Oxygen Dilution Among Portable Home Ventilators in Children: The Influence of Prescribed Settings and Leak

Ryne Simpson iD, Teresa Magruder, Ajay Kasi, Nicholas Fitzpatrick iD, Brittany Truitt, Carolyn Henderson, Amit Agarwal, Jacob Kaslow iD, William Coleburn iD, Guillermo Beltran-Ale iD, Pediatric Mechanical Ventilation Society

Pediatric Pulmonology · 2026-03 · 61 (3)

DOI: 10.1002/ppul.71543 ↗

Kurz gefasst

Leserbrief in der Zeitschrift Pediatric Pulmonology, online erschienen am 01.03.2026. Verfasst von Ryne Simpson und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Division of Pediatric Pulmonology and Sleep Medicine, University of Alabama at Birmingham and Children's of Alabama, Birmin…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Children requiring invasive home mechanical ventilation (HMV) often receive respiratory support via tracheostomy tubes and single-limb, passive circuits (SLPC). Advantages of SLPCs—compared to active circuits—include greater simplicity, availability of volume-targeting modes, and superior leak compensation [1]. Two primary disadvantages of SLPCs are the inability to directly measure exhaled tidal volume and the variability in delivered fraction of inspired oxygen (FiO2) [1, 2]. Since portable home ventilators (PHV) typically utilize low-pressure oxygen sources with fixed flow rates, any increase in ventilator output (which occurs through increased ambient airflow) will dilute the effective concentration of oxygen ultimately delivered to the patient. Certain factors leading to increased ventilator output include intentional and unintentional leaks as well as higher levels of baseline respiratory support. Several PHVs are approved for use in infants and children, and while their functionality is generally comparable, variations in ventilation modes and device-specific, proprietary algorithms (e.g., leak compensation) likely influence ventilator monitoring and output [3]. It is not known whether these differences independently affect oxygen delivery to the patient. This study aims to evaluate several different PHVs utilizing SLPCs with low-flow oxygen and characterize the influence of leak and setting prescription on effective FiO2 delivery. We analyzed the effective FiO2 delivery of multiple PHVs when subjected to different leak and respiratory support scenarios. Testing was completed at four pediatric centers; ventilators were chosen based on local availability. Testing was designed to simulate the care of pediatric patients with chronic respiratory failure requiring invasive HMV. Accordingly, PHVs were set in pressure-controlled ventilation using SLPCs with low-flow supplemental oxygen. The test lung was calibrated to mimic a pediatric patient. We tested five home ventilator models: the Trilogy EV300/EVO (Philips Respironics; three centers), Trilogy 100 (Philips Respironics; two centers), Vivo 45 LS (Breas; three centers), Astral 150 (ResMed, one center), and V*Home/V+Pro (React Health, four centers). Each ventilator was connected to a 0.5 L infant test lung (SmartLung; IMT Analytics AG). The test lung was set to a compliance of 5 mL/mbar and a resistance of 50 mbar/L/s. Each ventilator was connected to the test lung with an SLPC (Fischer and Paykel, adult) and a Whisper-Swivel exhalation valve (Respironics). An external oxygen analyzer (Handi+/Max O2+/Max O2 ME by Maxtec) was placed between the Whisper-Swivel and the test lung. Oxygen analyzers were chosen based on center availability; each analyzer was calibrated to 21% FiO2 in room air prior to use. Circuit compliance/resistance calibration (i.e., pre-use testing) was performed, when available, for each ventilator prior to testing. The ventilators were set in a pressure-controlled mode of ventilation. Low: Peak inspiratory pressure of 20 cmH2O, positive end expiratory pressure of 8 cmH2O, inspiratory time of 0.7 s, and mandatory breath rate of 20 breaths per minute. High: Peak inspiratory pressure of 30 cmH2O, positive end expiratory pressure of 8 cmH2O, inspiratory time of 0.7 s, and mandatory breath rate of 40 breaths per minute. Low leak: Intrinsic, fixed leak of exhalation valve; no additional leak. High leak: From the “low” leak setting, the adjustable leak port on the test lung (see Figure 1) was titrated to achieve a roughly 20% increase in ventilator-reported leak (example: Low leak = 35 LPM [liter per minute], high leak titrated to 41−43 LPM). Low settings, low leak (LS, LL) Low settings, high leak (LS, HL) High settings, low leak (HS, LL) High settings, high leak (HS, HL) The FiO2 results were compiled from all testing centers; averages and ranges were calculated in a typical fashion and are reported below. In this simulation of pediatric invasive HMV, PHVs with SLPCs show variability in delivered FiO2 when subjected to different levels of leak and respiratory support (Table 1). This variability appears less clinically relevant with lower flows of supplemental oxygen but becomes more significant as flow increases (Table 2). Branson et al. previously demonstrated that FiO2 delivered during noninvasive ventilation using PHV is influenced by ventilatory mode (pressure vs. volume), site of oxygen entrainment, and the minute ventilation generated [4]. Our findings reaffirm aspects of that analysis; however, to our knowledge, this is the first study to analyze actual delivered FiO2 for PHVs using SLPCs with a pediatric, invasive ventilatory model. All ventilators demonstrated similar performance trends. At low supplemental oxygen levels, increased circuit leak or respiratory support had minimal impact on delivered FiO2. At higher oxygen flows, FiO2 