Theoretical modeling and numerical simulation of a light‐field detector for x‐ray differential phase contrast imaging

Yuhang Tan, Jiongtao Zhu, Xin Zhang, Hairong Zheng, Dong Liang, Yongshuai Ge

Medical Physics · 2026-08 · 53 (8)

DOI: 10.1002/mp.70601 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 29.07.2026. Verfasst von Yuhang Tan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences…. Hauptthemen laut MEDLINE: Light, Models, Theoretical, Computer Simulation. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Shenzhen Science and Technology Program, National Key Research and Development Program of China, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Most recently, a novel color‐encoding light‐field detector based on perovskite material has been demonstrated for x‐ray differential phase contrast (DPC) imaging. However, it lacks performance evaluation and comparison between such a novel light‐field detector and a widely used grating interferometer before applying it into medical imaging field.

Purpose

This study aims at evaluating and comparing the DPC imaging performance of the light‐field detector and the grating interferometer.

Methods

In this work, a mathematical signal model was developed for the color‐encoding perovskite light‐field detector designed for x‐ray DPC imaging. Specifically, the beam refraction angle was modeled based on the changes in the emitted color within the standard CIE1931 color space. Upon it, noise responses were derived. Numerical simulations were conducted to compare the sensitivity and signal variance of the perovskite light‐field detector with those of a generic grating interferometer system. Additionally, impacts of the polychromatic x‐ray beam and the Compton scattering were also investigated.

Results

It was found that the sensitivity of the perovskite light‐field detector may surpass that of the grating interferometer if its color depth exceeds 30‐bit (three 10‐bit color channels). For perovskite light‐field detectors, the signal noise increases as the detected refraction signal increases. In addition, polychromatic x‐ray beam adds bias to the retrieved DPC signal, but has slight impact on the signal noise. Compton scattered x‐ray photons could dramatically change the retrieved DPC signal.

Conclusion

This study demonstrates that the novel perovskite light‐field detector may not be suitable for medical grade DPC imaging due to the inevitable Compton scatters. However, it might be a promising alternative to the grating interferometer in developing state‐of‐the‐art scientific DPC imaging instruments for small sample (), in which the impact of Compton scatters could be ignored.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LightModels, TheoreticalComputer SimulationOxidesTitaniumX-RaysColorCalcium Compounds

Stichwörter der Autor:innen

grating interferometerperovskite light‐field detectorx‐ray differential phase contrast imaging

Substanzen

OxidesperovskiteTitaniumCalcium Compounds

Autor:innen und Institutionen

  • Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
  • Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
  • State Key Laboratory of Biomedical Imaging Science and System, Shenzhen, Guangdong, China.
  • National Innovation Center for Advanced Medical Devices, Shenzhen, Guangdong, China.
  • Research Center for Medical Artificial Intelligence, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.

Förderung

  • Shenzhen Science and Technology Program
  • National Key Research and Development Program of China
  • National Natural Science Foundation of China

Zitieren

APA

Tan, Y., Zhu, J., Zhang, X., Zheng, H., Liang, D., & Ge, Y. (2026). Theoretical modeling and numerical simulation of a light‐field detector for x‐ray differential phase contrast imaging. Medical Physics, 53(8). https://doi.org/10.1002/mp.70601

Vancouver

Tan Y, Zhu J, Zhang X, Zheng H, Liang D, Ge Y. Theoretical modeling and numerical simulation of a light‐field detector for x‐ray differential phase contrast imaging. Medical Physics. 2026;53(8). doi:10.1002/mp.70601

BibTeX
@article{Tan2026,
  author = {Tan, Yuhang and Zhu, Jiongtao and Zhang, Xin and Zheng, Hairong and Liang, Dong and Ge, Yongshuai},
  title = {Theoretical modeling and numerical simulation of a light‐field detector for x‐ray differential phase contrast imaging},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {8},
  doi = {10.1002/mp.70601}
}
Kostenlos starten →