A Room‐Temperature Operational Alignment Magnetometer Utilizing Free‐Spin Precession
Advanced Quantum Technologies · 2024-03 · 7 (3)
DOI: 10.1002/qute.202300377 ↗Kurz gefasst
Fachartikel in der Zeitschrift Advanced Quantum Technologies, online erschienen am 10.01.2024. Verfasst von Ge Jin und 4 weiteren Autor:innen. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Beijing Municipality.
Abstract (Originalsprache)
An alignment magnetometer based on free‐spin precession (FSP) has been deployed in the geomagnetic range. A linearly polarized probe beam is utilized to create alignment in a buffer‐gas‐free paraffin‐coated natural abundance rubidium vapor cell at room temperature exposed to a pulsed radio frequency field. Furthermore, the utilization of hyperfine‐selective optical repumping employing linearly polarized beams amplifies the proportion of responsive atoms. As a consequence, this leads to a remarkable enhancement in the optical rotation signal. This paper reports measurements with a minor static magnetic field of about 1 inside a four‐layer ‐metal shield, achieving a system noise level of 0.16 ). With a static magnetic field amplitude of 48.5 (in proximity to Earth's magnetic field), the noise level is ).
Förderung
- National Natural Science Foundation of China
- Natural Science Foundation of Beijing Municipality
Zitieren
APA
Jin, G., shi, T., Zhao, Y., Zhang, H., & Zou, S. (2024). A Room‐Temperature Operational Alignment Magnetometer Utilizing Free‐Spin Precession. Advanced Quantum Technologies, 7(3). https://doi.org/10.1002/qute.202300377
Vancouver
Jin G, shi T, Zhao Y, Zhang H, Zou S. A Room‐Temperature Operational Alignment Magnetometer Utilizing Free‐Spin Precession. Advanced Quantum Technologies. 2024;7(3). doi:10.1002/qute.202300377
BibTeX
@article{Jin2024,
author = {Jin, Ge and shi, Tao and Zhao, Yan and Zhang, Hong and Zou, Sheng},
title = {A Room‐Temperature Operational Alignment Magnetometer Utilizing Free‐Spin Precession},
journal = {Advanced Quantum Technologies},
year = {2024},
volume = {7},
number = {3},
doi = {10.1002/qute.202300377}
}