Numerical Simulation and Experimental Study of a Dual‐Ionization Source (PI/EI) Quadrupole Mass Spectrometer Based on the CFD–DSMC Method
Rapid Communications in Mass Spectrometry · 30.06.2026 · 40 (12)
DOI: 10.1002/rcm.70069 ↗Kurz gefasst
Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, online erschienen am 30.03.2026. Verfasst von Jun Huang und 7 weiteren Autor:innen. Beteiligte Einrichtung: National Synchrotron Radiation Laboratory, University of Science and Technology of China, Hefei, China. Die Arbeit nennt 27 Literaturangaben. Gefördert von: National Natural Science Foundation of China, National Key Research and Development Program of China, Science and Technology Plan Project of China Tobacco Anhui Industrial Co. Ltd..
Abstract (Originalsprache)
Background
Online analysis of complex gas mixtures is often hindered by spectral congestion resulting from extensive fragmentation in traditional electron ionization (EI) sources. Although photoionization (PI) provides clean mass spectra by preserving molecular ion integrity, its comprehensive detection is limited by an inherent ionization selectivity toward species with low ionization energies.
Methods
To address these challenges, a dual‐ionization source quadrupole mass spectrometer (DISQMS) integrating PI and EI was developed. Its design and optimization were guided by a multiphysics ion trajectory simulation method coupling continuum‐rarefied flow fields with electric fields. The flow field, spanning from the sampling capillary to the quadrupole mass analyzer chamber, was numerically simulated using a hybrid computational fluid dynamics–direct simulation Monte Carlo (CFD–DSMC) method.
Results
Guided by the simulation, the key instrumental parameters were systematically optimized, and the reliability of the simulation model was validated experimentally. During the detection of acetone, benzene, and air mixtures, clean organic spectra were obtained in PI mode, whereas comprehensive detection of inorganic gases with high ionization energies (e.g., N 2 and O 2 ) and detailed fragment‐ion information were achieved in EI mode.
Conclusions
A DISQMS was developed and validated, and a multiphysics ion trajectory simulation framework was proposed to numerically optimize the instrument's sensitivity. It was demonstrated that complementary detection of organic and inorganic species is achieved through the dual‐mode capability. Specifically, spectral interpretation and molecular structure elucidation are effectively facilitated by combining the detailed fragment‐ion information from EI with the clean molecular ion data from PI.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- National Synchrotron Radiation Laboratory, University of Science and Technology of China, Hefei, China.
- Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, Anhui, China.
ORCID-Profile: Chengyuan Liu, Yang Pan
Förderung
- National Natural Science Foundation of China
- National Key Research and Development Program of China
- Science and Technology Plan Project of China Tobacco Anhui Industrial Co. Ltd.
- Key Project of Anhui Province Science and Technology Innovation Platform
- CAS Key Technology Talent Program, and USTC Research Funds of the Double First-Class Initiative
Zitieren
APA
Huang, J., Wen, Z., Gao, S., Wang, G., Tang, Y., Xu, M., Liu, C., & Pan, Y. (2026). Numerical Simulation and Experimental Study of a Dual‐Ionization Source (PI/EI) Quadrupole Mass Spectrometer Based on the CFD–DSMC Method. Rapid Communications in Mass Spectrometry, 40(12). https://doi.org/10.1002/rcm.70069
Vancouver
Huang J, Wen Z, Gao S, Wang G, Tang Y, Xu M, et al. Numerical Simulation and Experimental Study of a Dual‐Ionization Source (PI/EI) Quadrupole Mass Spectrometer Based on the CFD–DSMC Method. Rapid Communications in Mass Spectrometry. 2026;40(12). doi:10.1002/rcm.70069
BibTeX
@article{Huang2026,
author = {Huang, Jun and Wen, Zhiwei and Gao, Shijun and Wang, Guangqi and Tang, Yihan and Xu, Minggao and Liu, Chengyuan and Pan, Yang},
title = {Numerical Simulation and Experimental Study of a Dual‐Ionization Source (PI/EI) Quadrupole Mass Spectrometer Based on the CFD–DSMC Method},
journal = {Rapid Communications in Mass Spectrometry},
year = {2026},
volume = {40},
number = {12},
doi = {10.1002/rcm.70069}
}
EUVIMED