EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Accuracy of a whole‐body single‐photon emission computed tomography with a thallium‐bromide detector: Verification via Monte Carlo simulations

Toshimune Ito, Keitaro Hitomi, Michael Ljungberg, Sousei Kawasaki, Yuka Katayama, Akane Kato, Hirotatsu Tsuchikame, Kentaro Suzuki, Kyosuke Miyazaki, Ritsushi Mogi

Medical Physics · 2025

DOI: 10.1002/mp.17724 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Toshimune Ito und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Radiological, Faculty of Medical Technology, Teikyo University, Tokyo, Japan. Hauptthemen laut MEDLINE: Monte Carlo Method, Tomography, Emission-Computed, Single-Photon, Thallium, Whole Body Imaging. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Single‐photon emission computed tomography (SPECT) devices equipped with cadmium–zinc–telluride (CZT) detectors achieve high contrast resolution because of their enhanced energy resolution. Recently, thallium bromide (TlBr) has gained attention as a detector material because of its high atomic number and density.

Purpose

This study evaluated the clinical applicability of a SPECT system equipped with TlBr detectors using Monte Carlo simulations, focusing on 99mTc and 177Lu imaging.

Methods

This study used the Simulation of Imaging Nuclear Detectors Monte Carlo program to compare the imaging characteristics between a whole‐body SPECT system equipped with TlBr (T‐SPECT) and a system equipped with CZT detectors (C‐SPECT). The simulations were performed using a three‐dimensional brain phantom and a National Electrical Manufacturers Association body phantom to evaluate 99mTc and 177Lu imaging. The simulation parameters were accurately set by comparing them with the actual measurements.

Results

The T‐SPECT system demonstrated improved energy resolution and higher detection efficiency than the C‐SPECT system. In 99mTc imaging, T‐SPECT demonstrated 1.71 times higher photopeak counts and improved contrast resolution. T‐SPECT exhibited a significantly lower impact of hole tailing and higher‐energy resolution (4.50% for T‐SPECT vs. 7.34% for C‐SPECT). Furthermore, T‐SPECT showed higher peak signal‐to‐noise ratio (PSNR) and structural similarity (SSIM) values, indicating better image quality. In 177Lu imaging, T‐SPECT showed 2.76 times higher photopeak counts and improved energy resolution (3.94% for T‐SPECT vs. 5.20% for C‐SPECT). T‐SPECT demonstrated a higher contrast recovery coefficient (CRC) and contrast‐to‐noise ratio (CNR) across all acquisition times, maintaining sufficient counts even with shorter acquisition times. Moreover, T‐SPECT acquired higher low‐frequency values in power spectrum density (PSD), indicating more accurate internal image reproduction.

Conclusions

T‐SPECT offers superior energy resolution and detection efficiency than C‐SPECT. Moreover, T‐SPECT can provide higher contrast resolution and sensitivity in clinical imaging with 99mTc and 177Lu. Furthermore, the Monte Carlo simulations are confirmed to be a valuable guide for the development of T‐SPECT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodTomography, Emission-Computed, Single-PhotonThalliumWhole Body ImagingBromidesPhantoms, ImagingHumansTelluriumLutetiumZincCadmium

Stichwörter der Autor:innen

Monte Carlo modelingdevelopment (new technology and techniques)instrumentationphantoms–digital

Substanzen

ThalliumBromidesTelluriumLutetiumCdZnTeZincCadmium

Autor:innen und Institutionen

  • Department of Medical Radiological, Faculty of Medical Technology, Teikyo University, Tokyo, Japan.
  • Department of Quantum Science and Energy Engineering, Graduate School of Engineering, Tohoku University, Sendai, Japan.
  • Medical Radiation Physics, Lund University, Lund, Sweden.
  • Department of Radiology, Nippon Medical School Hospital, Tokyo, Japan.
  • Department of Radiological Technology, Showa University Hospital, Tokyo, Japan.
  • Department of Radiology, Institute of Science Tokyo Hospital, Tokyo, Japan.

Förderung

  • JSPS KAKENHI

Zitieren

APA

Ito, T., Hitomi, K., Ljungberg, M., Kawasaki, S., Katayama, Y., Kato, A., Tsuchikame, H., Suzuki, K., Miyazaki, K., & Mogi, R. (2025). Accuracy of a whole‐body single‐photon emission computed tomography with a thallium‐bromide detector: Verification via Monte Carlo simulations. Medical Physics. https://doi.org/10.1002/mp.17724

Vancouver

Ito T, Hitomi K, Ljungberg M, Kawasaki S, Katayama Y, Kato A, et al. Accuracy of a whole‐body single‐photon emission computed tomography with a thallium‐bromide detector: Verification via Monte Carlo simulations. Medical Physics. 2025. doi:10.1002/mp.17724

BibTeX
@article{Ito2025,
  author = {Ito, Toshimune and Hitomi, Keitaro and Ljungberg, Michael and Kawasaki, Sousei and Katayama, Yuka and Kato, Akane and Tsuchikame, Hirotatsu and Suzuki, Kentaro and Miyazaki, Kyosuke and Mogi, Ritsushi},
  title = {Accuracy of a whole‐body single‐photon emission computed tomography with a thallium‐bromide detector: Verification via Monte Carlo simulations},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17724}
}
Kostenlos starten →