Development and validation of an automated, accurate in‐house treatment planning system for pencil‐beam scanning carbon ion radiotherapy
Medical Physics · 2026
DOI: 10.1002/mp.70456 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Medical Physics, 2026. Verfasst von Sac Lee und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic …. Hauptthemen laut MEDLINE: Heavy Ion Radiotherapy, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 44 Literaturangaben. Gefördert von: Korea Medical Device Development Fund, National Research Foundation of Korea (NRF).
Abstract (Originalsprache)
Background
Carbon ion radiotherapy (CIRT) offers superior physical and biological advantages over photon and proton radiotherapy (PRT). However, due to a need for accurate modeling of the relative biological effectiveness (RBE), currently available treatment planning systems (TPS) for CIRT remain limited. This has constrained clinical and research applications by compelling reliance on a narrow set of systems and restricting flexibility for broader application.
Purpose
This study aimed to develop and validate a new efficient and accurate CIRT TPS for pencil‐beam scanning (PBS)‐based CIRT, tailored for the Heavy‐Ion Therapy Center (HITC) at Yonsei Cancer Center (YCC), incorporating modified microdosimetric kinetic model (mMKM)‐based RBE‐weighted dose calculation and spot weight optimization.
Methods
The proposed TPS was designed to automate the workflow from CT image import to CIRT plan optimization. Key refinements included 3D Siddon's ray tracing for spot trajectory tracking, air‐gap modeling for beam profile adaptation, and 3D stopping power ratio (SPR) calculation. Gaussian beam modeling was performed to calibrate in‐air measurements and Monte Carlo (MC)‐driven 3D dose kernels in water. These procedures produced dose influence matrices (DIMs) and MC‐estimated matrices required for spot weight optimization with mMKM‐based RBE‐weighted dose calculation, which was further accelerated by Adam optimizer. Validation was performed by generating plans for a water phantom (10 Gy‐RBE), a lung case (15 Gy‐RBE), and a prostate case (4.3 Gy‐RBE), using single‐field and multi‐field optimizations (MFO). The resulting plans were recalculated in RayStation (v.2025) and TOPAS MC, and compared using dose–volume histogram (DVH) and gamma passing rate (GPR) at 2%/2 mm.
Results
The proposed TPS successfully automated the CIRT plan optimization within approximately 1 min for ∼1000 spots, which produced uniform RBE‐weighted dose coverage in virtual water phantom, the lung patient, and the prostate patient cases. When the spot weights optimized by the proposed framework were recalculated using a commercial TPS and TOPAS MC, the resulting dose distributions closely matched those of the proposed TPS. Quantitatively, GPRs exceeding 98% were achieved for both physical and RBE‐weighted dose at the 2%/2 mm criterion, except for the prostate single‐field case, which yielded a GPR of 97.37% for the physical dose, relative to TOPAS MC, and a GPR of 96.28% for the RBE‐weighted dose, relative to RayStation. These findings confirmed strong agreement between the proposed TPS, MC simulations, and a commercial TPS.
Conclusion
A new, fully automated in‐house CIRT TPS was developed and validated, demonstrating high dosimetric accuracy and computational efficiency, even comparable to a commercial TPS and MC simulations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic of Korea.
Förderung
- Korea Medical Device Development Fund
- National Research Foundation of Korea (NRF)
Zitieren
APA
Lee, S., Yun, Y., Han, M. C., Kim, C., Hong, C. S., Lee, H., Kim, D. W., Jung, D. M., Kim, J. S., & Kim, H. (2026). Development and validation of an automated, accurate in‐house treatment planning system for pencil‐beam scanning carbon ion radiotherapy. Medical Physics. https://doi.org/10.1002/mp.70456
Vancouver
Lee S, Yun Y, Han MC, Kim C, Hong CS, Lee H, et al. Development and validation of an automated, accurate in‐house treatment planning system for pencil‐beam scanning carbon ion radiotherapy. Medical Physics. 2026. doi:10.1002/mp.70456
BibTeX
@article{Lee2026,
author = {Lee, Sac and Yun, Yongdo and Han, Min Cheol and Kim, Changhwan and Hong, Chae-Seon and Lee, Ho and Kim, Dong Wook and Jung, Dong Min and Kim, Jin Sung and Kim, Hojin},
title = {Development and validation of an automated, accurate in‐house treatment planning system for pencil‐beam scanning carbon ion radiotherapy},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70456}
}
EUVIMED