First in‐vivo magic angle directional imaging using dedicated low‐field MRI
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30332 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Mihailo Ristic und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Faculty of Engineering, Imperial College London, London, UK. Hauptthemen laut MEDLINE: Anterior Cruciate Ligament, Magnetic Resonance Imaging. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, The Wellcome Trust.
Abstract (Originalsprache)
Purpose
To report the first in‐vivo results from exploiting the magic angle effect, using a dedicated low‐field MRI scanner that can be rotated about two axes. The magic angle directional imaging (MADI) method is used to depict collagen microstructures with 3D collagen tractography of knee ligaments and the meniscus.
Methods
A novel low‐field MRI system was developed, based on a transverse field open magnet, where the magnet can be rotated about two orthogonal axes. Sets of volume scans at various orientations were obtained in healthy volunteers. The experiments focused on the anterior cruciate ligament (ACL) and the meniscus of the knee. The images were co‐registered, anatomical regions of interest (ROIs) were selected and the collagen fiber orientations in each voxel were estimated from the observed image intensity variations. The 3D collagen tractography was superimposed on conventional volume images.
Results
The MADI method was successfully employed for the first time producing in‐vivo results comparable to those previously reported for excised animal specimens using conventional MRI. Tractography plots were generated for the ACL and the menisci. These results are consistent with the known microstructure of collagen fibers in these tissues.
Conclusion
Images obtained using low‐field MRI with 1 mm 3 resolution were of sufficient quality for the MADI method, which was shown to produce high quality in‐vivo information of collagen microstructures. This was achieved using a cost effective and sustainable low‐field magnet making the technique potentially accessible and scalable, potentially changing the way we image injuries or disease in joints.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Faculty of Engineering, Imperial College London, London, UK.
- Department of Surgery and Cancer, Faculty of Medicine, Imperial College London, London, UK.
- Department of Radiology, Imperial College NHS Healthcare Trust, London, UK.
ORCID-Profile: Mihailo Ristic, Karyn E Chappell
Förderung
- Wellcome Trust
- The Wellcome Trust
Zitieren
APA
Ristic, M., Chappell, K. E., Lanz, H., McGinley, J., Gupte, C., & Amiras, D. (2025). First in‐vivo magic angle directional imaging using dedicated low‐field MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30332
Vancouver
Ristic M, Chappell KE, Lanz H, McGinley J, Gupte C, Amiras D. First in‐vivo magic angle directional imaging using dedicated low‐field MRI. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30332
BibTeX
@article{Ristic2025,
author = {Ristic, Mihailo and Chappell, Karyn E and Lanz, Harry and McGinley, John and Gupte, Chinmay and Amiras, Dimitris},
title = {First in‐vivo magic angle directional imaging using dedicated low‐field MRI},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30332}
}
EUVIMED