EUVIMEDEUVIMED entdecken
Open Access

Simulation study of wedge‐shaped light guide for compensating depth‐of‐interaction induced timing uncertainty in Cherenkov radiator

Hyungu Kang, Hyeong Seok Shim, Minseok Yi, Juhwan Kim, Yoonchan Jeong, Jae Sung Lee

Medical Physics · 2026

DOI: 10.1002/mp.70655 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Hyungu Kang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, Seoul National University, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Monte Carlo Method, Light, Positron-Emission Tomography. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Interdisciplinary Research Initiatives Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Coincidence timing resolution (CTR) is a key performance factor in PET systems. Cherenkov radiation, due to its ultrafast emission on the femtosecond scale, offers potential for achieving improved timing resolution compared to traditional scintillation‐based detectors. However, the inherently low photon yield in pure Cherenkov radiators makes depth‐of‐interaction (DOI) estimation and its associated timing uncertainty a significant challenge, especially when trying to achieve sub‐30 ps CTR.

Purpose

This study aims to reduce DOI‐induced timing uncertainty in Cherenkov‐based PET detectors by proposing a novel wedge‐shaped light guide design. The goal is to improve the CTR performance by compensating for optical path differences resulting from varying interaction depths.

Methods

A Monte Carlo simulation framework was developed to model gamma‐ray interactions within a Cherenkov radiator, incorporating electron scattering and Cherenkov photon generation. Two detector geometries were compared: one with a conventional planar light guide and another with the proposed wedge‐shaped light guide. The wedge design was tailored to delay early‐arriving photons and synchronize photon transit times, thereby reducing time spread. The simulation tracked photon paths through the wedge and evaluated performance using metrics such as spatial resolution, timing spread, and sensitivity.

Results

Simulations demonstrated that the wedge‐shaped light guide significantly reduced DOI‐induced timing uncertainty—by 26.3% in a 3 mm‐thick Cherenkov radiator—compared to the conventional configuration. The wedge geometry also enabled improved spatial resolution in multi‐layer detector configurations. However, a trade‐off with reduced sensitivity was observed. To address this, strategies such as utilizing discarded photons via refractive index manipulation and cascade detector structures were proposed.

Conclusions

The wedge‐shaped light guide effectively compensates for DOI‐related path length differences and enhances CTR in Cherenkov‐based PET detectors. This approach opens avenues for more accurate event localization without increasing photon yield, and could serve as a building block for advanced PET systems requiring ultrafast timing performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodLightPositron-Emission TomographyTime FactorsUncertaintyPhotonsComputer Simulation

Stichwörter der Autor:innen

Cherenkov radiationMonte Carlo simulationPET detectordepth‐of‐interactiontiming resolutionwedge‐shaped light guide

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, Seoul National University, Seoul, Republic of Korea.
  • Interdisciplinary Program in Bioengineering, College of Engineering, Seoul National University, Seoul, Republic of Korea.
  • Integrated Major in Innovative Medical Science, Seoul National University, Seoul, Republic of Korea.
  • Department of Nuclear Medicine, Seoul National University College of Medicine, Seoul, Republic of Korea.

Förderung

  • Interdisciplinary Research Initiatives Program

Zitieren

APA

Kang, H., Shim, H. S., Yi, M., Kim, J., Jeong, Y., & Lee, J. S. (2026). Simulation study of wedge‐shaped light guide for compensating depth‐of‐interaction induced timing uncertainty in Cherenkov radiator. Medical Physics. https://doi.org/10.1002/mp.70655

Vancouver

Kang H, Shim HS, Yi M, Kim J, Jeong Y, Lee JS. Simulation study of wedge‐shaped light guide for compensating depth‐of‐interaction induced timing uncertainty in Cherenkov radiator. Medical Physics. 2026. doi:10.1002/mp.70655

BibTeX
@article{Kang2026,
  author = {Kang, Hyungu and Shim, Hyeong Seok and Yi, Minseok and Kim, Juhwan and Jeong, Yoonchan and Lee, Jae Sung},
  title = {Simulation study of wedge‐shaped light guide for compensating depth‐of‐interaction induced timing uncertainty in Cherenkov radiator},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70655}
}
Kostenlos starten →