EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY-NC-ND 4.0

Experimental validation of Geant4 nuclear interaction models in dose calculations of therapeutic carbon ion beams

Yihan Jia, Martina Favaretto, Lisa Hartl, Markus Stock, Dietmar Georg, Loïc Grevillot, Andreas F Resch

Medical Physics · 2025-07 · 52 (7)

DOI: 10.1002/mp.17906 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Medical Physics, online erschienen am 29.05.2025. Verfasst von Yihan Jia und 6 weiteren Autor:innen. Beteiligte Einrichtung: MedAustron Ion Therapy Center, Wiener Neustadt, Austria. Hauptthemen laut MEDLINE: Monte Carlo Method, Heavy Ion Radiotherapy, Radiotherapy Planning, Computer-Assisted, Radiation Dosage. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: European Regional Development Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The choice of nuclear interaction models in Monte Carlo simulations affects the dose calculation accuracy for light ion beam therapy.

Purpose

This study aimed to evaluate the dose calculation accuracy and simulation time of three GATE‐RTiON/ Geant4 physics lists for therapeutic carbon ion beams, assessing their suitability for independent dose calculation in patient‐specific quality assurance (PSQA).

Methods

The normalized beam models for physics lists QGSP_BIC_HP_EMZ, QGSP_INCLXX_HP_EMZ, and Shielding_EMZ were validated against measurements regarding the accuracy of range, spot size and reference dose. Normalized transversal dose profiles () and field size factor (FSF) were compared with measurements. The accuracy of simulated target dose in 103 fields (various energies, field sizes, depths, and dose gradient complexity) of energy‐modulated scanned beams was evaluated at 3181 positions. The median of global dose difference was calculated at different depth ranges.

Results

The three physics lists with validated beam models showed similar accuracy in and FSF in the Bragg peak region and proximal depths, while QGSP_INCLXX_HP agreed most closely for in the fragmentation tail. Accounting for ‐related uncertainty, remained within ±1.1% for QGSP_INCLXX_HP, while exhibiting an overall increasing trend with depth for QGSP_BIC_HP (up to 2.3%) and a decreasing trend for Shielding (down to −4.1%), respectively. By tuning the number‐of‐primaries/monitor unit conversion () as a function of energy, of QGSP_BIC_HP was reduced to within ±1.3%, at the cost of reduced accuracy in the simulated reference dose. The simulation time of Shielding was 1.8 times that of QGSP_BIC_HP and 1.5 times that of QGSP_INCLXX_HP.

Conclusions

QGSP_INCLXX_HP demonstrated high dosimetric accuracy in the target region of energy‐modulated fields. QGSP_BIC_HP and Shielding showed physics model‐related inaccuracies in simulated target dose. Additional tuning improved their target dose calculation accuracy with a trade‐off of reference dose accuracy. The computationally efficient QGSP_INCLXX_HP and QGSP_BIC_HP are viable candidates for dose calculation applications of carbon ion beam therapy, such as in silico PSQA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodHeavy Ion RadiotherapyRadiotherapy Planning, Computer-AssistedRadiation DosageRadiotherapy DosageHumans

Stichwörter der Autor:innen

GATE‐RTiONGeant4carbon ion radiotherapydosimetric accuracy

Autor:innen und Institutionen

  • MedAustron Ion Therapy Center, Wiener Neustadt, Austria.
  • Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria.
  • Faculty of Electrical Engineering and Information Technology, Vienna University of Technology, Vienna, Austria.
  • Division Medical Physics, Karl Landsteiner University of Health Sciences, Krems an der Donau, Austria.

Förderung

  • European Regional Development Fund

Zitieren

APA

Jia, Y., Favaretto, M., Hartl, L., Stock, M., Georg, D., Grevillot, L., & Resch, A. F. (2025). Experimental validation of Geant4 nuclear interaction models in dose calculations of therapeutic carbon ion beams. Medical Physics, 52(7). https://doi.org/10.1002/mp.17906

Vancouver

Jia Y, Favaretto M, Hartl L, Stock M, Georg D, Grevillot L, et al. Experimental validation of Geant4 nuclear interaction models in dose calculations of therapeutic carbon ion beams. Medical Physics. 2025;52(7). doi:10.1002/mp.17906

BibTeX
@article{Jia2025,
  author = {Jia, Yihan and Favaretto, Martina and Hartl, Lisa and Stock, Markus and Georg, Dietmar and Grevillot, Loïc and Resch, Andreas F},
  title = {Experimental validation of Geant4 nuclear interaction models in dose calculations of therapeutic carbon ion beams},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {7},
  doi = {10.1002/mp.17906}
}
Kostenlos starten →