Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy
Medical Physics · 2025
DOI: 10.1002/mp.17683 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Neville D Gai und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory on Quantitative Medical Imaging, NIBIB/NIH, Bethesda, Maryland, USA. Hauptthemen laut MEDLINE: Povidone, Temperature. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Polyvinylpyrrolidone (PVP) water solutions could be used for cross‐site and cross‐vendor validation of diffusion‐related measurements. However, since water diffusivity varies as a function of temperature, knowing the temperature of the PVP solution at the time of the measurement is fundamental in accomplishing this task.
Purpose
MR spectroscopy (MRS) could provide absolute temperature measurements since the water peak moves relative to any stable peak as temperature changes. In this work, the PVP proton spectrum was investigated to see if any stable peaks would allow for temperature determination. Reproducibility and repeatability for three scanners from three vendors were also assessed.
Methods
A spherical 17 cm container filled with 40% PVP w/w in distilled water was used for the experiments. A Point REsolved Spectroscopy Sequence (PRESS) with water suppression was employed on three 3T scanners from different vendors—GE, Siemens, and Philips. Frequency separation (in ppm) between peaks was measured in a voxel at the location of a fiber optic temperature probe and mapped to probe measured temperature. The center peak of the first methylene proton triplet closest to water peak was selected for analysis in jMRUI due to its ease of identification and echo time shift invariance. Shift in ppm of the central methylene peak proton was mapped against measured temperatures. Repeatability and reproducibility across the three scanners were determined at room temperature using 10 repeated PRESS scans. MRS established ppm shift versus temperature relationship was used to predict temperature in different PVP phantoms which were then compared against fiber optic probe measured temperature values.
Results
Several 1 H peaks were identified on all scans of the PVP phantom. The water peak moved by ∼−0.01 ppm/°C on the three scanners relative to a central methylene peak. The maximum mean absolute temperature difference over a temperature range of 18–35°C between the three scanners was 0.16°C while the minimum was 0.057°C. Repeatability on each scanner was excellent (std range: 0.00–0.14°C) over 10 repeated PRESS scans. Reproducibility across the three scanners was also excellent with mean temperature difference between scanners ranging between 0.1 and 0.4°C. Temperature values from MRS were within prediction bounds on the three scanners for another in‐house prepared 40% PVP phantom (maximum difference
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory on Quantitative Medical Imaging, NIBIB/NIH, Bethesda, Maryland, USA.
- Magnetic Resonance Spectroscopy Core, NIMH/NIH, Bethesda, Maryland, USA.
- Biomedical & Metabolic Imaging Section, NIDDK/NIH, Bethesda, Maryland, USA.
Zitieren
APA
Gai, N. D., Dong, R., van der Veen, J. W., Ouwerkerk, R., & Pierpaoli, C. (2025). Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy. Medical Physics. https://doi.org/10.1002/mp.17683
Vancouver
Gai ND, Dong R, van der Veen JW, Ouwerkerk R, Pierpaoli C. Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy. Medical Physics. 2025. doi:10.1002/mp.17683
BibTeX
@article{Gai2025,
author = {Gai, Neville D and Dong, Ruifeng and van der Veen, Jan Willem and Ouwerkerk, Ronald and Pierpaoli, Carlo},
title = {Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17683}
}
EUVIMED