Comparison of Consumer Smartwatch and Research‐Grade Accelerometer‐Derived Step Counts in Amyotrophic Lateral Sclerosis
Muscle & Nerve · 2026-09 · 74 (3) · 580-589
DOI: 10.1002/mus.70323 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Muscle & Nerve, online erschienen am 19.06.2026. Verfasst von Gordon Jewett und 4 weiteren Autor:innen. Beteiligte Einrichtung: Data Intelligence for Health Lab, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada. Hauptthemen laut MEDLINE: Amyotrophic Lateral Sclerosis, Accelerometry, Wearable Electronic Devices, Walking. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: ALS Canada, ALS Society of Canada.
Abstract (Originalsprache)
Introduction/Aims Objective, scalable measures of function are needed in amyotrophic lateral sclerosis (ALS). Research‐grade accelerometers are promising but may be difficult to deploy for prolonged remote monitoring, whereas consumer smartwatches offer a practical alternative. We evaluated the feasibility of long‐term smartwatch monitoring in ALS and compared smartwatch‐ and accelerometer‐derived daily step counts.
Methods
In this single‐centre prospective observational study, participants wore a Fitbit Sense (smartwatch) on the wrist longitudinally and an ActiGraph GT9X Link (accelerometer) on the opposite wrist during clinic‐aligned assessment periods. Device comparison was restricted to concurrent valid days (≥ 10 h wear) and included paired comparison of mean daily step counts, Pearson correlation, Bland–Altman analysis, and proportional bias assessment.
Results
Forty participants were enrolled and 39 contributed data across 8093 smartwatch and 915 accelerometer days. Step count comparison included 503 concurrent valid days from 34 participants. Mean daily step counts were 3480 (SD 4365) for smartwatch and 4066 (SD 4797) for accelerometer ( p < 0.0001), with strong correlation ( r = 0.91, p < 0.0001). Mean bias was −586 steps for smartwatch, with 95% limits of agreement from −4193 to 3021. Among participants with ≥ 6 months of concurrent data, devices showed similar decline over time (interaction p = 0.76).
Discussion
Long‐term remote monitoring using a smartwatch was feasible in ALS. Smartwatch‐derived step counts were strongly correlated with accelerometer‐derived estimates but were systematically lower and showed wide day‐level limits of agreement. Smartwatch‐derived step counts may be useful for group‐level analyses and longitudinal monitoring, though device‐specific bias should be considered in interpretation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Data Intelligence for Health Lab, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
- Department of Clinical Neurosciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
- Hotchkiss Brain Institute, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
- Faculty of Kinesiology, Human Performance Laboratory, University of Calgary, Calgary, Alberta, Canada.
- Running Injury Clinic, University of Calgary, Calgary, Alberta, Canada.
- Department of Community Health Sciences, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada.
ORCID-Profile: Gordon Jewett, Reed Ferber, Colin B Josephson, Lawrence Korngut, Joon Lee
Förderung
- ALS Canada
- ALS Society of Canada
Zitieren
APA
Jewett, G., Ferber, R., Josephson, C. B., Korngut, L., & Lee, J. (2026). Comparison of Consumer Smartwatch and Research‐Grade Accelerometer‐Derived Step Counts in Amyotrophic Lateral Sclerosis. Muscle & Nerve, 74(3), 580-589. https://doi.org/10.1002/mus.70323
Vancouver
Jewett G, Ferber R, Josephson CB, Korngut L, Lee J. Comparison of Consumer Smartwatch and Research‐Grade Accelerometer‐Derived Step Counts in Amyotrophic Lateral Sclerosis. Muscle & Nerve. 2026;74(3):580-589. doi:10.1002/mus.70323
BibTeX
@article{Jewett2026,
author = {Jewett, Gordon and Ferber, Reed and Josephson, Colin B and Korngut, Lawrence and Lee, Joon},
title = {Comparison of Consumer Smartwatch and Research‐Grade Accelerometer‐Derived Step Counts in Amyotrophic Lateral Sclerosis},
journal = {Muscle & Nerve},
year = {2026},
volume = {74},
number = {3},
pages = {580--589},
doi = {10.1002/mus.70323}
}