Laser Desorption-IonizationLightLasers

Autor:innen und Institutionen

  • FTMS Laboratory for Human Health Research, Department of Chemistry, North Carolina State University, Raleigh, North Carolina, USA.

Förderung

  • NIGMS NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Joignant, A. N., Knizner, K. T., Xi, Y., & Muddiman, D. C. (2023). Evaluating the optimal tissue thickness for mass spectrometry imaging using infrared matrix‐assisted laser desorption electrospray ionization. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.9638

Vancouver

Joignant AN, Knizner KT, Xi Y, Muddiman DC. Evaluating the optimal tissue thickness for mass spectrometry imaging using infrared matrix‐assisted laser desorption electrospray ionization. Rapid Communications in Mass Spectrometry. 2023. doi:10.1002/rcm.9638

BibTeX
@article{Joignant2023,
  author = {Joignant, Alena N and Knizner, Kevan T and Xi, Ying and Muddiman, David C},
  title = {Evaluating the optimal tissue thickness for mass spectrometry imaging using infrared matrix‐assisted laser desorption electrospray ionization},
  journal = {Rapid Communications in Mass Spectrometry},
  year = {2023},
  doi = {10.1002/rcm.9638}
}
Kostenlos starten →