EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Hand‐held electron spin resonance scanner for subcutaneous oximetry using OxyChip

Nir Almog, Oleg Zgadzai, Periannan Kuppusamy iD, Yehonatan Zur, Limor Baruch, Marcelle Machluf, Aharon Blank iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30066 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Nir Almog und 6 weiteren Autor:innen. Beteiligte Einrichtung: Schulich Faculty of Chemistry, Technion-Israel Institute of Technology, Haifa, Israel. Hauptthemen laut MEDLINE: Phantoms, Imaging, Oximetry, Oxygen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NCI NIH HHS, NIBIB NIH HHS, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Electron spin resonance (ESR) is used to measure oxygen partial pressure (pO 2 ) in biological media with many clinical applications. Traditional clinical ESR involves large magnets that encompass the subject of measurement. However, certain applications might benefit from a scanner operating within local static magnetic fields. Our group recently developed such a compact scanner for transcutaneous (surface) pO 2 measurements of skin tissue. Here we extend this capability to subsurface (subcutaneous) pO 2 measurements and verify it using an artificial tissue emulating (ATE) phantom.

Methods

We introduce a new scanner, tailored for subcutaneous measurements up to 2 mm beneath the skin's surface. This scanner captures pulsed ESR signals from embedded approximate 1‐mm oxygen‐sensing solid paramagnetic implant, OxyChip. The scanner features a static magnetic field source, producing a uniform region outside its surface, and a compact microwave resonator, for exciting and receiving ESR signals.

Results

ESR readings derived from an OxyChip, positioned approximately 1.5 mm from the scanner's surface, embedded in ATE phantom, exhibited a linear relation of 1/T 2 versus pO 2 for pO 2 levels at 0, 7.6, 30, and 160 mmHg, with relative reading accuracy of about 10%.

Conclusion

The compact ESR scanner can report pO 2 data in ATE phantom from an external position relative to the scanner. Implementing this scanner in preclinical and clinical applications for subcutaneous pO 2 measurements is a feasible next phase for this development. This innovative design also has the potential to operate in conjunction with artificial skin graft for wound healing, combining therapeutic and pO 2 diagnostic features.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingOximetryOxygenElectron Spin Resonance SpectroscopyHumansEquipment DesignSkin

Stichwörter der Autor:innen

OxyChipcompact magnetoximetrypulsed ESRtissue pO2

Substanzen

Oxygen

Autor:innen und Institutionen

  • Schulich Faculty of Chemistry, Technion-Israel Institute of Technology, Haifa, Israel.
  • Department of Radiology, Radiation Oncology, and Medicine, Geisel School of Medicine, Dartmouth College, Lebanon, New Hampshire, USA.
  • Faculty of Biotechnology and Food Engineering, Technion-Israel Institute of Technology, Haifa, Israel.

ORCID-Profile: Periannan Kuppusamy, Aharon Blank

Förderung

  • NCI NIH HHS
  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Almog, N., Zgadzai, O., Kuppusamy, P., Zur, Y., Baruch, L., Machluf, M., & Blank, A. (2024). Hand‐held electron spin resonance scanner for subcutaneous oximetry using OxyChip. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30066

Vancouver

Almog N, Zgadzai O, Kuppusamy P, Zur Y, Baruch L, Machluf M, et al. Hand‐held electron spin resonance scanner for subcutaneous oximetry using OxyChip. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30066

BibTeX
@article{Almog2024,
  author = {Almog, Nir and Zgadzai, Oleg and Kuppusamy, Periannan and Zur, Yehonatan and Baruch, Limor and Machluf, Marcelle and Blank, Aharon},
  title = {Hand‐held electron spin resonance scanner for subcutaneous oximetry using OxyChip},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30066}
}