Feasibility study of surface deformation‐based dosimetry for breast cancer radiotherapy
Medical Physics · 2026
DOI: 10.1002/mp.70276 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Yongjin Deng und 7 weiteren Autor:innen. Beteiligte Einrichtung: The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, Guangdong, China. Hauptthemen laut MEDLINE: Breast Neoplasms. Die Arbeit nennt 35 Literaturangaben. Gefördert von: Medical Scientific Research Foundation of Guangdong Province of China.
Abstract (Originalsprache)
Background
With the increasing complexity of breast cancer radiotherapy techniques, accurate dose delivery has become more critical than ever. Real‐time dosimetry now plays an essential role in ensuring both treatment efficacy and patient safety.
Purpose
To propose a novel dosimetry approach for breast cancer radiotherapy that leverages surface deformation to enhance dose delivery accuracy.
Method
This retrospective study analyzed 29 left breast cancer radiotherapy cases from an existing clinical database. The methodology incorporated nonrigid registration between planning CT and CBCT for dose recalculation, and surface images computed from planning CT and CBCT were analyzed using monocular vision algorithms to simulate clinical monocular camera‐based surface deformation detection. A neural network architecture was subsequently developed to establish the predictive relationship between surface deformation and internal dose redistribution. Model performance was rigorously evaluated through RMSE and 3D gamma pass rate analysis.
Result
The analysis revealed a robust correlation between breast surface deformation features and internal dose redistribution, as evidenced by a first canonical correlation coefficient of 0.940. Quantitative evaluation demonstrated strong predictive performance, with RMSE of 0.005 pixels for the dose deformation field and 0.009 cGy for dose amplitude differences in the test set. Gamma analysis results indicated clinically acceptable accuracy, with all of cases achieving 100% pass rate under 2 mm/2% criteria.
Conclusions
This study presents a novel noninvasive dosimetry approach for breast cancer radiotherapy that enables real‐time monitoring without requiring additional radiation exposure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, Guangdong, China.
Förderung
- Medical Scientific Research Foundation of Guangdong Province of China
Zitieren
APA
Deng, Y., Qiu, M., Xiao, Z., Huang, B., Liu, Y., Guan, Q., Zhong, J., & Bao, Y. (2026). Feasibility study of surface deformation‐based dosimetry for breast cancer radiotherapy. Medical Physics. https://doi.org/10.1002/mp.70276
Vancouver
Deng Y, Qiu M, Xiao Z, Huang B, Liu Y, Guan Q, et al. Feasibility study of surface deformation‐based dosimetry for breast cancer radiotherapy. Medical Physics. 2026. doi:10.1002/mp.70276
BibTeX
@article{Deng2026,
author = {Deng, Yongjin and Qiu, Minmin and Xiao, Zhenhua and Huang, Botian and Liu, Yujian and Guan, Qi and Zhong, Jiajian and Bao, Yong},
title = {Feasibility study of surface deformation‐based dosimetry for breast cancer radiotherapy},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70276}
}
EUVIMED