Evaluating the accuracy and reliability of compound‐specific carbon isotopic analysis using gas chromatography‐combustion‐isotope ratio mass spectrometry with the addition of a reduction furnace

Yunning Cao, Hu Liu, Jing Hu, Zheng Wang, Mengshu Zhu, Xu Liu, Kaili Yang, Haijiao Gan, Weiguo Liu iD

Rapid Communications in Mass Spectrometry · 2023

DOI: 10.1002/rcm.9450 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, 2023. Verfasst von Yunning Cao und 8 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Loess and Quaternary Geology, Institute of Earth Environment, Chinese Academy of Sciences, Xi'an, China. Hauptthemen laut MEDLINE: Caffeine, Nitrogen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: the Chinese Academy of Sciences, the Technology Development and Innovation Programme of Chinese Academy of Sciences, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

Gas chromatography‐combustion‐isotope ratio mass spectrometry (GC/C/IRMS) is widely used for compound‐specific carbon isotopic analysis. However, current isotopic analysis systems utilize the GC IsoLink combustion reactor, and independent reduction furnaces are not implemented. Therefore, whether this limitation in furnace use affects the precision of compound‐specific carbon isotopic analysis needs to be evaluated.

Methods

We attempted to add a separate reduction furnace to the GC IsoLink interface and compared the δ 13 C values of n ‐alkanes (including C and H elements), fatty acid methyl ester (including C, H, and O elements), caffeine (USGS61 and USGS62, including C, H, O, and N elements), and 9‐ethylcarbazole (including C, H, and N elements) before and after the addition of the reduction furnace using the GC IsoLink combustion reactor.

Results

For n ‐alkanes and fatty acid methyl esters, the δ 13 C differences between the measured values and their standard values were basically falling within 0.5‰ whether or not an independent reduction furnace was added. However, for the nitrogen‐containing compounds (caffeine and 9‐ethylcarbazole), the δ 13 C differences between the measured values and their standard values were much larger without an independent reduction furnace (1.0–3.71‰ for USGS61, 1.78–2.19‰ for USGS62, and 0.39–1.13‰ for 9‐ethylcarbazole) than with a reduction furnace (−0.31–0.68‰ for USGS61, −0.44–0.06‰ for USGS62, and −0.04–0.25‰ for 9‐ethylcarbazole).

Conclusions

The addition of an independent reduction furnace had no significant effect on the δ 13 C of n ‐alkanes and fatty acid methyl esters, but it had a significant effect on the δ 13 C of nitrogen‐containing compounds. It is suggested that GC IsoLink needs an independent reduction furnace to effectively eliminate the interference of NO x on CO 2 isotopic determination to improve the accuracy of δ 13 C for nitrogen‐containing compounds.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CaffeineNitrogenCarbon IsotopesReproducibility of ResultsGas Chromatography-Mass SpectrometryAlkanes

Substanzen

Carbon IsotopesCaffeineNitrogenAlkanes

Autor:innen und Institutionen

  • State Key Laboratory of Loess and Quaternary Geology, Institute of Earth Environment, Chinese Academy of Sciences, Xi'an, China.
  • CAS Center for Excellence in Quaternary Science and Global Change, Xi'an, China.
  • Xi'an Institute for Innovative Earth Environment Research, Xi'an, China.
  • University of Chinese Academy of Sciences, Beijing, China.

ORCID-Profile: Weiguo Liu

Förderung

  • the Chinese Academy of Sciences
  • the Technology Development and Innovation Programme of Chinese Academy of Sciences
  • National Natural Science Foundation of China
  • University of Chinese Academy of Sciences
  • Chinese Academy of Sciences

Zitieren

APA

Cao, Y., Liu, H., Hu, J., Wang, Z., Zhu, M., Liu, X., Yang, K., Gan, H., & Liu, W. (2023). Evaluating the accuracy and reliability of compound‐specific carbon isotopic analysis using gas chromatography‐combustion‐isotope ratio mass spectrometry with the addition of a reduction furnace. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.9450

Vancouver

Cao Y, Liu H, Hu J, Wang Z, Zhu M, Liu X, et al. Evaluating the accuracy and reliability of compound‐specific carbon isotopic analysis using gas chromatography‐combustion‐isotope ratio mass spectrometry with the addition of a reduction furnace. Rapid Communications in Mass Spectrometry. 2023. doi:10.1002/rcm.9450

BibTeX
@article{Cao2023,
  author = {Cao, Yunning and Liu, Hu and Hu, Jing and Wang, Zheng and Zhu, Mengshu and Liu, Xu and Yang, Kaili and Gan, Haijiao and Liu, Weiguo},
  title = {Evaluating the accuracy and reliability of compound‐specific carbon isotopic analysis using gas chromatography‐combustion‐isotope ratio mass spectrometry with the addition of a reduction furnace},
  journal = {Rapid Communications in Mass Spectrometry},
  year = {2023},
  doi = {10.1002/rcm.9450}
}
Kostenlos starten →