Fachartikel in der Zeitschrift Otolaryngology–Head and Neck Surgery, 2024. Verfasst von Vishal Narwani und 2 weiteren Autor:innen. Beteiligte Einrichtung: Division of Otolaryngology, Department of Surgery, Yale School of Medicine, New Haven, Connecticut, USA. Hauptthemen laut MEDLINE: Tracheostomy, Hemorrhage. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 30 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Tracheotomy is one of the most common procedures. Although tracheostomy complications have been extensively studied, literature related to device complications is scarce. The objective of this study is to describe complications associated with tracheostomies utilizing the Manufacturer and User Facility Device Experience (MAUDE) database. Study Design Retrospective cross‐sectional study.

Setting

The US Food and Drug Administration's (FDA) MAUDE database (2015‐2020).

Methods

The FDA's MAUDE database was queried for all reports on adverse events related to tracheostomy from January 1, 2015 to December 31, 2020.

Results

A total of 3086 adverse events related to open tracheostomy and 52 related to percutaneous tracheostomy were identified. For open tracheostomy, 2872 (93%), were related to device malfunction, and 214 (7%) consisted of patient‐related factors. The most frequently reported device‐related adverse event was cuff malfunction, with 1834 (59%) reported events, which includes cuff deflation, pilot balloon malfunction, and cuff inflation line malfunction. The most frequently reported patient‐related adverse events were tracheostomy tube obstruction with 67 events (2%). For percutaneous tracheostomy, 38 (73%) events were related to device malfunction, and 14 (27%) were related to patient injury. The most frequently reported adverse events were cuff malfunction (29%), safety ridge malfunction (17%), and bleeding (10%).

Conclusion

The MAUDE database is a useful tool that can be utilized to complement existing literature in identifying common and rare adverse events associated with tracheostomy device‐related failures, which are mostly reliant on isolated, published case reports.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TracheostomyHemorrhageHumansUnited StatesRetrospective StudiesCross-Sectional StudiesPostoperative ComplicationsDatabases, FactualUnited States Food and Drug Administration

Stichwörter der Autor:innen

MAUDEadverse eventballoon malfunctiontracheostomytracheotomy

Autor:innen und Institutionen

  • Division of Otolaryngology, Department of Surgery, Yale School of Medicine, New Haven, Connecticut, USA.
  • Warren Alpert Medical School of Brown University, Providence, Rhode Island, USA.

Zitieren

APA

Narwani, V., Dacey, S., & Lerner, M. Z. (2024). Adverse Events Associated With Tracheostomy: A MAUDE Database Analysis. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.501

Vancouver

Narwani V, Dacey S, Lerner MZ. Adverse Events Associated With Tracheostomy: A MAUDE Database Analysis. Otolaryngology–Head and Neck Surgery. 2024. doi:10.1002/ohn.501

BibTeX
@article{Narwani2024,
  author = {Narwani, Vishal and Dacey, Sydney and Lerner, Michael Z},
  title = {Adverse Events Associated With Tracheostomy: A MAUDE Database Analysis},
  journal = {Otolaryngology–Head and Neck Surgery},
  year = {2024},
  doi = {10.1002/ohn.501}
}
Kostenlos starten →