Dosimetric response of Gafchromic™ EBT‐XD film to therapeutic protons

Fada Guan, Xiaochun Wang, Ming Yang, Emily Draeger, Dae Han, Kiminori Iga, Fanqing Guo, Luis Perles, Yuting Li, Narayan Sahoo, Radhe Mohan, Zhe Chen

Precision Radiation Oncology · 2023-03 · 7 (1) · 15-26

DOI: 10.1002/pro6.1187 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Precision Radiation Oncology, online erschienen am 02.03.2023. Verfasst von Fada Guan und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center,1515 Holcombe Boulevard, Houston, Texas 77030, USA. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NCI NIH HHS, National Cancer Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The EBT‐XD model of Gafchromic™ films has a broader optimal dynamic dose range, up to 40 Gy, compared with its predecessor models. This characteristic has made EBT‐XD films suitable for high‐dose applications, such as stereotactic body radiotherapy and stereotactic radiosurgery, as well as ultra‐high dose rate FLASH radiotherapy. The purpose of the current study was to characterize the dependence of EBT‐XD film response on linear energy transfer (LET) and dose rate of therapeutic protons from a synchrotron. A clinical spot‐scanning proton beam was used to study LET dependence at three dose‐averaged LET values of 1.0 keV/μm, 3.6 keV/μm, and 7.6 keV/μm. A research proton beamline was used to study dose rate dependence at 150 Gy/s in the FLASH mode and 0.3 Gy/s in the non‐FLASH mode. Film response data from dose‐averaged LET values of 0.9 keV/μm and 9.0 keV/μm of the proton FLASH beam were also compared. Film response data from a clinical 6‐MV photon beam were used as a reference. Both the gray value method and optical density (OD) method were used in film calibration. Calibration results using a specific OD calculation method and a generic OD calculation method were compared. The four‐parameter NIH Rodbard function and three‐parameter rational function were compared in fitting the calibration curves. Experimental results showed that the response of EBT‐XD film is proton LET dependent, but independent of dose rate. Goodness‐of‐fit analysis showed that using the NIH Rodbard function is superior for both protons and photons. Using the “specific OD + NIH Rodbard function” method for EBT‐XD film calibration is recommended.

Schlagwörter

Stichwörter der Autor:innen

EBT-XD filmFLASH dose ratelinear energy transfer (LET)proton therapy

Autor:innen und Institutionen

  • Department of Radiation Physics, The University of Texas MD Anderson Cancer Center,1515 Holcombe Boulevard, Houston, Texas 77030, USA.
  • Department of Therapeutic Radiology, Yale University School of Medicine, 333 Cedar Street, New Haven, Connecticut 06510, USA; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holc…
  • Particle Therapy Division, Hitachi America, Ltd, 2535 Augustine Drive, Santa Clara, California 95054, USA.

Förderung

  • NCI NIH HHS
  • National Cancer Institute

Zitieren

APA

Guan, F., Wang, X., Yang, M., Draeger, E., Han, D., Iga, K., Guo, F., Perles, L., Li, Y., Sahoo, N., Mohan, R., & Chen, Z. (2023). Dosimetric response of Gafchromic™ EBT‐XD film to therapeutic protons. Precision Radiation Oncology, 7(1), 15-26. https://doi.org/10.1002/pro6.1187

Vancouver

Guan F, Wang X, Yang M, Draeger E, Han D, Iga K, et al. Dosimetric response of Gafchromic™ EBT‐XD film to therapeutic protons. Precision Radiation Oncology. 2023;7(1):15-26. doi:10.1002/pro6.1187

BibTeX
@article{Guan2023,
  author = {Guan, Fada and Wang, Xiaochun and Yang, Ming and Draeger, Emily and Han, Dae and Iga, Kiminori and Guo, Fanqing and Perles, Luis and Li, Yuting and Sahoo, Narayan and Mohan, Radhe and Chen, Zhe},
  title = {Dosimetric response of Gafchromic™ EBT‐XD film to therapeutic protons},
  journal = {Precision Radiation Oncology},
  year = {2023},
  volume = {7},
  number = {1},
  pages = {15--26},
  doi = {10.1002/pro6.1187}
}
Kostenlos starten →