Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 29.09.2023. Verfasst von Luca Zilberti und 9 weiteren Autor:innen. Beteiligte Einrichtung: Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy. Hauptthemen laut MEDLINE: Electricity, Magnetic Resonance Imaging. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Metrology Programme for Innovation and Research, European Partnership on Metrology.
Abstract (Originalsprache)
Purpose
To investigate whether the risk of peripheral nerve stimulation increases in the presence of bulky metallic prostheses implanted in a patient's body.
Methods
A computational tool was used to calculate the electric field (E‐field) induced in a realistic human model due to the action of gradient fields. The calculations were performed both on the original version of the anatomical model and on a version modified through “virtual surgery” to incorporate knee, hip, and shoulder prostheses. Five exam positions within a body gradient coil and one position using a head gradient coil were simulated, subjecting the human model to the readout gradient from an EPI sequence. The induced E‐field in models with and without prostheses was compared, focusing on the nerves and all other tissues (both including and excluding the bones from the analysis).
Results
In the nerves, the most pronounced increase in the E‐field (+24%) was observed around the knee implant during an abdominal MRI ( Y axis readout). When extending the analysis to encompass all tissues (excluding bones), the greatest amplification (+360%) occurred around the knee implant during pelvic MRI ( Z axis readout). Notable increases in E‐field peaks were also identified around the shoulder and hip implants in multiple scenarios.
Conclusion
Based on the presented results, further investigations aimed at quantifying the threshold of nerve stimulation in the presence of bulky implants are desirable.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy.
- Department of Industrial Engineering, Università degli Studi di Padova, Padova, Italy.
- IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy.
- Research Center for Medical Artificial Intelligence, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
ORCID-Profile: Luca Zilberti, Alessandro Arduino, Riccardo Torchio, Umberto Zanovello, Oriano Bottauscio
Förderung
- European Metrology Programme for Innovation and Research
- European Partnership on Metrology
Zitieren
APA
Zilberti, L., Arduino, A., Torchio, R., Zanovello, U., Baruffaldi, F., Sanchez-Lopez, H., Bettini, P., Alotto, P., Chiampi, M., & Bottauscio, O. (2024). Orthopedic implants affect the electric field induced by switching gradients in MRI. Magnetic Resonance in Medicine, 91(1), 398-412. https://doi.org/10.1002/mrm.29861
Vancouver
Zilberti L, Arduino A, Torchio R, Zanovello U, Baruffaldi F, Sanchez-Lopez H, et al. Orthopedic implants affect the electric field induced by switching gradients in MRI. Magnetic Resonance in Medicine. 2024;91(1):398-412. doi:10.1002/mrm.29861
BibTeX
@article{Zilberti2024,
author = {Zilberti, Luca and Arduino, Alessandro and Torchio, Riccardo and Zanovello, Umberto and Baruffaldi, Fabio and Sanchez-Lopez, Hector and Bettini, Paolo and Alotto, Piergiorgio and Chiampi, Mario and Bottauscio, Oriano},
title = {Orthopedic implants affect the electric field induced by switching gradients in MRI},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {91},
number = {1},
pages = {398--412},
doi = {10.1002/mrm.29861}
}
EUVIMED