Real‐Time Bladder Volume Monitoring for Pediatric Patients Using a Commercially Available Wearable Ultrasound Device
Neurourology and Urodynamics · 2025
DOI: 10.1002/nau.70060 ↗Kurz gefasst
Fachartikel in der Zeitschrift Neurourology and Urodynamics, 2025. Verfasst von Peter Y Cai und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Boston Children's Hospital, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Urinary Bladder, Wearable Electronic Devices. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 6 Literaturangaben. Gefördert von: P.Y.C. is a research fellow supported by National Institute of Health L30 grant (Award #1L30DK134207-01) and T32 training grant (Award # T32DK060442-19)..
Abstract (Originalsprache)
Introduction
We tested whether a commercially available, wearable bladder ultrasound device (DFree Ⓡ , Triple W Japan Inc.) could accurately estimate bladder volume in children with urologic conditions by comparing sensor readings to instilled volume during cystometrography (CMG). Materials and Methods We prospectively enrolled 40 children (6 months to 18 years old). The device was taped to the suprapubic region and volume measurements obtained during CMG. Primary outcome was device accuracy, defined as device indicating ≥ 8 (0−10 scale) when the instilled volume was ≥ 80% of CMG‐determined capacity. Mixed effects linear regression modeled the relationship between sensor indication and instilled volume after adjusting for clinical variables.
Results
Median age was 7.2 years, 20 (50%) were male, 27 (67.5%) were on clean intermittent catheterization, 30 (75%) had spina bifida. The sensor accurately measured capacity in 46.2% of patients with a sensitivity of 0.70, specificity of 0.13, positive predictive value of 0.53, and negative predictive value of 0.22. Device readings were statistically associated with instilled volume; single‐digit (10%) increase in sensor reading was associated with an average increase of 10% (95% CI: 9% to 12%, p < 0.001) in bladder volume.
Conclusion
A commercially available, wearable ultrasound bladder sensor accurately measured bladder capacity in 46.2% of children undergoing UDS. Our findings caution clinicians in recommending these types of direct‐to‐consumer devices. Repeated measures testing is needed to determine if the accurate device indicators observed in a minority of patients herein, are reproducible. Trial Registration Our study did not need to be registered per the International Committee of Medical Journal Editors Uniform Requirements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, Boston Children's Hospital, Boston, Massachusetts, USA.
Förderung
- P.Y.C. is a research fellow supported by National Institute of Health L30 grant (Award #1L30DK134207-01) and T32 training grant (Award # T32DK060442-19).
Zitieren
APA
Cai, P. Y., Logvinenko, T., Omar, B., Oottamasathien, S., Kurtz, M., Estrada, C., Bauer, S., & Nelson, C. P. (2025). Real‐Time Bladder Volume Monitoring for Pediatric Patients Using a Commercially Available Wearable Ultrasound Device. Neurourology and Urodynamics. https://doi.org/10.1002/nau.70060
Vancouver
Cai PY, Logvinenko T, Omar B, Oottamasathien S, Kurtz M, Estrada C, et al. Real‐Time Bladder Volume Monitoring for Pediatric Patients Using a Commercially Available Wearable Ultrasound Device. Neurourology and Urodynamics. 2025. doi:10.1002/nau.70060
BibTeX
@article{Cai2025,
author = {Cai, Peter Y and Logvinenko, Tanya and Omar, Badar and Oottamasathien, Siam and Kurtz, Michael and Estrada, Carlos and Bauer, Stuart and Nelson, Caleb P},
title = {Real‐Time Bladder Volume Monitoring for Pediatric Patients Using a Commercially Available Wearable Ultrasound Device},
journal = {Neurourology and Urodynamics},
year = {2025},
doi = {10.1002/nau.70060}
}