Bicomponent Mapping of Cortical Bone Using a New Interleaved UTE Imaging Sequence
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70245 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Soo Hyun Shin und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California San Diego, La Jolla, California, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Tibia, Cortical Bone, Image Interpretation, Computer-Assisted. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIAMS NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
Ultrashort echo time (UTE) MRI enables direct imaging of cortical bone and quantification of its water compartments via bicomponent T 2 * modeling. However, conventional approaches require multiple separate dual‐echo scans due to limitations in gradient power. This approach is prone to inter‐scan inconsistencies such as motion and signal drift, which degrade fitting accuracy. This study proposes an interleaved dual‐echo acquisition sequence that acquires multiple echo time (TE) images in a single scan to improve bicomponent T 2 * quantification in cortical bone.
Methods
The proposed UTE sequence utilizes interleaved dual‐echo acquisitions with flexible TE spacings. This sequence was tested on five healthy subjects' tibial midshafts and compared to conventional separate dual‐echo scans with and without image registration. Bicomponent T 2 * modeling was performed, and fitting accuracy was evaluated using normalized‐root‐mean‐squared error (NRMSE). Three subjects were scanned three times to evaluate the scan repeatability.
Results
The interleaved method significantly reduced NRMSE (3.2% ± 2.3% vs. 6.2% ± 3.1%, p = 0.0231) and yielded lower and more stable T 2 * (T 2s *; 0.50 ± 0.10 ms vs. 0.76 ± 0.13 ms, p < 0.0001) and fraction (F s ; 78.2 ± 5.1 vs. 84.2% ± 7.1%, p = 0.0006) of short T 2 components compared to separate scans without registration. Image registration had a minimal improvement on mapping results for separate scans. Parameter maps from the interleaved scans confirmed more homogeneous distributions of T 2s * and F s with lower fitting errors. The much lower coefficients of variance of the interleaved scans demonstrated improved repeatability compared with separate scans.
Conclusion
The proposed interleaved UTE dual‐echo sequence improves the robustness of bicomponent T 2 * mapping of the cortical bone by reducing inter‐scan inconsistencies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University of California San Diego, La Jolla, California, USA.
- Shu Chien-Gene Lay Department of Bioengineering, University of California San Diego, La Jolla, California, USA.
ORCID-Profile: Soo Hyun Shin, Jiyo S Athertya, James Lo
Förderung
- NIAMS NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Shin, S. H., Athertya, J. S., Suprana, A., Lo, J., Wang, J., Moazamian, D., Chapelin, F., & Ma, Y. (2026). Bicomponent Mapping of Cortical Bone Using a New Interleaved UTE Imaging Sequence. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70245
Vancouver
Shin SH, Athertya JS, Suprana A, Lo J, Wang J, Moazamian D, et al. Bicomponent Mapping of Cortical Bone Using a New Interleaved UTE Imaging Sequence. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70245
BibTeX
@article{Shin2026,
author = {Shin, Soo Hyun and Athertya, Jiyo S and Suprana, Arya and Lo, James and Wang, Jiaji and Moazamian, Dina and Chapelin, Fanny and Ma, Yajun},
title = {Bicomponent Mapping of Cortical Bone Using a New Interleaved UTE Imaging Sequence},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70245}
}
EUVIMED