EUVIMEDEUVIMED entdecken
VergleichsstudieMultizentrische StudieRandomisierte kontrollierte Studie

Automated Versus Manual Endotracheal Tube Cuff Pressure Control in Mechanically Ventilated Children: A Randomized Crossover Trial

Ekin Soydan iD, Esat Erdem Gokpınar iD, Ozlem Karakaya iD, Zeynep Gökoglu Sevinc iD, Gokhan Ceylan iD, Mehmet Gurkan Erbas iD, Selma Albayrak iD, Özlem Demirel iD, Ece Dorsan Yay iD, Kaan Aslan iD, Nihal Karaöz Özdamar iD, Nuriye Turgut iD und 2 weitere

Pediatric Anesthesia · 2026

DOI: 10.1002/pan.70268 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Pediatric Anesthesia, 2026. Verfasst von Ekin Soydan und 13 weiteren Autor:innen. Beteiligte Einrichtung: Aydın Maternity and Children's Hospital, Aydın, Turkey. Hauptthemen laut MEDLINE: Intubation, Intratracheal, Respiration, Artificial. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Optimal endotracheal tube cuff pressure is key for airway sealing and minimizing microaspiration and tracheal injury in ventilated children. However, intermittent manual monitoring can allow significant pressure changes and be affected by measurement artifacts, potentially reducing airway protection.

Aims

This study evaluated whether automated cuff pressure control better maintains target pressure than manual monitoring in ventilated children, and assessed its effects on off‐target pressure and workload.

Methods

In this prospective, randomized crossover study, mechanically ventilated pediatric patients (1 month–18 years) expected to require ventilation for ≥ 48 h underwent two randomized 24‐h monitoring phases (manual and automated). During the manual phase, cuff pressure was continuously displayed via a closed system without repeated disconnection of the manometer and was intermittently adjusted by bedside nurses according to a predefined protocol. During the automated phase, cuff pressure was continuously regulated using an automated control system to maintain a target range of 15–25 cmH 2 O. The primary outcomes were the frequency and cumulative duration of off‐target cuff pressure exposure; secondary outcomes included nursing workload.

Results

Thirty‐two patients completed both phases (768 measurements). Low cuff pressure episodes (< 15 cmH 2 O) were more frequent during manual monitoring than automated monitoring (12.7% vs. 1.3%, p < 0.001), while high‐pressure episodes (> 25 cmH 2 O) occurred only during manual monitoring (2.6% vs. 0%, p < 0.001). Total time outside the target range was markedly reduced with automated monitoring (118 vs. 10 h, p < 0.001). Nursing workload was significantly lower in the automated phase (3.65 h vs. 0.08 h, p < 0.001).

Conclusions

Automated cuff pressure control more reliably maintains target pressure and minimizes clinically important deviations compared to manual management. These results support adopting continuous automated regulation as a preferred strategy to enhance airway safety and reduce workload in mechanically ventilated children. Trial Registration ClinicalTrials.gov (NCT06965400)

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intubation, IntratrachealRespiration, ArtificialAdolescentChildChild, PreschoolFemaleHumansInfantMaleAutomationCross-Over StudiesPressureProspective Studies

Stichwörter der Autor:innen

automated monitoringcuff pressureendotracheal tubemechanical ventilationpediatric intensive care

Autor:innen und Institutionen

  • Aydın Maternity and Children's Hospital, Aydın, Turkey.
  • University of Health Sciences, Dr. Behçet Uz Children's Hospital, İzmir, Turkey.

ORCID-Profile: Ekin Soydan, Esat Erdem Gokpınar, Ozlem Karakaya, Zeynep Gökoglu Sevinc, Gokhan Ceylan, Mehmet Gurkan Erbas, Selma Albayrak, Özlem Demirel, Ece Dorsan Yay, Kaan Aslan, Nihal Karaöz Özdamar, Nuriye Turgut, Utku Karaarslan, Hasan Agin

Zitieren

APA

Soydan, E., Gokpınar, E. E., Karakaya, O., Sevinc, Z. G., Ceylan, G., Erbas, M. G., Albayrak, S., Demirel, Ö., Yay, E. D., Aslan, K., Özdamar, N. K., Turgut, N., Karaarslan, U., & Agin, H. (2026). Automated Versus Manual Endotracheal Tube Cuff Pressure Control in Mechanically Ventilated Children: A Randomized Crossover Trial. Pediatric Anesthesia. https://doi.org/10.1002/pan.70268

Vancouver

Soydan E, Gokpınar EE, Karakaya O, Sevinc ZG, Ceylan G, Erbas MG, et al. Automated Versus Manual Endotracheal Tube Cuff Pressure Control in Mechanically Ventilated Children: A Randomized Crossover Trial. Pediatric Anesthesia. 2026. doi:10.1002/pan.70268