variability became more pronounced due to the greater relative influence of increased ambient minute ventilation/inspiratory flow on entrained oxygen. These findings highlight the need for close monitoring of infants and children sensitive to oxygen fluctuations, especially when high ventilatory support or circuit leak is present. In these situations, the use of ventilators with active circuits or oxygen blenders, which allow for precise and consistent FiO2 delivery, may be indicated. It should be noted that depending on the type of device or level of respiratory support required, use of these ventilators may preclude transition to the outpatient setting. For instance, some durable medical equipment companies or home health providers may lack experience managing children with active circuit configurations at home, and oxygen blenders typically require high-pressure oxygen input not currently available for outpatient use. When comparing ventilator models, there is clear variability in delivered FiO2, even when facing identical clinical circumstances. Similarly, this appears to be less of a concern with lower amounts of supplemental oxygen, but may be more significant at higher flow levels. This finding is especially relevant when transitioning between ventilators. Children requiring greater amounts of supplemental oxygen may need adjustments in oxygen flow following ventilator transition; this should not be assumed to be from changing clinical status. The authors suspect that variations in how ventilator parameters are monitored, circuit calibration programs (i.e., pre-use testing), and leak compensation algorithms affect each ventilator's response to external factors. These differing responses likely influence the degree of ambient airflow output, and in turn, the dilution of supplemental oxygen delivered to the patient. Finally, this analysis provides an important reference for estimating delivered FiO2 in pediatric patients requiring invasive HMV. Our findings reinforce that certain guidelines such as the “4:1 rule” for effective FiO2 in nasal cannula delivery (i.e., a 4% increase in FiO2 for every LPM) have no utility when discussing mechanical ventilation with SLPCs [5]. Children transitioning from active-circuit, intensive care ventilators to low-flow PHVs with SLPCs may benefit from higher than anticipated supplemental oxygen levels to maximize the likelihood of a successful transition. Additionally, consistent clinical use of external oxygen analyzers is encouraged in these situations. Although pressure and respiratory rate were not assessed as independent variables, the evidence suggests that delivered FiO2% is influenced by ventilator-generated minute ventilation. The authors suspect that both pressure and mandatory rate independently contribute to FiO2 dilution, with the magnitude ultimately determined by resultant changes in minute ventilation. Additionally, while certain ventilators trended towards greater internal FiO2 variability across testing sites, directly comparing brands is challenging because certain models were evaluated at only one or two centers. While this bench analysis was highly controlled, there are several limitations worth mentioning. First, although leak titration was objective, any slight over/underestimation may have led to outsized differences in FiO2 measurements between centers, especially at higher flows of oxygen. Second, each center utilized wall-mounted medical gas flow-regulators to deliver supplemental oxygen. These regulators, while standard of care, can be imprecise at higher flow amounts, potentially contributing to inter-center variability. In a similar vein, home oxygen concentrators, which were not tested, are unable to achieve 100% oxygen concentrations, potentially contributing to additional oxygen variability. Third, alternative interfaces (e.g., noninvasive) and modes of ventilation (e.g., volume-targeted, pressure support) likely influence ventilator output in different ways. As such, further evaluation of these clinical scenarios is likely worthwhile. Finally, it is important to note that other commercially available, passive exhalation valves provide different levels of intentional circuit leak; therefore, the results of this study cannot be directly extrapolated to other circuit-exhalation valve combinations, although the authors suspect the reported trends generally hold true. In a simulated model of pediatric invasive HMV, effective FiO2 is influenced by the amount of circuit leak and respiratory support. This effect is more pronounced at higher flows of supplemental oxygen. Providers should be aware of these differences when transitioning patients between ventilators, especially in children who are sensitive to changes in FiO2. Dr. Ryne Simpson conceptualized the project, drafted the manuscript, performed testing, curated data, and reviewed and revised the manuscript critically for important intellectual content. Drs. Guillermo Beltran-Ale, Teresa Magruder, Amit Agarwal, Ajay Kasi, Nicholas Fitzpatrick, Brittany Truitt, Carolyn Henderson, and Jacob Kaslow conceptualized the project, performed testing, and reviewed and revised the manuscript critically for important intellectual content. Mr. William Coleburn performed testing and reviewed the manuscript for critically important intellectual content. The authors received no specific funding for this work. Dr. Ryne Simpson serves on the Scientific Advisory Board for React Health. The other authors declare no conflicts of interest. The data that support the findings of this study are available from the corresponding author upon reasonable request.

