Comparison of organ volumes and standardized uptake values in [18F]FDG‐PET/CT images using MOOSE and TotalSegmentator to segment CT images
Medical Physics · 2025
DOI: 10.1002/mp.70025 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Medical Physics, 2025. Verfasst von Julie Auriac und 9 weiteren Autor:innen. Beteiligte Einrichtung: LITO U1288, Institut Curie, PSL University, Inserm, Orsay, France. Hauptthemen laut MEDLINE: Fluorodeoxyglucose F18, Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted, Software. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Manual segmentation of organs from PET/CT images is a time‐consuming and highly operator‐dependent task. Open software solutions are now available to automatically segment all major anatomical structures in CT images.
Purpose
We compared the volumes and standardized uptake values (SUV) extracted from [18F]FDG‐PET/CT patient scans for 33 anatomical structures segmented using two deep learning (DL) algorithms to determine if they are interchangeable.
Methods
Baseline [18F]FDG‐PET/CT images were collected retrospectively for 315 women with metastatic breast cancer. A total of 33 anatomical volumes of interest (VOI) were segmented from the whole‐body CT scans using both MOOSE v.3.0.14 and TotalSegmentator v.2.0.5 and copied onto the corresponding PET images. For each VOI, the volume from the CT image and SUVmax, SUVpeak and SUVmean from the PET image were extracted. The resulting values were compared using the relative difference for each feature.
Results
Following DL segmentation, resulting organ volumes differed by less than 10% for 19/33 organs in more than 80% (252/315) of patients. Four organs were segmented with volume differences greater than 20% in 1/5th of patients: bladder (48%, p < 0.0001), portal and splenic veins (34%, p < 0.0001), thyroid (16%, p < 0.0001), adrenal glands (15%, p < 0.0001). SUVmax and SUVpeak were affected by the choice of DL algorithms, with values differing by less than 10% in more than 80% of patients for only 16 and 19 out of 33 organs, respectively. In contrast, SUVmean was less affected with differences of less than 10% in more than 80% of patients for all anatomical structures, except the bladder, lungs and skull.
Conclusions
The two software tools produce similar results in volume estimates for most anatomical structures. SUVmean is less dependent on the segmentation algorithm than SUVmax and SUVpeak and shows excellent reproducibility for all anatomical structures studied except for the bladder, the lungs and the skull.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- LITO U1288, Institut Curie, PSL University, Inserm, Orsay, France.
- Institut Curie, Department of Nuclear Medicine, Paris-St Cloud, France.
- Quantitative Imaging and Medical Physics (QIMP) Team, Medical University of Vienna, Vienna, Austria.
Zitieren
APA
Auriac, J., Nioche, C., Hovhannisyan-Baghdasarian, N., Loisel, C., Seban, R. D., Jehanno, N., Sundar, L. K. S., Beyer, T., Buvat, I., & Orlhac, F. (2025). Comparison of organ volumes and standardized uptake values in [18F]FDG‐PET/CT images using MOOSE and TotalSegmentator to segment CT images. Medical Physics. https://doi.org/10.1002/mp.70025
Vancouver
Auriac J, Nioche C, Hovhannisyan-Baghdasarian N, Loisel C, Seban RD, Jehanno N, et al. Comparison of organ volumes and standardized uptake values in [18F]FDG‐PET/CT images using MOOSE and TotalSegmentator to segment CT images. Medical Physics. 2025. doi:10.1002/mp.70025
BibTeX
@article{Auriac2025,
author = {Auriac, Julie and Nioche, Christophe and Hovhannisyan-Baghdasarian, Narinée and Loisel, Charlotte and Seban, Romain-David and Jehanno, Nina and Sundar, Lalith Kumar Shiyam and Beyer, Thomas and Buvat, Irène and Orlhac, Fanny},
title = {Comparison of organ volumes and standardized uptake values in [18F]FDG‐PET/CT images using MOOSE and TotalSegmentator to segment CT images},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70025}
}
EUVIMED