Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants

Petr Kordac, Milan Hajek, Petr Sedivy, Dita Pajuelo, Martin Kveton, Ondrej Fabian, Iva Subhanová, Ferdinand Los, Monika Dezortova

NMR in Biomedicine · 2026-12 · 39 (12)

DOI: 10.1002/nbm.70440 ↗

Kurz gefasst

Fachartikel in der Zeitschrift NMR in Biomedicine, online erschienen am 16.11.2026. Verfasst von Petr Kordac und 8 weiteren Autor:innen. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministerstvo Zdravotnictví Ceské Republiky.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

MR relaxometry has become the preferred noninvasive method for estimating hepatic iron from R2 and R2* relaxation rates. However, the precise quantification of liver iron concentration (LIC) requires calibration, and published equations relating relaxation rates to LIC have limited transferability across scanners and acquisition protocols. This study aimed to establish local 3‐T calibration equations relating R2 and R2* relaxation rates to chemically determined LIC using human liver explants and to evaluate their applicability in independent in vivo cohorts. Thirty‐six human liver explants underwent MR examination on a 3‐T system using multi‐echo spin‐echo and multi‐echo gradient‐echo sequences to obtain R2 and R2* relaxation rates. LIC was determined chemically by atomic absorption spectrophotometry. Calibration equations were derived using linear regression and compared with previously published models. The R2*‐based calibration was subsequently applied to independent in vivo cohorts examined using the same MR system and comparable acquisition protocols. Both MR methods correlated strongly with chemically determined LIC (R2*: r = 0.95, p < 0.001; R2: r = 0.92, p < 0.001). Compared with published models, the local calibrations demonstrated lower prediction errors and narrower limits of agreement within the studied explant cohort. Application of the R2* equation to in vivo datasets yielded physiologically plausible LIC estimates, with the lowest values observed in patients with primary sclerosing cholangitis, intermediate values in healthy controls and the highest values in subjects with elevated serum ferritin concentrations. Explant‐derived 3‐T calibration equations demonstrated strong association with chemically measured LIC and generated physiologically plausible LIC estimates in vivo. These findings support the feasibility of deriving local MR relaxometry calibrations from human liver explants, while further prospective validation against an independent LIC reference is required before clinical application.

Förderung

  • Ministerstvo Zdravotnictví Ceské Republiky

Zitieren

APA

Kordac, P., Hajek, M., Sedivy, P., Pajuelo, D., Kveton, M., Fabian, O., Subhanová, I., Los, F., & Dezortova, M. (2026). Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants. NMR in Biomedicine, 39(12). https://doi.org/10.1002/nbm.70440

Vancouver

Kordac P, Hajek M, Sedivy P, Pajuelo D, Kveton M, Fabian O, et al. Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants. NMR in Biomedicine. 2026;39(12). doi:10.1002/nbm.70440

BibTeX
@article{Kordac2026,
  author = {Kordac, Petr and Hajek, Milan and Sedivy, Petr and Pajuelo, Dita and Kveton, Martin and Fabian, Ondrej and Subhanová, Iva and Los, Ferdinand and Dezortova, Monika},
  title = {Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants},
  journal = {NMR in Biomedicine},
  year = {2026},
  volume = {39},
  number = {12},
  doi = {10.1002/nbm.70440}
}
Kostenlos starten →