Physical phantom validation of clustering‐initiated factorization in dynamic PET
Medical Physics · 2025
DOI: 10.1002/mp.17902 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Medical Physics, 2025. Verfasst von Valerie Kobzarenko und 7 weiteren Autor:innen. Beteiligte Einrichtung: BioComputing Lab, Department of Electrical Engineering and Computer Science, Florida Institute of Technology, Melbourne, Florida, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Positron-Emission Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, NIBIB NIH HHS, National Science Foundation.
Abstract (Originalsprache)
Background
Dynamic positron emission tomography (PET) enables the quantification of physiological parameters of radiotracers employed in the investigation of neuropsychiatric disorders. We previously introduced a factor analysis‐based algorithm, Cluster‐Initialized Factor Analysis (CIFA), designed to overcome the problem of specifying reference regions. CIFA is capable of automatically extracting distinct radiotracer binding distributions across many modalities based on the differences in tracer dynamics, and thus can distinguish regions of specific‐ and non‐specific binding without requiring prior segmentation.
Purpose
Our goal is to quantitatively validate the ability of CIFA to resolve different dynamic biological processes by comparing the output of the algorithm to an independent benchmark. As an intermediate goal, we aim to create a physical phantom capable of modeling unique aspects of dynamic imaging and to use this phantom as the benchmark in evaluating CIFA.
Methods
CIFA was used to reconstruct 18 F‐flortaucipir dynamic brain PET datasets acquired at Lawrence Berkeley National Lab. The resulting factor curves served as the foundation for creating dynamic input time‐activity curve (TAC) combinations in a physical brain phantom specifically constructed for this purpose. The phantom represented three components: two overlapping tissue types and free radiotracer, constructed with a combination of small hydraulic elements. The physical components were scanned separately to generate a library of images, allowing us to reproduce scans of any duration with prescribed dynamics and realistic partial volume effects. The phantom was designed to produce noisy instances with compartment mixing of dynamic scans with desired activity TACs for free, non‐specifically bound, and specifically bound radiotracers. Ten distinct dynamic simulations with varying levels of TAC similarity were estimated with CIFA.
Results
We directly evaluated CIFA's performance in analyzing each of the 10 dynamic datasets by computing the Pearson correlation coefficient between the estimated outputs and the ground truth tissue TACs and corresponding tissue distributions. For seven out of 10 modeled dynamics, which captured the full spectrum of realistically expected tissue TAC shapes, the curve correlation of the specific binding tissue was above 95%.
Conclusions
This work formulated an innovative process by combining a physical phantom design with PET images for evaluating the application of CIFA in the extraction of dynamic TACs from dynamic PET image data. In most cases the CIFA algorithm accurately reproduced the dynamics of the phantom simulated data.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BioComputing Lab, Department of Electrical Engineering and Computer Science, Florida Institute of Technology, Melbourne, Florida, USA.
- Molecular Biophysics and Integrated Bioimaging, Lawrence Berkeley National Laboratory, Berkeley, California, USA.
- Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA.
- SolvingDynamics, Inc.
Förderung
- NIH HHS
- NIBIB NIH HHS
- National Science Foundation
- NIA NIH HHS
- NHLBI NIH HHS
- National Institutes of Health
Zitieren
APA
Kobzarenko, V., Baker, S. L., Janabi, M., Choong, W. S., Gullberg, G. T., Seo, Y., Boutchko, R., & Mitra, D. (2025). Physical phantom validation of clustering‐initiated factorization in dynamic PET. Medical Physics. https://doi.org/10.1002/mp.17902
Vancouver
Kobzarenko V, Baker SL, Janabi M, Choong WS, Gullberg GT, Seo Y, et al. Physical phantom validation of clustering‐initiated factorization in dynamic PET. Medical Physics. 2025. doi:10.1002/mp.17902
BibTeX
@article{Kobzarenko2025,
author = {Kobzarenko, Valerie and Baker, Suzanne L and Janabi, Mustafa and Choong, Woon-Seng and Gullberg, Grant T and Seo, Youngho and Boutchko, Rostyslav and Mitra, Debasis},
title = {Physical phantom validation of clustering‐initiated factorization in dynamic PET},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17902}
}
EUVIMED