Magnetic Resonance Imaging Artifact Associated With Transcutaneous Bone Conduction Implants: Cholesteatoma and Vestibular Schwannoma Surveillance
Otolaryngology–Head and Neck Surgery · 2024
DOI: 10.1002/ohn.474 ↗Kurz gefasst
Fachartikel in der Zeitschrift Otolaryngology–Head and Neck Surgery, 2024. Verfasst von Ashley M Nassiri und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, University of Colorado Anschutz School of Medicine, Aurora, Colorado, USA. Hauptthemen laut MEDLINE: Neuroma, Acoustic, Cholesteatoma, Cochlear Implants. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
Objective
To evaluate the magnetic resonance (MR) image artifact and image distortion associated with the two transcutaneous bone conduction implants currently available in the United States. Study Design Cadaveric study.
Methods
Two cadaveric head specimens (1 male, 1 female) were unilaterally implanted according to manufacturer guidelines and underwent MR imaging (General Electric and Siemens 1.5 T scanners) under the following device conditions: (1) no device, (2) Cochlear Osia with magnet and headwrap, (3) Cochlear Osia without magnet, and (4) MED‐EL Bonebridge with magnet. Maximum metal mitigation techniques were employed in all conditions, and identical sequences were obtained. Blinded image scoring (diagnostic vs nondiagnostic image) was performed by experienced neuroradiologists according to anatomical subsites.
Results
All device conditions produced artifact and image distortion. The Osia with magnet produced diagnostic T1‐ and T2‐weighted images of the ipsilateral temporal bone, however, non‐echo planar imaging diffusion‐weighted imaging (DWI) was nondiagnostic. The Osia without magnet scanned on the Siemens MR imaging demonstrated the least amount of artifact and was the only condition that allowed for diagnostic imaging of the ipsilateral temporal bone on DWI. The Bonebridge produced a large area of artifact and distortion with the involvement of the ipsilateral and contralateral temporal bones.
Conclusion
In summary, of the three device conditions (Osia with magnet, Osia without magnet, and Bonebridge), Osia without magnet offered the least amount of artifact and distortion and was the only condition in which diagnostic DWI was available for the middle ear and mastoid regions on the Siemens MR imaging scanner.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, University of Colorado Anschutz School of Medicine, Aurora, Colorado, USA.
- Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
- Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, Minnesota, USA.
ORCID-Profile: Ashley M Nassiri
Zitieren
APA
Nassiri, A. M., Messina, S. A., Benson, J. C., Lane, J. I., McGee, K. P., Trzasko, J. D., & Carlson, M. L. (2024). Magnetic Resonance Imaging Artifact Associated With Transcutaneous Bone Conduction Implants: Cholesteatoma and Vestibular Schwannoma Surveillance. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.474
Vancouver
Nassiri AM, Messina SA, Benson JC, Lane JI, McGee KP, Trzasko JD, et al. Magnetic Resonance Imaging Artifact Associated With Transcutaneous Bone Conduction Implants: Cholesteatoma and Vestibular Schwannoma Surveillance. Otolaryngology–Head and Neck Surgery. 2024. doi:10.1002/ohn.474
BibTeX
@article{Nassiri2024,
author = {Nassiri, Ashley M and Messina, Steven A and Benson, John C and Lane, John I and McGee, Kiaran P and Trzasko, Joshua D and Carlson, Matthew L},
title = {Magnetic Resonance Imaging Artifact Associated With Transcutaneous Bone Conduction Implants: Cholesteatoma and Vestibular Schwannoma Surveillance},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2024},
doi = {10.1002/ohn.474}
}
EUVIMED