Combining Cartesian and Non‐Cartesian Acceleration Techniques With SPARKLING for 1 mm Isotropic Whole‐Brain MPRAGE in a Minute
Magnetic Resonance in Medicine · 15.11.2026
DOI: 10.1002/mrm.70657 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 15.11.2026. Verfasst von Chaithya Giliyar Radhakrishna und 4 weiteren Autor:innen. Beteiligte Einrichtung: CEA, NeuroSpin, CNRS, Université Paris-Saclay, Gif-sur-Yvette, France. Die Arbeit nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CCRT High-Performance Computing (HPC), HPC resources of IDRIS, CEA BlueSky project.
Abstract (Originalsprache)
Purpose ‐weighted MPRAGE remains a cornerstone of clinical anatomical imaging, yet its long acquisition times constrain routine use. Established acceleration techniques, namely Parallel Imaging ( PI ) and Compressed Sensing ( CS ), tend to introduce substantial noise and blurring when pushed to high acceleration factors. Although they rely on fundamentally different redundancies, combining them synergistically remains an open challenge.
Methods
The GoLF ‐ SPARKLING framework was extended to jointly exploit two acceleration mechanisms: PI ‐based uniform undersampling in the central k‐space region and variable‐density CS at the periphery, with independent acceleration factors in each zone. To preserve smooth signal evolution throughout the inversion‐recovery period and avoid modulation artifacts, the acquisition trajectory was reordered accordingly. The resulting method was evaluated prospectively in vivo at 1 mm isotropic resolution and benchmarked against Wave‐ CAIPI and Poisson‐disk sampling as respective PI and CS baselines.
Results
The proposed hybrid approach produced sharper, less noisy, and more stable whole‐brain images in approximately 1 min than either acceleration strategy alone. Purely PI ‐based reconstructions were degraded by high g‐factor noise, while exclusively CS ‐based reconstructions exhibited pronounced blurring. Furthermore, this method yielded lower average volumetric errors in downstream automated brain segmentation than state‐of‐the‐art acceleration techniques, demonstrating its clinical utility.
Conclusion
By jointly leveraging PI and CS , GoLF ‐ SPARKLING achieves high acceleration factors that enable sub‐minute, high‐quality anatomical MRI . This translates into greater clinical throughput and more reliable imaging in patients who are challenging to scan.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- CEA, NeuroSpin, CNRS, Université Paris-Saclay, Gif-sur-Yvette, France.
- MIND, Inria, Palaiseau, France.
- Siemens Healthcare SAS, Courbevoie, France.
- Department of Neuroradiology, AP-HP, Henri Mondor University Hospital, Créteil, France.
- EA 4391, Université Paris Est Créteil, Créteil, France.
ORCID-Profile: Chaithya Giliyar Radhakrishna, Aurélien Massire, Blanche Bapst, Alexandre Vignaud, Philippe Ciuciu
Förderung
- CCRT High-Performance Computing (HPC)
- HPC resources of IDRIS
- CEA BlueSky project
Zitieren
APA
Giliyar Radhakrishna, C., Massire, A., Bapst, B., Vignaud, A., & Ciuciu, P. (2026). Combining Cartesian and Non‐Cartesian Acceleration Techniques With SPARKLING for 1 mm Isotropic Whole‐Brain MPRAGE in a Minute. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70657
Vancouver
Giliyar Radhakrishna C, Massire A, Bapst B, Vignaud A, Ciuciu P. Combining Cartesian and Non‐Cartesian Acceleration Techniques With SPARKLING for 1 mm Isotropic Whole‐Brain MPRAGE in a Minute. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70657
BibTeX
@article{GiliyarRadhakrishna2026,
author = {Giliyar Radhakrishna, Chaithya and Massire, Aurélien and Bapst, Blanche and Vignaud, Alexandre and Ciuciu, Philippe},
title = {Combining Cartesian and Non‐Cartesian Acceleration Techniques With SPARKLING for 1 mm Isotropic Whole‐Brain MPRAGE in a Minute},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70657}
}
EUVIMED