Microdosimetry calculations in situ for clinically relevant photon sources and their correlation with the early DNA damage response
Medical Physics · 2025
DOI: 10.1002/mp.17979 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Mirta Dumančić und 6 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada. Hauptthemen laut MEDLINE: Photons, Radiometry, DNA Damage. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 102 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Canada Research Chairs, MEDTEQ+, Institut TransMedTech.
Abstract (Originalsprache)
Background
Radiobiological data suggests variations in relative biological effectiveness (RBE) between clinically used photon‐based sources. A microdosimetric formalism using Monte Carlo (MC) methods can mechanistically describe the photon RBE. Experimentally derived RBE based on DNA double‐strand breaks () has been shown to scale with the microdosimetry quantity dose‐mean lineal energy ().
Purpose
To calculate microdosimetric spectra for clinically relevant photon sources, spanning from soft x‐rays produced by a 50 kVp x‐ray source through various brachytherapy sources up to a 6 MV medical linac. Furthermore, we investigated the correlation between and of different photon sources.
Methods
Photon sources simulated include low‐energy x‐rays (50 kVp), orthovoltage x‐rays (225 kVp), high‐dose‐rate brachytherapy sources ( 75 Se, 192 Ir and 60 Co), and a 6 MV medical linac. Secondary electron spectra at the cellular level were calculated for in vitro cell irradiation setups using Geant4 MC‐based packages, RapidBrachyMCTPS and RapidExternalBeam. The obtained spectra were used in MicroDose, a microdosimetry simulation software, to obtain microdosimetric quantities, including single‐event lineal energy () and specific energy () spectra, and dose‐mean and frequency‐mean quantities (, , , ). Uniform spherical targets (1–14 radius) and realistic HeLa and PC3 cell nucleus models were simulated using cell size data obtained from literature and nuclei size data from confocal microscopy imaging. Radiobiological experiments using foci quantified DNA double‐strand breaks for HeLa and PC3 cells after irradiations with 50 and 225 kVp, 192 Ir, and 6 MV linac, and was determined using 225 kVp as the reference.
Results
The calculated () is within the 3.5–1.2 keV/ range (1.8–0.2 keV/) for 1 simulated target size between the lowest energy 50 kVp x‐ray source and the highest energy 6 MV linac source, respectively. For the HeLa and PC3 cell nuclei models based on microscopy data, () spans from 1.6 to 0.6 keV/ (0.7 to 0.2 keV/). When compared between different target sizes, () ranges from 3.5 to 1.0 (1.8–0.4) keV/ between 1 and 10 radius targets for the 50 kVp x‐ray source. A smaller change is observed for 6 MV linac, ranging from 1.2 to 0.5 keV/ and 0.23 to 0.22 keV/ for and , respectively. For the simulated 75 Se source currently under investigation, the calculated values are 11%–24% higher relative to those of 192 Ir in the range of target sizes between 1 and 14 in radius. for HeLa cells was 1.40.7 for 50 kVp x‐rays, 0.50.2 for 192 Ir, and 0.70.4 for 6 MV linac irradiations. For PC3 cells, was 1.30.6, 0.80.4 and 0.50.3 for 50 kVp, 192 Ir and 6 MV linac, respectively. Measured values are consistent with ratios of the corresponding photon sources for HeLa and PC3 nucleus models.
Conclusions
Microdosimetric spectra strongly depend on the simulated energy of photon sources and target size, with and decreasing by a factor of 2–3 between diagnostic 50 kVp and 6 MV therapeutic x‐rays for target sizes from 1–14 in radius. The early damage indicates this stochastic change in energy density between various photon sources as the yields of foci per nucleus scale with of the source.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada.
- Research Institute of the McGill University Health Centre, Montreal, Quebec, Canada.
- Lady Davis Institute for Medical Research, Jewish General Hospital, Montreal, Quebec, Canada.
- Medical Physics Program, The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, USA.
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, USA.
ORCID-Profile: Mirta Dumančić, Jonathan Kalinowski, Joanna Li, Joseph M DeCunha, Shirin A Enger
Förderung
- Canada Research Chairs
- MEDTEQ+
- Institut TransMedTech
- McGill University Health Centre
- Alpha Tau Medical Ltd.
Zitieren
APA
Dumančić, M., Kalinowski, J., Diaz-Martinez, V. D., Li, J., Behmand, B., DeCunha, J. M., & Enger, S. A. (2025). Microdosimetry calculations in situ for clinically relevant photon sources and their correlation with the early DNA damage response. Medical Physics. https://doi.org/10.1002/mp.17979
Vancouver
Dumančić M, Kalinowski J, Diaz-Martinez VD, Li J, Behmand B, DeCunha JM, et al. Microdosimetry calculations in situ for clinically relevant photon sources and their correlation with the early DNA damage response. Medical Physics. 2025. doi:10.1002/mp.17979
BibTeX
@article{Dumani2025,
author = {Dumančić, Mirta and Kalinowski, Jonathan and Diaz-Martinez, Victor D and Li, Joanna and Behmand, Behnaz and DeCunha, Joseph M and Enger, Shirin A},
title = {Microdosimetry calculations in situ for clinically relevant photon sources and their correlation with the early DNA damage response},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17979}
}
EUVIMED