Conservative Safety Assessment of a Nitinol Microhook Coaptation Device on Intact Nerve

Jonathan Isaacs, Geetanjali S Bendale, Ananya Gomatam, Hope T Richard iD, Charles Reiter, Kush Savsani, Marc Mazur, Taylor Allen, Noah Clough

Muscle & Nerve · 2026-09 · 74 (3) · 720-727

DOI: 10.1002/mus.70346 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Muscle & Nerve, online erschienen am 14.07.2026. Verfasst von Jonathan Isaacs und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Hand Surgery, Department of Orthopaedic Surgery, Virginia Commonwealth University Health System, Richmond, Virginia, USA. Hauptthemen laut MEDLINE: Alloys, Tibial Nerve. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: BCT, Bat Conservation Trust.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction/Aims Nerve Tape is a novel nerve coaptation device that eliminates the technical challenges of microsuturing. The device incorporates tissue‐engaging Nitinol microhooks within a flexible backing of decellularized porcine small intestine submucosa (SIS). Preclinical nerve injury models have demonstrated device safety and effective promotion of regeneration and recovery. Here, we further assess the impact of the device's components on nerve tissue and function, using a highly conservative model in which the device is applied to intact nerves.

Methods

Nerve Tape, sutures, or SIS were applied to rabbit tibial nerves in a manner simulating transection repairs but leaving the nerves intact. After three months, “repaired” nerves and target gastrocnemius muscles along with contralateral control nerves and muscles were evaluated for perineural scarring, inflammation, and function utilizing gross inspection, morphology measurements, electrophysiology, and histology.

Results

Analysis of endpoint data, including muscle mass/girth, nerve compound action potential (NCAP) amplitude, conduction velocity, axon counts, and g‐ratio revealed no significant differences between experimental or control groups (distal axon count/g‐ratio: Nerve Tape 6546 ± 2300, 0.647 ± 0.04; Suture 7979 ± 666, 0.664 ± 0.033; SIS 7537 ± 2117, 0.689 ± 0.02; control 6158 ± 1958, 0.631 ± 0.02). Gross adhesion thickness at the treatment site was higher in the Suture group versus the Nerve Tape group. Granulomas and neovascularization were most pronounced in the Suture group. Macrophage/lymphocyte/plasma cell presence was associated with SIS material. No additional tissue response was observed in response to Nitinol components.

Discussion

Nerve Tape's SIS and Nitinol microhook design maintained the health of an intact nerve with sequelae comparable to sutures or SIS without microhooks.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlloysTibial NerveAnimalsRabbitsSwineMuscle, SkeletalNerve RegenerationFemaleAction PotentialsNeural ConductionNerve Conduction Studies

Stichwörter der Autor:innen

Nerve TapeSISmicrosuturesnerve coaptation

Substanzen

nitinolAlloys

Autor:innen und Institutionen

  • Division of Hand Surgery, Department of Orthopaedic Surgery, Virginia Commonwealth University Health System, Richmond, Virginia, USA.
  • William & Mary, Williamsburg, Virginia, USA.
  • Department of Pathology, Virginia Commonwealth University Health System, Richmond, Virginia, USA.
  • Department of Plastic Surgery, University Medical Center Utrecht, Utrecht, the Netherlands.

ORCID-Profile: Hope T Richard

Förderung

  • BCT
  • Bat Conservation Trust

Zitieren

APA

Isaacs, J., Bendale, G. S., Gomatam, A., Richard, H. T., Reiter, C., Savsani, K., Mazur, M., Allen, T., & Clough, N. (2026). Conservative Safety Assessment of a Nitinol Microhook Coaptation Device on Intact Nerve. Muscle & Nerve, 74(3), 720-727. https://doi.org/10.1002/mus.70346

Vancouver

Isaacs J, Bendale GS, Gomatam A, Richard HT, Reiter C, Savsani K, et al. Conservative Safety Assessment of a Nitinol Microhook Coaptation Device on Intact Nerve. Muscle & Nerve. 2026;74(3):720-727. doi:10.1002/mus.70346

BibTeX
@article{Isaacs2026,
  author = {Isaacs, Jonathan and Bendale, Geetanjali S and Gomatam, Ananya and Richard, Hope T and Reiter, Charles and Savsani, Kush and Mazur, Marc and Allen, Taylor and Clough, Noah},
  title = {Conservative Safety Assessment of a Nitinol Microhook Coaptation Device on Intact Nerve},
  journal = {Muscle & Nerve},
  year = {2026},
  volume = {74},
  number = {3},
  pages = {720--727},
  doi = {10.1002/mus.70346}
}
Kostenlos starten →