Fachartikel in der Zeitschrift NMR in Biomedicine, 2026. Verfasst von Kirsten J Schmidt und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Lung, Magnetic Resonance Imaging, Perfusion. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: German Center for Lung Research (DZL), Deutsches Zentrum für Lungenforschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Phase‐resolved functional lung (PREFUL) MRI enables contrast‐free lung perfusion quantification (Q Quant ). Current quantification uses a single median value per slice of maximal blood exchange fraction (ΔS), which limits regional assessment, whereas regional ΔS causes artifacts. This retrospective study aims to optimize Q Quant by applying Gaussian filtering to ΔS to improve intrascan/interscan repeatability in healthy volunteers, while evaluating whether the optimized Q Quant retains agreement with dynamic contrast‐enhanced (DCE) MRI in chronic obstructive pulmonary disease (COPD) and chronic thromboembolic pulmonary hypertension (CTEPH) patients and exhibits a physiologically plausible gravity‐dependent perfusion gradient. Analysis was performed on 29 healthy subjects (11 female, age: 19–33 years), 20 COPD patients (three female, age: 46–76 years) and 15 CTEPH patients (six female, age: 44–82 years), scanned at 1.5 T. Healthy subjects underwent four PREFUL scans in two sessions. Patients just underwent one PREFUL and DCE scan. Q Quant maps were calculated while applying different Gaussian filter options to ΔS. In patients, PREFUL‐derived Q Quant maps were coregistered to DCE‐derived pulmonary blood flow (PBF) maps for global and voxel‐wise correlations. In healthy subjects, the gravity‐dependent perfusion gradient was quantified by linear regression of slice‐wise Q Quant across five coronal slices. Gaussian filtering ΔS with σ = 1.5 improved intrascan and interscan repeatability in healthy subjects (mean bias: ≤ 2.94 mL/100 mL; Cohen's d : 0.2) compared with the current slice median setting (mean bias: ≥ 3.75 mL/100 mL; Cohen's d : 0.68). Correlations with PBF remained stable after filtering (σ = 1.5) in COPD (filtered vs. default ΔS: global r = 0.57 vs. r = 0.59, voxel‐wise r = 0.32 vs. r = 0.32) and CTEPH (global r = 0.32 vs. r = 0.28, voxel‐wise r = 0.40 vs. r = 0.41), all significant except global CTEPH. The filtered Q Quant maps showed a significant gravity‐dependent perfusion gradient (ΔS filtered σ = 1.5) of 1.47 mL/min/100 mL/cm. Gaussian filtering ΔS with σ = 1.5 improved Q Quant repeatability while preserving DCE‐MRI agreement and is therefore recommended. The observed gravity‐dependent perfusion gradient confirms physiological plausibility, supporting potential in clinical use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LungMagnetic Resonance ImagingPerfusionHumansFemaleMiddle AgedAgedAdultNormal DistributionMaleAged, 80 and overReproducibility of ResultsPulmonary Disease, Chronic ObstructiveYoung AdultPerfusion Magnetic Resonance Imaging

Stichwörter der Autor:innen

DCE‐MRIPREFUL MRIperfusionpulmonary MRIrepeatabilityvalidation

Autor:innen und Institutionen

  • Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research, Hannover, Germany.
  • Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
  • Department of Radiation Protection and Medical Physics, Hannover Medical School, Hannover, Germany.

ORCID-Profile: Kirsten J Schmidt, Andreas Voskrebenzev, Julian M W Glandorf, Marius M Klein, Lea Behrendt, Frank Wacker, Filip Klimeš

Förderung

  • German Center for Lung Research (DZL)
  • Deutsches Zentrum für Lungenforschung

Zitieren

APA

Schmidt, K. J., Voskrebenzev, A., Glandorf, J. M. W., Klein, M. M., Behrendt, L., Wacker, F., Vogel-Claussen, J., & Klimeš, F. (2026). Evaluating Perfusion Quantification for Phase‐Resolved Functional Lung MRI Using Gaussian Filtering. NMR in Biomedicine. https://doi.org/10.1002/nbm.70420

Vancouver

Schmidt KJ, Voskrebenzev A, Glandorf JMW, Klein MM, Behrendt L, Wacker F, et al. Evaluating Perfusion Quantification for Phase‐Resolved Functional Lung MRI Using Gaussian Filtering. NMR in Biomedicine. 2026. doi:10.1002/nbm.70420

BibTeX
@article{Schmidt2026,
  author = {Schmidt, Kirsten J and Voskrebenzev, Andreas and Glandorf, Julian M W and Klein, Marius M and Behrendt, Lea and Wacker, Frank and Vogel-Claussen, Jens and Klimeš, Filip},
  title = {Evaluating Perfusion Quantification for Phase‐Resolved Functional Lung MRI Using Gaussian Filtering},
  journal = {NMR in Biomedicine},
  year = {2026},
  doi = {10.1002/nbm.70420}
}
Kostenlos starten →