Schlagwörter

Stichwörter der Autor:innen

mechanical ventilationpediatricssupplemental oxygenventilator

Autor:innen und Institutionen

  • Department of Pediatrics, Division of Pediatric Pulmonology and Sleep Medicine, University of Alabama at Birmingham and Children's of Alabama, Birmingham, Alabama, USA.
  • Department of Pediatrics, Division of Pediatric Pulmonology Medicine, Emory University and Children's Healthcare of Atlanta, Atlanta, Georgia, USA.
  • Department of Pediatrics, Division of Pediatric Pulmonology, Baylor University and Texas Children's Hospital, Houston, Texas, USA.
  • Department of Pediatrics, Division of Pulmonary Medicine, Vanderbilt University and Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • University of Tennessee at Knoxville, Knoxville, Tennessee, USA.

ORCID-Profile: Ryne Simpson, Nicholas Fitzpatrick, Jacob Kaslow, William Coleburn, Guillermo Beltran-Ale

Zitieren

APA

Simpson, R., Magruder, T., Kasi, A., Fitzpatrick, N., Truitt, B., Henderson, C., Agarwal, A., Kaslow, J., Coleburn, W., Beltran-Ale, G., & Pediatric Mechanical Ventilation Society (2026). Oxygen Dilution Among Portable Home Ventilators in Children: The Influence of Prescribed Settings and Leak. Pediatric Pulmonology, 61(3). https://doi.org/10.1002/ppul.71543

Vancouver

Simpson R, Magruder T, Kasi A, Fitzpatrick N, Truitt B, Henderson C, et al. Oxygen Dilution Among Portable Home Ventilators in Children: The Influence of Prescribed Settings and Leak. Pediatric Pulmonology. 2026;61(3). doi:10.1002/ppul.71543

BibTeX
@article{Simpson2026,
  author = {Simpson, Ryne and Magruder, Teresa and Kasi, Ajay and Fitzpatrick, Nicholas and Truitt, Brittany and Henderson, Carolyn and Agarwal, Amit and Kaslow, Jacob and Coleburn, William and Beltran-Ale, Guillermo and Pediatric Mechanical Ventilation Society, Pediatric Mechanical Ventilation Society},
  title = {Oxygen Dilution Among Portable Home Ventilators in Children: The Influence of Prescribed Settings and Leak},
  journal = {Pediatric Pulmonology},
  year = {2026},
  volume = {61},
  number = {3},
  doi = {10.1002/ppul.71543}
}
Kostenlos starten →