Direct Probe Ionisation Mass Spectrometry for Rapid and Accurate Determination of Perfluoroalkyl and Polyfluoroalkyl Substance Concentrations
Rapid Communications in Mass Spectrometry · 30.09.2025 · 39 (18)
DOI: 10.1002/rcm.10079 ↗Kurz gefasst
Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, online erschienen am 11.06.2025. Verfasst von Tim P Sidnell und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Chemistry and Chemical Engineering, University of Surrey, Guildford, UK. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 77 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Rationale
Analysis of perfluoroalkyl and polyfluoroalkyl substances (PFAS) is typically achieved using chromatographic techniques, such as high‐performance liquid chromatography, combined with mass spectrometry. These techniques can be complex, expensive, time consuming and can lead to inaccurate quantification due to sample container sorption and contamination by polytetrafluoroethylene (PTFE) components.
Methods
Here, we demonstrate the novel application of a Direct Probe ionisation Mass Spectrometer (DPiMS) to quantify four aqueous anionic PFAS; perfluorooctanoic acid (PFOA), perfluorooctane sulfonic acid (PFOS), perfluorononanoic acid (PFNA) and 1H,1H,2H,2H‐perfluorooctanesulfonic acid (6:2 fluorotelomer surfactant). The DPiMS was operated in both Single Ion Monitoring (SIM) and broad m/z scan range (50–500) modes and the cleaning methodology, ionising voltage and desolvation line heating temperature were optimised.
Results
The derived method quantified both single component and mixtures of PFAS from 0.5 nM to 200 μM (0.2 ppb to 108 ppm) without use of pre‐concentration. The advantages of this technique over chromatographic techniques are; the speed of analysis (≈3.5 min per sample, including blanks/cleaning), lack of PTFE components and simplified methodology.
Conclusions
The DPiMS is a useful tool for the rapid screening and estimation of PFAS concentrations with results that are comparable to existing methods. It is anticipated that better sensitivity could be achieved with the use of a triple quadrupole instrument.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Chemistry and Chemical Engineering, University of Surrey, Guildford, UK.
- Ecology & Environment Research Centre, Manchester Metropolitan University, Manchester, UK.
ORCID-Profile: Simone C Mathias, Patrick J Sears
Zitieren
APA
Sidnell, T. P., Mathias, S. C., Megson, D. P., Sears, P. J., & Bussemaker, M. J. (2025). Direct Probe Ionisation Mass Spectrometry for Rapid and Accurate Determination of Perfluoroalkyl and Polyfluoroalkyl Substance Concentrations. Rapid Communications in Mass Spectrometry, 39(18). https://doi.org/10.1002/rcm.10079
Vancouver
Sidnell TP, Mathias SC, Megson DP, Sears PJ, Bussemaker MJ. Direct Probe Ionisation Mass Spectrometry for Rapid and Accurate Determination of Perfluoroalkyl and Polyfluoroalkyl Substance Concentrations. Rapid Communications in Mass Spectrometry. 2025;39(18). doi:10.1002/rcm.10079
BibTeX
@article{Sidnell2025,
author = {Sidnell, Tim P and Mathias, Simone C and Megson, David P and Sears, Patrick J and Bussemaker, Madeleine J},
title = {Direct Probe Ionisation Mass Spectrometry for Rapid and Accurate Determination of Perfluoroalkyl and Polyfluoroalkyl Substance Concentrations},
journal = {Rapid Communications in Mass Spectrometry},
year = {2025},
volume = {39},
number = {18},
doi = {10.1002/rcm.10079}
}
EUVIMED