Optimisation of Heated Electrospray Ionisation Parameters to Minimise In‐Source Generated Impurities in the Analysis of Oligonucleotide Therapeutics

Mollie A Glenister iD, Ulrik Mistarz, Ken Cook, James L Stephenson, Mark J Dickman iD

Rapid Communications in Mass Spectrometry · 2025

DOI: 10.1002/rcm.10033 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, 2025. Verfasst von Mollie A Glenister und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Chemical, Materials and Biological Engineering, University of Sheffield, Sheffield, UK. Hauptthemen laut MEDLINE: Spectrometry, Mass, Electrospray Ionization, Oligonucleotides, Drug Contamination, Oligonucleotides, Antisense. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 61 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Biotechnology and Biological Sciences Research Council UK (BBSRC), Engineering and Physical Sciences Research Council (EPSRC).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

Oligonucleotides have emerged as an important new class of therapeutic. Due to their structural complexity, this presents significant challenges for the development of analytical methods to characterise and determine their impurity profile. In this study, we introduce a sensitive ion‐pair reverse phase method interfaced with mass spectrometry for analysis of antisense oligonucleotides and small interfering RNAs.

Methods

Liquid chromatography–mass spectrometry analysis of antisense oligonucleotides and small interfering RNAs was performed using hexylamine: hexafluoro‐2‐propanol mobiles phases. LC‐MS analysis was performed in both negative and positive ion mode. Electrospray ionisation source conditions including collision energy and temperature were optimised to minimise in‐source generated impurities and alkylamine adducts in the analysis of oligonucleotide therapeutics.

Results

The results show that under low or no in‐source collision energy the presence of hexylamine adducts are observed and are predominantly on the lowest charge states present. As the in‐source collision energy is increased, a reduction of hexylamine adducts is observed in conjunction with an increase in nucleobase loss in the gas phase, therefore generating in‐source impurities. In comparison to tributylammonium acetate, increased MS sensitivity, higher charge states and effective removal of hexylamine adducts using mild source conditions was achieved.

Conclusions

Optimisation of the mild source conditions in conjunction with high pH mobile phases was combined with high‐resolution accurate mass spectrometry analysis and automated deconvolution workflows to develop a simplified and streamlined approach for characterising oligonucleotide therapeutics and their related impurities.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spectrometry, Mass, Electrospray IonizationOligonucleotidesDrug ContaminationOligonucleotides, AntisenseChromatography, Liquid

Stichwörter der Autor:innen

electrospray ionisationion‐pair reversed phaseoligonucleotide therapeutics

Substanzen

OligonucleotidesOligonucleotides, Antisense

Autor:innen und Institutionen

  • School of Chemical, Materials and Biological Engineering, University of Sheffield, Sheffield, UK.
  • ThermoFisher Scientific, Allerød, Denmark.
  • ThermoFisher Scientific, Hemel Hempstead, UK.
  • ThermoFisher Scientific, Lexington, Massachusetts, USA.

ORCID-Profile: Mollie A Glenister, Mark J Dickman

Förderung

  • Biotechnology and Biological Sciences Research Council UK (BBSRC)
  • Engineering and Physical Sciences Research Council (EPSRC)

Zitieren

APA

Glenister, M. A., Mistarz, U., Cook, K., Stephenson, J. L., & Dickman, M. J. (2025). Optimisation of Heated Electrospray Ionisation Parameters to Minimise In‐Source Generated Impurities in the Analysis of Oligonucleotide Therapeutics. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.10033

Vancouver

Glenister MA, Mistarz U, Cook K, Stephenson JL, Dickman MJ. Optimisation of Heated Electrospray Ionisation Parameters to Minimise In‐Source Generated Impurities in the Analysis of Oligonucleotide Therapeutics. Rapid Communications in Mass Spectrometry. 2025. doi:10.1002/rcm.10033

BibTeX
@article{Glenister2025,
  author = {Glenister, Mollie A and Mistarz, Ulrik and Cook, Ken and Stephenson, James L and Dickman, Mark J},
  title = {Optimisation of Heated Electrospray Ionisation Parameters to Minimise In‐Source Generated Impurities in the Analysis of Oligonucleotide Therapeutics},
  journal = {Rapid Communications in Mass Spectrometry},
  year = {2025},
  doi = {10.1002/rcm.10033}
}
Kostenlos starten →