Inter‐Shot Motion Correction of Segmented 3D ‐ GRASE ASL Perfusion Imaging With Self‐Navigation and CAIPI
Magnetic Resonance in Medicine · 2026-10 · 96 (4) · 1557-1571
DOI: 10.1002/mrm.70437 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 24.05.2026. Verfasst von Minhao Hu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Oxford Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Brain, Image Processing, Computer-Assisted, Magnetic Resonance Angiography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, NIHR Oxford Biomedical Research Centre, Podium Institute for Sports Medicine and Technology, University of Oxford.
Abstract (Originalsprache)
Purpose
Segmented 3D Gradient and Spin Echo (GRASE) is commonly used in Arterial Spin Labeling (ASL) perfusion imaging. However, it is vulnerable to inter‐shot motion, leading to subtraction errors that cannot be corrected. We developed a retrospective self‐navigated inter‐shot motion correction method for segmented 3D‐GRASE ASL imaging with Controlled Aliasing in Parallel Imaging (CAIPI).
Methods
Multiple shots, each uniformly covering k‐space at distinct sample locations, allow a self‐navigator image to be reconstructed using SENSE for each shot. Rigid‐body motion estimation across the self‐navigators is incorporated into a motion‐compensated forward model for image reconstruction. To support self‐navigation, two CAIPI‐sampled segmented 3D‐GRASE trajectories ensuring full k‐space coverage were explored for point spread function profiles and g‐factor effects. Our approach was evaluated against conventional inter‐volume registration and a previously proposed method, alignedSENSE. Additionally, we compared tag‐control interleaving strategies to assess the impact on motion robustness in five healthy volunteers with instructed head motion.
Results
With instructed moderate head motion, our method effectively reduced motion artifacts and outperformed conventional inter‐volume correction by 12.3% in Pearson correlation coefficient, 4.5% in Structural Similarity Index Measure, and 40.1% in temporal SNR. It matched alignedSENSE performance while requiring only 20% of the computational time. All evaluated CAIPI sampling variants enabled robust motion correction, although tradeoffs were observed between through‐plane blurring and SNR. The tag‐control (T/C) inner loop acquisition yielded better motion robustness across quantitative metrics.
Conclusion
Self‐navigated inter‐shot motion correction using CAIPI sampling and a T/C inner loop for segmented 3D‐GRASE ASL can improve image quality and motion robustness.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Oxford Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
- Department of Diagnostic, Interventional and Pediatric Radiology (DIPR), Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
- Translational Imaging Center (TIC), Swiss Institute for Translational and Entrepreneurial Medicine, Bern, Switzerland.
- Physical Sciences, Sunnybrook Research Institute, Toronto, Ontario, Canada.
- Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada.
- The Podium Institute for Sports Medicine and Technology, Department of Engineering Science, University of Oxford, Oxford, UK.
ORCID-Profile: Minhao Hu, Peter Jezzard, Joseph G Woods, Mark Chiew, Thomas W Okell
Förderung
- Wellcome Trust
- NIHR Oxford Biomedical Research Centre
- Podium Institute for Sports Medicine and Technology, University of Oxford
- Vivensa Foundation
- Royal Society
- Oxford Health Biomedical Research Centre
- Jardine Foundation
- Canada Research Chairs Program
Zitieren
APA
Hu, M., Lange, F. J., Jezzard, P., Woods, J. G., Chiew, M., & Okell, T. W. (2026). Inter‐Shot Motion Correction of Segmented 3D ‐ GRASE ASL Perfusion Imaging With Self‐Navigation and CAIPI. Magnetic Resonance in Medicine, 96(4), 1557-1571. https://doi.org/10.1002/mrm.70437
Vancouver
Hu M, Lange FJ, Jezzard P, Woods JG, Chiew M, Okell TW. Inter‐Shot Motion Correction of Segmented 3D ‐ GRASE ASL Perfusion Imaging With Self‐Navigation and CAIPI. Magnetic Resonance in Medicine. 2026;96(4):1557-1571. doi:10.1002/mrm.70437
BibTeX
@article{Hu2026,
author = {Hu, Minhao and Lange, Frederik J and Jezzard, Peter and Woods, Joseph G and Chiew, Mark and Okell, Thomas W},
title = {Inter‐Shot Motion Correction of Segmented 3D ‐ GRASE ASL Perfusion Imaging With Self‐Navigation and CAIPI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {4},
pages = {1557--1571},
doi = {10.1002/mrm.70437}
}
EUVIMED