Efficacy of a prototype inspiratory‐synchronized small particle versus conventional vibrating mesh nebulizer during pediatric and neonatal mechanical ventilation
Pediatric Pulmonology · 2025
DOI: 10.1002/ppul.27356 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Pediatric Pulmonology, 2025. Verfasst von Jie Li und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiopulmonary Sciences, Division of Respiratory Care, Rush University, Chicago, Illinois, USA. Hauptthemen laut MEDLINE: Nebulizers and Vaporizers, Albuterol, Respiration, Artificial, Bronchodilator Agents. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben. Gefördert von: None.
Themen: Beatmung und Weaning
Abstract (Originalsprache)
Background
An inspiration‐synchronized vibrating mesh nebulizer (VMN) has been reported to improve aerosol delivery during adult mechanical ventilation. A prototype VMN generating smaller particles was developed. We aimed to compare the aerosol delivery efficiency of small‐particle and conventional VMNs in inspiration‐synchronized and continuous modes during neonatal and pediatric mechanical ventilation.
Methods
A critical care ventilator with heated humidified circuits connected to an endotracheal tube (ETT) and passive test lung was set to pediatric and neonate parameters. Albuterol (2.5 mg/ml, 1 ml) was administered using both small‐particle and conventional VMNs in inspiration‐synchronization and continuous modes. For the pediatric model, VMN was placed at the humidifier inlet, inspiratory limb at Y‐piece, and between Y‐piece and ETT (Y‐ETT). For the neonatal model, VMN was placed at the humidifier inlet and between Y‐ETT. Each setup was repeated five times. Albuterol collected on the filter distal to the ETT was eluted and assayed with UV spectrophotometry (276 nm).
Results
The inspiration‐synchronized VMN generated higher inhaled doses compared to continuous VMN across all nebulizer placements, particle sizes, and aerosol generation models (all p < .05). The highest inhaled doses (42.2 ± 2.0% and 40.7 ± 1.0% for pediatric and neonate, respectively) were observed with the small‐particle inspiration‐synchronized VMN placed at Y‐ETT. In the pediatric model, the inhaled dose with inspiration‐synchronized conventional VMN was similar, independent of nebulizer placements (24.4 ~ 27.0%). In contrast, the inhaled dose was greatest with continuous VMN placed at the humidifier inlet. With the neonatal model, VMN placed at Y‐ETT yielded higher doses than the humidifier inlet, and small‐particle VMNs outperformed conventional VMNs across all settings (all p < .05).
Conclusion
The prototype small‐particle VMN positioned between Y‐piece and ETT in an inspiration‐synchronized mode optimized aerosol delivery during mechanical ventilation in both pediatric and neonatal models.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiopulmonary Sciences, Division of Respiratory Care, Rush University, Chicago, Illinois, USA.
- Department of Respiratory Care, Advocate Children's Hospital in Park Ridge, Park Ridge, Illinois, USA.
- Aerogen Pharma Corp., San Mateo, California, USA.
ORCID-Profile: Jie Li
Förderung
- None
Zitieren
APA
Li, J., Karabelas, P., Gong, L., Sheridan, C. A., & Fink, J. B. (2025). Efficacy of a prototype inspiratory‐synchronized small particle versus conventional vibrating mesh nebulizer during pediatric and neonatal mechanical ventilation. Pediatric Pulmonology. https://doi.org/10.1002/ppul.27356
Vancouver
Li J, Karabelas P, Gong L, Sheridan CA, Fink JB. Efficacy of a prototype inspiratory‐synchronized small particle versus conventional vibrating mesh nebulizer during pediatric and neonatal mechanical ventilation. Pediatric Pulmonology. 2025. doi:10.1002/ppul.27356
BibTeX
@article{Li2025,
author = {Li, Jie and Karabelas, Paul and Gong, Lingyue and Sheridan, Caylie A and Fink, James B},
title = {Efficacy of a prototype inspiratory‐synchronized small particle versus conventional vibrating mesh nebulizer during pediatric and neonatal mechanical ventilation},
journal = {Pediatric Pulmonology},
year = {2025},
doi = {10.1002/ppul.27356}
}
EUVIMED