Validierungsstudie in der Zeitschrift Medical Physics, 2024. Verfasst von Claire Keun Sun Park und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Physics and Biophysics, Department of Radiation Oncology, Brigham and Women's Hospital and Dana-Farber Cancer Institute, Harvard Medical …. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Signal-To-Noise Ratio, Phantoms, Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 55 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council (NSERC) of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Stereotactic MR‐guided Adaptive Radiation Therapy (SMART) dose painting for hypoxia has potential to improve treatment outcomes, but clinical implementation on low‐field MR‐Linac faces substantial challenges due to dramatically lower signal‐to‐noise ratio (SNR) characteristics. While quantitative MRI and T 1 mapping of hypoxia biomarkers show promise, T 1 ‐to‐noise ratio (T 1 NR) optimization at low fields is paramount, particularly for the clinical implementation of oxygen‐enhanced (OE)‐MRI. The 3D Magnetization Prepared (2) Rapid Gradient Echo (MP2RAGE) sequence stands out for its ability to acquire homogeneous T 1 ‐weighted contrast images with simultaneous T 1 mapping.

Purpose

To optimize MP2RAGE for low‐field T 1 mapping; conduct experimental validation in a ground‐truth phantom; establish feasibility and reproducibility of low‐field MP2RAGE acquisition and T 1 mapping in healthy volunteers.

Methods

The MP2RAGE optimization was performed to maximize the contrast‐to‐noise ratio (CNR) of T 1 values in white matter (WM) and gray matter (GM) brain tissues at 0.35T. Low‐field MP2RAGE images were acquired on a 0.35T MR‐Linac (ViewRay MRIdian) using a multi‐channel head coil. Validation of T 1 mapping was performed with a ground‐truth Eurospin phantom, containing inserts of known T 1 values (400–850 ms), with one and two average (1A and 2A) MP2RAGE scans across four acquisition sessions, resulting in eight T 1 maps. Mean (± SD) T 1 relative error, T 1 NR, and intersession coefficient of variation (CV) were determined. Whole‐brain MP2RAGE scans were acquired in 5 healthy volunteers across two sessions (A and B) and T 1 maps were generated. Mean (± SD) T 1 values for WM and GM were determined. Whole‐brain T 1 histogram analysis was performed, and reproducibility was determined with the CV between sessions. Voxel‐by‐voxel T 1 difference maps were generated to evaluate 3D spatial variation.

Results

Low‐field MP2RAGE optimization resulted in parameters: MP2RAGE TR of 3250 ms, inversion times (TI 1 /TI 2 ) of 500/1200 ms, and flip angles ( α 1 / α 2 ) of 7/5°. Eurospin T 1 maps exhibited a mean (± SD) relative error of 3.45% ± 1.30%, T 1 NR of 20.13 ± 5.31, and CV of 2.22% ± 0.67% across all inserts. Whole‐brain MP2RAGE images showed high anatomical quality with clear tissue differentiation, resulting in mean (± SD) T 1 values: 435.36 ± 10.01 ms for WM and 623.29 ± 14.64 ms for GM across subjects, showing excellent concordance with literature. Whole‐brain T 1 histograms showed high intrapatient and intersession reproducibility with characteristic intensity peaks consistent with voxel‐level WM and GM T 1 values. Reproducibility analysis revealed a CV of 0.46% ± 0.31% and 0.35% ± 0.18% for WM and GM, respectively. Voxel‐by‐voxel T 1 difference maps show a normal 3D spatial distribution of noise in WM and GM.

Conclusions

Low‐field MP2RAGE proved effective in generating accurate, reliable, and reproducible T 1 maps with high T 1 NR in phantom studies and in vivo feasibility established in healthy volunteers. While current work is focused on refining the MP2RAGE protocol to enable clinically efficient OE‐MRI, this study establishes a foundation for TOLD T 1 mapping for hypoxia biomarkers. This advancement holds the potential to facilitate a paradigm shift toward MR‐guided biological adaptation and dose painting by leveraging 3D hypoxic spatial distributions and improving outcomes in conventionally challenging‐to‐treat cancers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingSignal-To-Noise RatioPhantoms, ImagingHumansBiomarkersImage Processing, Computer-AssistedRadiation DosageHypoxiaMaleAdultReproducibility of ResultsBrainFemale

Stichwörter der Autor:innen

MR‐LinacT1 mappingadaptive radiation therapylow‐field MRIquantitative imaging

Substanzen

Biomarkers

Autor:innen und Institutionen

  • Division of Physics and Biophysics, Department of Radiation Oncology, Brigham and Women's Hospital and Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts, USA.
  • Harvard-MIT Health Sciences and Technology, Harvard Medical School, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.

ORCID-Profile: Claire Keun Sun Park, Atchar Sudhyadhom

Förderung

  • Natural Sciences and Engineering Research Council (NSERC) of Canada

Zitieren

APA

Park, C. K. S., Warner, N. S., Kaza, E., & Sudhyadhom, A. (2024). Optimization and validation of low‐field MP2RAGE T 1 mapping on 0.35T MR‐Linac: Toward adaptive dose painting with hypoxia biomarkers. Medical Physics. https://doi.org/10.1002/mp.17353

Vancouver

Park CKS, Warner NS, Kaza E, Sudhyadhom A. Optimization and validation of low‐field MP2RAGE T 1 mapping on 0.35T MR‐Linac: Toward adaptive dose painting with hypoxia biomarkers. Medical Physics. 2024. doi:10.1002/mp.17353

BibTeX
@article{Park2024,
  author = {Park, Claire Keun Sun and Warner, Noah Stanley and Kaza, Evangelia and Sudhyadhom, Atchar},
  title = {Optimization and validation of low‐field MP2RAGE T 1 mapping on 0.35T MR‐Linac: Toward adaptive dose painting with hypoxia biomarkers},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17353}
}
Kostenlos starten →