Comparison of Rapid Arc and Intensity Modulated Radiotherapy in a True Beam Linear Accelerator for 6 MV: Application of AAPM TG‐119 tests in treatment planning and quality assurance
Precision Radiation Oncology · 2023-12 · 7 (4) · 256-267
DOI: 10.1002/pro6.1212 ↗Kurz gefasst
Fachartikel in der Zeitschrift Precision Radiation Oncology, online erschienen am 29.11.2023. Verfasst von Soumya Roy und 2 weiteren Autor:innen. Beteiligte Einrichtung: GLA University Mathura Uttar Pradesh India. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Objective
We developed rapid arc (RA) and intensity‐modulated radiotherapy (IMRT) plans based on the American Association of Physicists in Medicine Task Group‐119 (AAPM TG‐119) proposals and compared the planning and quality assurance results. Materials and Methods Two treatment plans were used for each study patient: one using 7–9 IMRT fields and the other using the two full‐arc RA methods. Plan optimization and dose calculations were performed using 6 MV photons and Eclipse TM with the anisotropic analytical algorithm (AAA). Task group (TG)‐119 described the planning objectives used to evaluate the treatment plans generated for this study. Point dose, planar fluence measurement, and trajectory log file analysis are used for quality assurance.
Results
The conformity index (CI) for treatment plans varied between 0.76–0.91 when using IMRT and 0.75–0.93 when using RA. The homogeneity index (HI) was approximately 0.10–0.24 (IMRT) and 0.07–0.23 (RA). The ratio of the total number of monitor units required for IMRT to that required for RA was between 0.87 and 2.14. The treatment log files exhibited higher gamma passing with RA than with IMRT.
Conclusion
RA plans were more effective than IMRT plans in achieving the test goals. Point dose measurements and electronic portal imaging device (EPID)‐based planar fluence measurements showed statistically insignificant differences between the IMRT and RA plan quality assurance (QA) results. However, the trajectory log file analysis exhibited better gamma‐passing results for RA than for IMRT.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- GLA University Mathura Uttar Pradesh India.
- Chief Medical Physicist, Department of Radiation Oncology Manipal Hospital, Dwarka New Delhi India.
- Chief Medical Physicist Apollo Gleneagles Hospitals Kolkata India.
- Centre for Cosmology, Astrophysics and Space Science GLA University Mathura Uttar Pradesh India.
ORCID-Profile: Soumya Roy, Biplab Sarkar, Anirudh Pradhan
Zitieren
APA
Roy, S., Sarkar, B., & Pradhan, A. (2023). Comparison of Rapid Arc and Intensity Modulated Radiotherapy in a True Beam Linear Accelerator for 6 MV: Application of AAPM TG‐119 tests in treatment planning and quality assurance. Precision Radiation Oncology, 7(4), 256-267. https://doi.org/10.1002/pro6.1212
Vancouver
Roy S, Sarkar B, Pradhan A. Comparison of Rapid Arc and Intensity Modulated Radiotherapy in a True Beam Linear Accelerator for 6 MV: Application of AAPM TG‐119 tests in treatment planning and quality assurance. Precision Radiation Oncology. 2023;7(4):256-267. doi:10.1002/pro6.1212
BibTeX
@article{Roy2023,
author = {Roy, Soumya and Sarkar, Biplab and Pradhan, Anirudh},
title = {Comparison of Rapid Arc and Intensity Modulated Radiotherapy in a True Beam Linear Accelerator for 6 MV: Application of AAPM TG‐119 tests in treatment planning and quality assurance},
journal = {Precision Radiation Oncology},
year = {2023},
volume = {7},
number = {4},
pages = {256--267},
doi = {10.1002/pro6.1212}
}
EUVIMED