EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Simulation of dB /dt‐Over‐Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems

Simulation der dB/dt-über-E-Feld-Kardialmagnetstimulations-Sicherheitsverhältnisse in 75 Körpermodellen und 18 Gradientsystemen

Valerie Klein iD, Jonathan Edmonson, Mathias Davids iD, Natalie G Ferris, Matthias Gebhardt, Dominik Rattenbacher, Johan S van den Brink, Michael Steckner iD, Lawrence L Wald, Bastien Guérin

Magnetic Resonance in Medicine · 2026-03 · 95 (3) · 1753-1761

DOI: 10.1002/mrm.70130 ↗

Auf einen Blick

Simulationen in 75 Körpermodellen und 13 Gradientsystemen zeigten dB/dt-über-E-Feld-Raten von 12–53 (T/s)*(V/m)−1. Die Worst-Case-Raten lagen in beiden Studien bei 12–13 (T/s)*(V/m)−1.

Fragestellung

Welche dB/dt-über-E-Feld-Verhältnisse ergeben sich in realistischen Körpermodellen und Gradientsystemen für den Herzschutz bei MRT?

Methode

Zwei unabhängige Simulationsstudien in 75 realistischen Körpermodellen und 13 kommerziellen Gradientsystemen; E-Feld-Karten wurden am 99. Perzentil (E99) gethresholdet.

Ergebnisse

E99-Thresholding beseitigte Staircasing-Artefakte. Studie #1: 13–53 (T/s)*(V/m)−1, Studie #2: 12–35 (T/s)*(V/m)−1. In beiden Studien lagen die Worst-Case-Raten bei 12–13 (T/s)*(V/m)−1.

Schlussfolgerung der Autor:innen

Die Simulationen decken einen breiten Bereich realistischer klinischer Szenarien ab. Eine Erhöhung des zulässigen dB/dt über den aktuellen CS-Grenzwert der IEC 60601-2-33 könnte möglich sein.

Kernaussagen

  • E99-Thresholding reduzierte Staircasing-Artefakte
  • Worst-Case-Raten in beiden Studien: 12–13 (T/s)*(V/m)−1
  • Mögliche Erhöhung des zulässigen dB/dt wird diskutiert

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 07.10.2025. Verfasst von Valerie Klein und 9 weiteren Autor:innen. Beteiligte Einrichtung: A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, Massachusetts, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Heart. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The IEC 60601‐2‐33 standard provides consensus‐based safety provisions for MRI equipment. Protection of patients against cardiac stimulation (CS) is based on limiting the maximum E‐field induced by MRI gradient coils. In practice, this is achieved by imposing a conservative dB/dt threshold on any gradient waveform. The dB/dt‐over‐E‐field conversion ratio currently used in IEC 60601‐2‐33 was derived in a homogeneous ellipsoid exposed to a uniform B‐field and is 10 (T/s)*(V/m) −1 . This limit is becoming increasingly restrictive in high performance clinical systems. We therefore evaluate dB/dt‐over‐E‐field ratios in realistic body models and coils using state‐of‐the‐art electromagnetic simulations.

Methods

We performed two independent simulation studies in a total of 75 realistic body models and 13 commercial gradient systems and derived dB/dt‐over‐E‐field ratios in the heart. We thresholded the E‐field maps to mitigate the impact of staircasing artifacts in boundary voxels between the myocardium and the lungs.

Results

Thresholding the E‐field maps at the 99th percentile E‐field value (E99) eliminates staircasing artifacts in both simulation studies. Study #1 predicts a larger range of dB/dt‐over‐E99 ratios (13–53 (T/s)*(V/m) −1 ) than study #2 (12–35 (T/s)*(V/m) −1 ). Despite differences in EM solvers, body models, coils, mesh resolution, and post‐processing, both studies find similar worst‐case ratios of dB/dt‐over‐E99 of 12–13 (T/s)*(V/m) −1 .

Conclusion

Our simulations of dB/dt‐over‐E‐field ratios for cardiac safety in MRI cover a large range of realistic clinical scenarios. An increase of the allowable dB/dt beyond the current CS limit in IEC 60601‐2‐33 may be feasible.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingHeartHumansComputer SimulationElectromagnetic FieldsPhantoms, Imaging

Autor:innen und Institutionen

  • A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Harvard Medical School, Boston, Massachusetts, USA.
  • Medtronic Cardiac Rhythm Management, Minneapolis, Minnesota, USA.
  • Harvard-MIT Division of Health Sciences and Technology, Cambridge, Massachusetts, USA.
  • Siemens Healthineers, Erlangen, Germany.
  • Philips Healthcare, Best, the Netherlands.

ORCID-Profile: Valerie Klein, Mathias Davids, Michael Steckner

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Klein, V., Edmonson, J., Davids, M., Ferris, N. G., Gebhardt, M., Rattenbacher, D., van den Brink, J. S., Steckner, M., Wald, L. L., & Guérin, B. (2026). Simulation of dB /dt‐Over‐Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems. Magnetic Resonance in Medicine, 95(3), 1753-1761. https://doi.org/10.1002/mrm.70130

Vancouver

Klein V, Edmonson J, Davids M, Ferris NG, Gebhardt M, Rattenbacher D, et al. Simulation of dB /dt‐Over‐Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems. Magnetic Resonance in Medicine. 2026;95(3):1753-1761. doi:10.1002/mrm.70130

BibTeX
@article{Klein2026,
  author = {Klein, Valerie and Edmonson, Jonathan and Davids, Mathias and Ferris, Natalie G and Gebhardt, Matthias and Rattenbacher, Dominik and van den Brink, Johan S and Steckner, Michael and Wald, Lawrence L and Guérin, Bastien},
  title = {Simulation of dB /dt‐Over‐Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {95},
  number = {3},
  pages = {1753--1761},
  doi = {10.1002/mrm.70130}
}
Kostenlos starten →