Analysis of different evaluation indexes for prostate stereotactic body radiation therapy plans: conformity index, homogeneity index and gradient index

Tingting Cao, Zhitao Dai, Zhen Ding, Wuzhou Li, Hong Quan

Precision Radiation Oncology · 2019-09 · 3 (3) · 72-79

DOI: 10.1002/pro6.1072 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Precision Radiation Oncology, online erschienen am 12.09.2019. Verfasst von Tingting Cao und 4 weiteren Autor:innen. Die Arbeit wurde bisher 39-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: China Postdoctoral Science Foundation.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study analyzed different definitions of the conformity index (CI), homogeneity index (HI), and gradient index (GI) used in evaluation of prostate cancer stereotactic body radiation therapy.

Methods

A total of 10 patients with localized prostate cancer staged T1–T2a were selected randomly, from which two stereotactic body radiation therapy plans were designed with CyberKnife and EDGE systems for each patient based on the same images and contours. CI, HI, and GI with different definitions were calculated based on a dose‐volume histogram of treatment plans.

Results

For four definitions of CI, the results showed that the values calculated by the Radiation Therapy Oncology Group, Lomax and Scheib, and Van't Riet were closer to 1 for EDGE plans, except for that by the Saint‐Anne, Lariboisiere, and Tenon group. Meanwhile, CI values of Van't Riet et al were lower than other definitions, which showed high accuracy to describe the conformity between the target volume and prescribed isodose line. For five definitions of HI, CyberKnife plans showed lower values than EDGE plans. HI calculated based on the D 2 ( D 5 ) and D 98 ( D 95 ) were lower than that calculated by the D max to D min , whereas the standard deviation was higher than others. For the dose GI, the results of effective radius and modified GI showed that EDGE plans had a steeper dose fall‐off, which is inverse to the conventional GI.

Conclusions

These indexes can all be objective tools for the evaluation of plan quality. Our results showed better conformity of dose distribution and dose gradient for EDGE plans, but better uniformity for CyberKnife plans. The conformation number and S‐index were recommended to accurately describe the conformity and uniformity of dose distribution. The ∆R iso and modified GI were also recommended to calculate the dose gradient.

Förderung

  • China Postdoctoral Science Foundation

Zitieren

APA

Cao, T., Dai, Z., Ding, Z., Li, W., & Quan, H. (2019). Analysis of different evaluation indexes for prostate stereotactic body radiation therapy plans: conformity index, homogeneity index and gradient index. Precision Radiation Oncology, 3(3), 72-79. https://doi.org/10.1002/pro6.1072

Vancouver

Cao T, Dai Z, Ding Z, Li W, Quan H. Analysis of different evaluation indexes for prostate stereotactic body radiation therapy plans: conformity index, homogeneity index and gradient index. Precision Radiation Oncology. 2019;3(3):72-79. doi:10.1002/pro6.1072

BibTeX
@article{Cao2019,
  author = {Cao, Tingting and Dai, Zhitao and Ding, Zhen and Li, Wuzhou and Quan, Hong},
  title = {Analysis of different evaluation indexes for prostate stereotactic body radiation therapy plans: conformity index, homogeneity index and gradient index},
  journal = {Precision Radiation Oncology},
  year = {2019},
  volume = {3},
  number = {3},
  pages = {72--79},
  doi = {10.1002/pro6.1072}
}
Kostenlos starten →