Double‐bowtie filter design for pediatric spectral CT imaging: A fan beam simulation study

Doppel-Bowtie-Filterdesign für pädiatrische Spektral-CT-Bildgebung: eine Fan-Beam-Simulationsstudie

Yinglin Ge iD, Olivia F Sandvold iD, Roland Proksa, Amy E Perkins iD, Thomas Koehler iD, Kevin M Brown iD, Yannan Jin, Heiner Daerr, Ravindra M Manjeshwar, Peter B Noël iD

Medical Physics · 2026-11 · 53 (11)

DOI: 10.1002/mp.70752 ↗

Auf einen Blick

In einer Simulation verbesserte ein Double-Bowtie-Filter die spektrale Bildqualität in der pädiatrischen CT. Die Teflon+Gd-Konfiguration erzielte die besten Ergebnisse.

Fragestellung

Lässt sich mit einem Double-Bowtie-Filter aus K-Kanten-Material und Teflon die spektrale SNR und die spektrale Trennung bei pädiatrischer CT verbessern?

Methode

Simulationsstudie mit Rapid-kVp-Switching-CT bei 70/110 kVp; pädiatrische Phantome in 3 Größen; Vergleich von 5 Filtermaterialien mit Teflon-Bowtie als Referenz.

Ergebnisse

Teflon+Gd zeigte die höchste Leistung. Die mittlere AUMC über alle Fan-Winkel stieg im Mittel um 53,6 % und beim größten Phantom um bis zu 56 %. VMI-SNR nahm um etwa 46 % bei 40 keV und 42 % bei 70 keV zu.

Schlussfolgerung der Autor:innen

Das Double-Bowtie-Konzept verbessert die spektrale Leistung deutlich. Teflon+Gd wird als herstellbare, pädiatrisch optimierte Lösung für kVp-Switching und andere Spektral-CT-Architekturen beschrieben.

Kernaussagen

  • Teflon+Gd erzielte die beste Leistung unter 5 getesteten Materialien
  • AUMC im Mittel um 53,6 % erhöht
  • VMI-SNR um etwa 46 % bei 40 keV und 42 % bei 70 keV gestiegen

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 05.11.2026. Verfasst von Yinglin Ge und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Filtration. Die Arbeit nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, Philips Healthcare, NIBIB NIH HHS.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Despite concerns about radiation dose in pediatric patients, computed tomography (CT) remains a first‐line imaging modality for children. Spectral CT provides additional diagnostic information that benefits pediatric imaging. However, most existing spectral CT systems are primarily optimized for adult patients.

Purpose

To develop and evaluate a novel double bowtie filter integrating a K‐edge material layer with a conventional Teflon filter for CT. The proposed design aims to enhance spectral signal‐to‐noise ratio (SNR) and spectral separation within the projection domain while maintaining radiation dose levels suitable for pediatric imaging.

Methods

A simulation framework was developed and used to model a rapid kVp‐switching CT system operating at 70/110 kVp with realistic tube power and geometric constraints. Pediatric phantoms of three sizes (100–200 mm anterior–posterior width) were used to evaluate performance. Five accessible and safe filter materials—gadolinium (Gd), holmium (Ho), erbium (Er), silver (Ag), and tin (Sn)—were tested in combination with a Teflon bowtie. System performance was quantified using virtual monoenergetic image (VMI) SNR at 40 and 70 keV, and the area under the monoenergetic SNR curve (AUMC) as a comprehensive spectral image quality metric. Dose consistency was ensured using a traditional Teflon bowtie as the reference.

Results

The Teflon + Gd configuration achieved the highest performance, improving mean AUMC over all fan angles by 53.6% on average and up to 56% for the largest phantom. VMI SNR increased by approximately 46% at 40 keV and 42% at 70 keV.

Conclusions

The double‐bowtie concept substantially enhances spectral performance. The Teflon + Gd design provides a manufacturable, pediatric‐optimized solution adaptable to kVp‐switching and other spectral CT architectures, offering improved diagnostic quality at low dose levels.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedFiltrationSignal-To-Noise RatioPhantoms, ImagingHumansEquipment DesignChildRadiation Dosage

Stichwörter der Autor:innen

K‐edge materialbowtie filterfan beam simulationkVp‐switching CTpediatric imagingspectral CT

Autor:innen und Institutionen

  • Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Philips Healthcare, Cleveland, Ohio, USA.
  • Philips Innovative Technologies, Hamburg, Germany.

ORCID-Profile: Yinglin Ge, Olivia F Sandvold, Amy E Perkins, Thomas Koehler, Kevin M Brown, Peter B Noël

Förderung

  • NIH HHS
  • Philips Healthcare
  • NIBIB NIH HHS
  • National Institutes of Health

Zitieren

APA

Ge, Y., Sandvold, O. F., Proksa, R., Perkins, A. E., Koehler, T., Brown, K. M., Jin, Y., Daerr, H., Manjeshwar, R. M., & Noël, P. B. (2026). Double‐bowtie filter design for pediatric spectral CT imaging: A fan beam simulation study. Medical Physics, 53(11). https://doi.org/10.1002/mp.70752

Vancouver

Ge Y, Sandvold OF, Proksa R, Perkins AE, Koehler T, Brown KM, et al. Double‐bowtie filter design for pediatric spectral CT imaging: A fan beam simulation study. Medical Physics. 2026;53(11). doi:10.1002/mp.70752

BibTeX
@article{Ge2026,
  author = {Ge, Yinglin and Sandvold, Olivia F and Proksa, Roland and Perkins, Amy E and Koehler, Thomas and Brown, Kevin M and Jin, Yannan and Daerr, Heiner and Manjeshwar, Ravindra M and Noël, Peter B},
  title = {Double‐bowtie filter design for pediatric spectral CT imaging: A fan beam simulation study},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {11},
  doi = {10.1002/mp.70752}
}
Kostenlos starten →