Validation of liver quantitative susceptibility mapping across imaging parameters at 1.5 T and 3.0 T using SQUID susceptometry as reference
Magnetic Resonance in Medicine · 2023-04 · 89 (4) · 1418-1428
DOI: 10.1002/mrm.29529 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 21.11.2022. Verfasst von Ruiyang Zhao und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of Wisconsin, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Iron, Iron Overload. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NIBIB NIH HHS, NIDDK NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To validate QSM‐based biomagnetic liver susceptometry (BLS) to measure liver iron overload at 1.5 T and 3.0 T using superconducting quantum interference devices (SQUID)‐based BLS as reference.
Methods
Subjects with known or suspected iron overload were recruited for QSM‐BLS at 1.5 T and 3.0 T using eight different protocols. SQUID‐BLS was also obtained in each subject to provide susceptibility reference. A recent QSM method based on data‐adaptive regularization was used to obtain susceptibility and maps. Measurements of susceptibility and were obtained in the right liver lobe. Linear mixed‐effects analysis was used to estimate the contribution of specific acquisition parameters to QSM‐BLS. Linear regression and Bland–Altman analyses were used to assess the relationship between QSM‐BLS and SQUID‐BLS/.
Results
Susceptibility maps showed high subjective quality for each acquisition protocol across different iron levels. High linear correlation was observed between QSM‐BLS and SQUID‐BLS at 1.5 T ( r 2 range, [0.82, 0.84]) and 3.0 T ( r 2 range, [0.77, 0.85]) across different acquisition protocols. QSM‐BLS and were highly correlated at both field strengths ( r 2 range at 1.5 T, [0.94, 0.99]; 3.0 T, [0.93, 0.99]). High correlation ( r 2 = 0.99) between 1.5 T and 3.0 T QSM‐BLS, with narrow reproducibility coefficients (range, [0.13, 0.21] ppm) were observed for each protocol.
Conclusion
This work evaluated the feasibility and performance of liver QSM‐BLS across iron levels and acquisition protocols at 1.5 T and 3.0 T. High correlation and reproducibility were observed between QSM‐BLS and SQUID‐BLS across protocols and field strengths. In summary, QSM‐BLS may enable reliable and reproducible quantification of liver iron concentration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Biomedical Engineering, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Medicine, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Emergency Medicine, University of Wisconsin, Madison, Wisconsin, USA.
- Department of Radiology, Stanford University, Stanford, California, USA.
ORCID-Profile: Ruiyang Zhao, Julia Velikina, Scott B Reeder, Diego Hernando
Förderung
- NIBIB NIH HHS
- NIDDK NIH HHS
- National Institutes of Health
- GE Healthcare
- University of Wisconsin-Madison
- Stanford University
- Wisconsin Alumni Research Foundation
Zitieren
APA
Zhao, R., Velikina, J., Reeder, S. B., Vasanawala, S., Jeng, M., & Hernando, D. (2023). Validation of liver quantitative susceptibility mapping across imaging parameters at 1.5 T and 3.0 T using SQUID susceptometry as reference. Magnetic Resonance in Medicine, 89(4), 1418-1428. https://doi.org/10.1002/mrm.29529
Vancouver
Zhao R, Velikina J, Reeder SB, Vasanawala S, Jeng M, Hernando D. Validation of liver quantitative susceptibility mapping across imaging parameters at 1.5 T and 3.0 T using SQUID susceptometry as reference. Magnetic Resonance in Medicine. 2023;89(4):1418-1428. doi:10.1002/mrm.29529
BibTeX
@article{Zhao2023,
author = {Zhao, Ruiyang and Velikina, Julia and Reeder, Scott B and Vasanawala, Shreyas and Jeng, Michael and Hernando, Diego},
title = {Validation of liver quantitative susceptibility mapping across imaging parameters at 1.5 T and 3.0 T using SQUID susceptometry as reference},
journal = {Magnetic Resonance in Medicine},
year = {2023},
volume = {89},
number = {4},
pages = {1418--1428},
doi = {10.1002/mrm.29529}
}