EUVIMEDEUVIMED entdecken
Crossref

An Artifact in Concentric Needle EMG Recordings

Sanjeev D Nandedkar iD, Paul E Barkhaus iD

Muscle & Nerve · 2025-02 · 71 (2) · 216-222

DOI: 10.1002/mus.28312 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Muscle & Nerve, online erschienen am 12.12.2024. Verfasst von Sanjeev D Nandedkar und 1 weiteren Autor:innen. Beteiligte Einrichtung: Natus Medical Inc, Hopewell Junction, New York, USA. Hauptthemen laut MEDLINE: Electromyography, Artifacts, Needles, Muscle, Skeletal. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 11 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction/Aims In healthy subjects, we observed high amplitude motor unit potential (MUP) waveforms that resembled the cannula potential (CP) with a positive sharp wave (PSW)‐like waveform. We analyzed the source of this signal, its prevalence, and its effects on the analysis of electromyographic waveforms.

Methods

Three channel recordings were performed to explore the contribution of the needle core and cannula to the MUP. In 7 control subjects with no neuromuscular disease, the EMG interference pattern was quantified for the amplitude asymmetry (AASM) of its positive and negative going peaks.

Results

The large amplitude MUP resembling a CP with a PSW‐type waveform is recorded by the needle core, not the cannula. With a slight change in needle position, its amplitude decreased and the waveform had a normal appearance. Thus, it is an artifact. Such potentials (AASM > 150%) are commonly encountered (frequency 7%) and occur in all muscles and subjects in this study.

Discussion

The artifact is an enigma and appears related to the interaction between the needle and muscle fiber(s) similar to that in PSW signals. Failure to recognize these waveforms may result in a false impression of abnormality. AASM calculations can help confirm this artifact in automated analysis algorithms to preclude false positive neurogenic patterns.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectromyographyArtifactsNeedlesMuscle, SkeletalHumansMaleAdultFemaleMiddle AgedYoung AdultMotor NeuronsAction Potentials

Stichwörter der Autor:innen

amplitudeasymmetrycannula potentialconcentric needleelectromyographyneedle electrode artifact

Autor:innen und Institutionen

  • Natus Medical Inc, Hopewell Junction, New York, USA.
  • Medical College of Wisconsin, Milwaukee, Wisconsin, USA.

ORCID-Profile: Sanjeev D Nandedkar, Paul E Barkhaus

Zitieren

APA

Nandedkar, S. D., & Barkhaus, P. E. (2025). An Artifact in Concentric Needle EMG Recordings. Muscle & Nerve, 71(2), 216-222. https://doi.org/10.1002/mus.28312

Vancouver

Nandedkar SD, Barkhaus PE. An Artifact in Concentric Needle EMG Recordings. Muscle & Nerve. 2025;71(2):216-222. doi:10.1002/mus.28312

BibTeX
@article{Nandedkar2025,
  author = {Nandedkar, Sanjeev D and Barkhaus, Paul E},
  title = {An Artifact in Concentric Needle EMG Recordings},
  journal = {Muscle & Nerve},
  year = {2025},
  volume = {71},
  number = {2},
  pages = {216--222},
  doi = {10.1002/mus.28312}
}
Kostenlos starten →