Medical Physics 3.0 approach to optimizing image quality in a breast screening program
Medical Physics · 2025
DOI: 10.1002/mp.17878 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Elizabeth Keavey und 3 weiteren Autor:innen. Beteiligte Einrichtung: BreastCheck, National Cancer Screening Service, Galway, Ireland. Hauptthemen laut MEDLINE: Mammography, Breast Neoplasms, Health Physics, Early Detection of Cancer. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
Background
Sensitivity and specificity of screening mammography is linked to image quality (IQ) and should be optimized and consistent for all women screened. Optimization, typically based on equipment performance metrics from quality control tests, do not correlate directly with clinical performance.
Purpose
The principles of Medical Physics 3.0 have been applied to combine medical physics Quality Control (QC) with screening clinical outcome to optimize the quality in a national breast screening program. Medical Physics 3.0 has been advocated as an initiative to enhance the practice of Medical Physics and leverage the value of medical physics to drive improved patient care.
Methods
A retrospective analysis of the cancer detection rate (CDR), of approximately 1.5 million screening examinations, was conducted within a screening program following the introduction of a new mammography system type. Significative differences in CDR (with nonoverlapping 95% confidence intervals [CI]) and performance metrics (lower [IQ] and mean glandular dose [MGD]) were observed between this new system type and the other system types in the screening program. The CDR for individual mammography system types was correlated with medical physics QC metrics of IQ and dosimetry. Preliminary investigations lead to an increase of the operational dose multiplier from 1.2 to 1.6 and the new setting was applied over all the new systems.
Results
Following optimization, the medical physics metrics revealed a 19% increase in MGD accompanied by a 12% improvement in IQ for the new system type. Meanwhile, a retrospective analysis of CDR showed overlapping 95% CI, indicating convergence between the two system types.
Conclusions
Applying Medical Physics 3.0 principles to aggregated screening and clinical outcome data in support of medical physics quality assurance processes has demonstrated potential to deliver optimization of quality for individual women and for the screened population.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BreastCheck, National Cancer Screening Service, Galway, Ireland.
- BreastCheck, National Cancer Screening Service, Dublin, Ireland.
Zitieren
APA
Keavey, E., Baldelli, P., Power, G., & Phelan, N. (2025). Medical Physics 3.0 approach to optimizing image quality in a breast screening program. Medical Physics. https://doi.org/10.1002/mp.17878
Vancouver
Keavey E, Baldelli P, Power G, Phelan N. Medical Physics 3.0 approach to optimizing image quality in a breast screening program. Medical Physics. 2025. doi:10.1002/mp.17878
BibTeX
@article{Keavey2025,
author = {Keavey, Elizabeth and Baldelli, Paola and Power, Gillian and Phelan, Niall},
title = {Medical Physics 3.0 approach to optimizing image quality in a breast screening program},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17878}
}
EUVIMED