Validation of an automatic method for reconstruction, delineation, and measurement of levator hiatus in clinical practice

Ying Chen, Xin Lin, Man Zhang, Enze Qu, Dongmei Huang, Yongjiang Mao, Zeping Huang, Xinling Zhang iD

Neurourology and Urodynamics · 2023

DOI: 10.1002/nau.25231 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Neurourology and Urodynamics, 2023. Verfasst von Ying Chen und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ultrasound, The Third Affiliated Hospital of Sun Yat-Sen University, Guangzhou, Guangdong Province, China. Hauptthemen laut MEDLINE: Muscle Contraction, Pelvic Floor. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Gefördert von: Natural Science Foundation of Guangdong Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate the concordance between an automatic software program and manual evaluation in reconstructing, delineating, and measuring the levator hiatus (LH) on maximal Valsalva maneuver.

Methods

This was a retrospective study analyzing archived raw ultrasound imaging data of 100 patients underwent transperineal ultrasound (TPUS) examination. Each data were assessed by the automatic Smart Pelvic System software program and manual evaluation. The Dice similarity index (DSI), mean absolute distance (MAD), and Hausdorff distance (HDD) were calculated to quantify delineation accuracy of LH. Agreement between automatic and manual measurement of levator hiatus area was assessed by intraclass correlation coefficient (ICC) and Bland–Altman method.

Results

The satisfaction rate of automatic reconstruction was 94%. Six images were recognized as unsatisfactory reconstructed images for some gas in the rectum and anal canal. Compared with satisfactory reconstructed images, DSI of unsatisfactory reconstructed images was lower, MAD and HDD were larger ( p = 0.001, p = 0.001, p = 0.006, respectively). The ICC was up to 0.987 in 94 satisfactory reconstructed images.

Conclusions

The Smart Pelvic System software program had good performance in reconstruction, delineation, and measurement of LH on maximal Valsalva maneuver in clinical practice, despite misidentification of the border of posterior aspect of LH due to the influence of gas in the rectum.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Muscle ContractionPelvic FloorHumansRetrospective StudiesImaging, Three-DimensionalUltrasonographyValsalva Maneuver

Stichwörter der Autor:innen

automatic delineationautomatic reconstructionlevator hiatustransperineal ultrasound

Autor:innen und Institutionen

  • Department of Ultrasound, The Third Affiliated Hospital of Sun Yat-Sen University, Guangzhou, Guangdong Province, China.

ORCID-Profile: Xinling Zhang

Förderung

  • Natural Science Foundation of Guangdong Province

Zitieren

APA

Chen, Y., Lin, X., Zhang, M., Qu, E., Huang, D., Mao, Y., Huang, Z., & Zhang, X. (2023). Validation of an automatic method for reconstruction, delineation, and measurement of levator hiatus in clinical practice. Neurourology and Urodynamics. https://doi.org/10.1002/nau.25231

Vancouver

Chen Y, Lin X, Zhang M, Qu E, Huang D, Mao Y, et al. Validation of an automatic method for reconstruction, delineation, and measurement of levator hiatus in clinical practice. Neurourology and Urodynamics. 2023. doi:10.1002/nau.25231

BibTeX
@article{Chen2023,
  author = {Chen, Ying and Lin, Xin and Zhang, Man and Qu, Enze and Huang, Dongmei and Mao, Yongjiang and Huang, Zeping and Zhang, Xinling},
  title = {Validation of an automatic method for reconstruction, delineation, and measurement of levator hiatus in clinical practice},
  journal = {Neurourology and Urodynamics},
  year = {2023},
  doi = {10.1002/nau.25231}
}
Kostenlos starten →