Comprehensive characterization and validation of a fast‐resolving (1000 Hz) plastic scintillator for ultra‐high dose rate electron dosimetry

Lixiang Guo, Banghao Zhou, Yi-Chun Tsai, Kai Jiang, Viktor Iakovenko, Ken Kang-Hsin Wang

Medical Physics · 2025

DOI: 10.1002/mp.70006 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Medical Physics, 2025. Verfasst von Lixiang Guo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biomedical Imaging and Radiation Technology Laboratory (BIRTLab), Department of Radiation Oncology, University of Texas Southwestern Medical Center, …. Hauptthemen laut MEDLINE: Electrons, Radiometry, Scintillation Counting, Plastics. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cancer Prevention and Research Institute of Texas.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The normal tissue sparing effect of ultra‐high dose rate irradiation (≥40 Gy/s, UHDR), as compared to conventional dose rate (CONV), has attracted significant research interest for FLASH radiotherapy (RT). Accurate, dose rate independent, fast‐responding dosimeters capable of resolving the spatiotemporal characteristics of UHDR beams are urgently needed to facilitate FLASH research and support its clinical translation. Tissue‐equivalent scintillators, with millimeter‐level spatial resolution and millisecond‐level temporal resolution, possess these required characteristics and show strong potential for use in UHDR dosimetry.

Purpose

We investigated the performance of the HYPERSCINT RP‐FLASH scintillator system at up to 1000 Hz sampling frequency ( f s ) for UHDR electron beam dosimetry.

Methods

The scintillator was exposed to CONV and UHDR electron irradiation using a LINAC‐based FLASH platform. Its spectral characteristics were delineated with a four‐component calibration, followed by a signal‐to‐dose calibration using 18 MeV CONV electron beam. The dose linearity and dosimetric accuracy in response to CONV and UHDR irradiation at 1 and 1000 Hz f s were quantified against ion chamber and EBT‐XD film measurements. The response of the scintillator system was investigated as a function of beam energy (6 and 18 MeV), field size (2 × 2 to 25 × 25 cm 2 ), dose per pulse (DPP, 0.8–2.3 Gy/pulse), and pulse repetition frequency (PRF, 30–180 Hz). Relative signal sensitivity was quantified against accumulated dose to account for the scintillator's radiation degradation. Pulse‐resolved dose measurements at 18 MeV UHDR, obtained using the scintillator with 1000 Hz f s for a train of 10 pulses at 180 Hz PRF, were validated with a PMT‐fiber optic scattered radiation detector.

Results

The scintillator system at 1 Hz f s demonstrated high accuracy in dose measurements, remaining within 0.5% of ion chamber measurements over the dose range of 0.1–35 Gy under CONV irradiation. For the UHDR irradiation, the scintillator showed

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsRadiometryScintillation CountingPlasticsRadiation DosageTime Factors

Stichwörter der Autor:innen

FLASH radiotherapydosimetryscintillatorultrahigh dose rate

Substanzen

Plastics

Autor:innen und Institutionen

  • Biomedical Imaging and Radiation Technology Laboratory (BIRTLab), Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
  • Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.

Förderung

  • Cancer Prevention and Research Institute of Texas

Zitieren

APA

Guo, L., Zhou, B., Tsai, Y. C., Jiang, K., Iakovenko, V., & Wang, K. K. (2025). Comprehensive characterization and validation of a fast‐resolving (1000 Hz) plastic scintillator for ultra‐high dose rate electron dosimetry. Medical Physics. https://doi.org/10.1002/mp.70006

Vancouver

Guo L, Zhou B, Tsai YC, Jiang K, Iakovenko V, Wang KK. Comprehensive characterization and validation of a fast‐resolving (1000 Hz) plastic scintillator for ultra‐high dose rate electron dosimetry. Medical Physics. 2025. doi:10.1002/mp.70006

BibTeX
@article{Guo2025,
  author = {Guo, Lixiang and Zhou, Banghao and Tsai, Yi-Chun and Jiang, Kai and Iakovenko, Viktor and Wang, Ken Kang-Hsin},
  title = {Comprehensive characterization and validation of a fast‐resolving (1000 Hz) plastic scintillator for ultra‐high dose rate electron dosimetry},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70006}
}
Kostenlos starten →