Cardiac diffusion kurtosis imaging in the human heart in vivo using 300 mT/m gradients
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30626 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Maryam Afzali und 10 weiteren Autor:innen. Beteiligte Einrichtung: Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK. Hauptthemen laut MEDLINE: Heart, Diffusion Tensor Imaging, Image Processing, Computer-Assisted, Diffusion Magnetic Resonance Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swedish Cancer Society, Wellcome Trust, EPSRC.
Abstract (Originalsprache)
Purpose
Diffusion tensor imaging (DTI) is commonly used in cardiac diffusion magnetic resonance imaging (dMRI). However, the tissue's microstructure (cells, membranes, etc.) restricts the movement of the water molecules, making the spin displacements deviate from Gaussian behavior. This effect may be observed with diffusion kurtosis imaging (DKI) using sufficiently high b‐values (), which are presently outside the realm of routine cardiac dMRI due to the limited gradient strength of clinical scanners. The Connectom scanner with enables high b‐values at echo times (TE) similar to DTI on standard clinical scanners, therefore facilitating cardiac DKI in humans.
Methods
Cardiac‐gated, second‐order motion‐compensated dMRI was performed with in 10 healthy volunteers on a 3T MRI scanner with . The signal was fitted to a cumulant expansion up to and including the kurtosis term, and diffusion metrics such as fractional anisotropy (FA), mean diffusivity (MD), mean kurtosis (MK), axial kurtosis (AK), and radial kurtosis (RK) were calculated.
Results
We demonstrate deviation of the signal from monoexponential decay for b‐values (). Radial kurtosis () was observed slightly larger than axial kurtosis (), and the difference is statistically significant (, ).
Conclusion
This work demonstrates the feasibility of quantifying kurtosis effect in the human heart in vivo (at an echo time shorter than typical TEs reported for cardiac DTI), using high‐performance gradient systems (which are 4–8 times stronger than on standard clinical scanners). Our work lays the foundation for exploring new biomarkers in cardiac dMRI beyond DTI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
- Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, UK.
- Siemens Healthcare Ltd., Camberly, UK.
- Siemens Healthcare GmbH, Erlangen, Germany.
- Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden.
ORCID-Profile: Maryam Afzali, Sam Coveney, Lars Mueller, Fabrizio Fasano, C John Evans, Irvin Teh, Erica Dall'Armellina, Filip Szczepankiewicz, Derek K Jones, Jürgen E Schneider
Förderung
- Swedish Cancer Society
- Wellcome Trust
- EPSRC
- Wolfson Foundation
- British Heart Foundation
- Cancerfonden
- Engineering and Physical Sciences Research Council
Zitieren
APA
Afzali, M., Coveney, S., Mueller, L., Jones, S., Fasano, F., Evans, C. J., Teh, I., Dall'Armellina, E., Szczepankiewicz, F., Jones, D. K., & Schneider, J. E. (2025). Cardiac diffusion kurtosis imaging in the human heart in vivo using 300 mT/m gradients. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30626
Vancouver
Afzali M, Coveney S, Mueller L, Jones S, Fasano F, Evans CJ, et al. Cardiac diffusion kurtosis imaging in the human heart in vivo using 300 mT/m gradients. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30626
BibTeX
@article{Afzali2025,
author = {Afzali, Maryam and Coveney, Sam and Mueller, Lars and Jones, Sarah and Fasano, Fabrizio and Evans, C John and Teh, Irvin and Dall'Armellina, Erica and Szczepankiewicz, Filip and Jones, Derek K and Schneider, Jürgen E},
title = {Cardiac diffusion kurtosis imaging in the human heart in vivo using 300 mT/m gradients},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30626}
}
EUVIMED