Reassessment of peripheral nerve stimulation thresholds for the Impulse model‐optimized asymmetric head gradient coil
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30523 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von David A Feinberg und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neuroscience and Helen Wills Neuroscience Institute, Brain Imaging Center, University of California, Berkeley, California, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Peripheral Nerves. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIMH NIH HHS, BRAIN Initiative.
Abstract (Originalsprache)
Purpose
Peripheral nerve stimulation (PNS) remains a physiologic limitation to boosting spatiotemporal resolution with more powerful gradients. We investigate discrepancies in previous measurements and model predictions from PNS experienced by volunteers scanned with the investigational “Impulse” gradient coil on the NexGen 7T scanner.
Methods
Twenty‐nine volunteers (18 males, mean ± standard deviation age 52.2 ± 17.1 years) underwent PNS characterizations in the scanner. The process was repeated after the subject positions were moved by 2 and 4 cm toward the feet, away from isocenter. These new data were compared with prior experimental data acquired at the factory (32 volunteers, 16 males, mean ± standard deviation age 58.3 ± 13.5 years) and to modeling results initially used to guide the gradient winding pattern.
Results
The PNS threshold for the x‐axis (left–right) was significantly below the threshold level predicted by the model used to optimize the wiring pattern and thresholds measured in the factory, whereas there was closer agreement for the y‐axis (anterior–posterior) and z‐axis (superior–inferior). The x‐axis threshold increased as the subject was moved in the Z‐direction toward the foot end of the magnet, at the expense of gradient nonlinearity distortions. Sensitivity of the threshold for the x‐axis was measured as 20 mT/m per centimeter Z‐offset.
Conclusion
The PNS threshold of the x‐axis measured in the scanner was much lower than predicted by the optimization model and as measured at the factory. Our measurements verified that PNS thresholds of asymmetric head gradient coils were sensitive to head position, subject variability, and age. The discrepancy of the PNS prediction model remains to be elucidated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neuroscience and Helen Wills Neuroscience Institute, Brain Imaging Center, University of California, Berkeley, California, USA.
- Advanced MRI Technologies, Sebastopol, California, USA.
- Siemens Medical Solutions, Malvern, Pennsylvania, USA.
- Siemens Healthcare, Erlangen, Germany.
- A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
- Harvard Medical School, Boston, Massachusetts, USA.
ORCID-Profile: Samantha J Ma, Alexander J S Beckett, Mathias Davids, Nicolas Boulant
Förderung
- NIBIB NIH HHS
- NIMH NIH HHS
- BRAIN Initiative
- NINDS NIH HHS
- Horizon 2020 Framework Programme
- National Institute of Mental Health
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Feinberg, D. A., Ma, S. J., Walker, E., Beckett, A. J. S., Rattenbacher, D., Rummert, E., Dietz, P., Davids, M., & Boulant, N. (2025). Reassessment of peripheral nerve stimulation thresholds for the Impulse model‐optimized asymmetric head gradient coil. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30523
Vancouver
Feinberg DA, Ma SJ, Walker E, Beckett AJS, Rattenbacher D, Rummert E, et al. Reassessment of peripheral nerve stimulation thresholds for the Impulse model‐optimized asymmetric head gradient coil. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30523
BibTeX
@article{Feinberg2025,
author = {Feinberg, David A and Ma, Samantha J and Walker, Erica and Beckett, Alexander J S and Rattenbacher, Dominik and Rummert, Elmar and Dietz, Peter and Davids, Mathias and Boulant, Nicolas},
title = {Reassessment of peripheral nerve stimulation thresholds for the Impulse model‐optimized asymmetric head gradient coil},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30523}
}