Performance evaluation of iterative PET reconstruction with resolution recovery incorporating Gallium‐68 positron range correction

Prodromos Gavriilidis, Michel Koole, Anouk Marinus, Floris P Jansen, Timothy W Deller, Felix M Mottaghy, Roel Wierts

Medical Physics · 2024-09 · 51 (9) · 5927-5942

DOI: 10.1002/mp.17259 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Medical Physics, online erschienen am 18.06.2024. Verfasst von Prodromos Gavriilidis und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands. Hauptthemen laut MEDLINE: Gallium Radioisotopes, Image Processing, Computer-Assisted, Phantoms, Imaging, Positron-Emission Tomography. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The distance traveled by the positron before annihilation with an electron, the so‐called positron range, negatively effects the positron emission tomography (PET) image quality for radionuclides emitting high‐energy positrons such as Gallium‐68 ( 68 Ga).

Purpose

In this study, the effect of a tissue‐independent positron range correction for Gallium‐68 ( 68 Ga‐PRC) was investigated based on phantom measurements. The effect of the 68 Ga‐PRC was also explored in four patients.

Methods

The positron range distribution profile of 68 Ga in water was generated via Monte Carlo simulation. That profile was mapped to a spatially invariant 3D convolution kernel which was incorporated in the OSEM and Q.Clear reconstruction algorithms to perform the 68 Ga‐PRC. In addition, each reconstruction method included point spread function (PSF) modeling and time‐of‐flight information. For both Fluorine‐18 ( 18 F) and 68 Ga, the NEMA IQ phantom was filled with a sphere‐to‐background ratio of 10:1 and scanned with the GE Discovery MI 5R PET/CT system. Standard non‐positron range correction (PRC) reconstructions were performed for both radionuclides, while also PRC reconstructions were performed for 68 Ga. Reconstructions parameters (OSEM: number of updates, Q.Clear: beta value) were adapted to achieve similar noise levels between the corresponding reconstructions. The effect of 68 Ga‐PRC was assessed for both OSEM and Q.Clear reconstructions and compared to non‐PRC reconstructions for 68 Ga and 18 F in terms of image contrast, noise, recovery coefficient (RC), and spatial resolution. For the clinical validation, 68 Ga‐labeled prostate‐specific membrane antigen ( 68 Ga‐PSMA) and 68 Ga‐DOTATOC PET scans were included of two patients each. For each PET scan, patients were injected with 1.5 MBq/kg of 68 Ga‐PSMA or 68 Ga‐DOTATOC and the contrast‐to‐noise ratio (CNR) was calculated and compared to the non‐PRC reconstructions.

Results

For OSEM reconstructions, including the 68 Ga‐PRC improved the RC by 9.4% (3.7%–19.3%) and spatial resolution by 21.7% (4.6 mm vs. 3.6 mm) for similar noise levels. For Q.Clear reconstructions, 68 Ga‐PRC modeling improved the RC by 6.7% (2.8%–10.5%) and spatial resolution by 15.3% (5.9 mm vs. 5.0 mm) while obtaining similar noise levels. In the patient data, the use of 68 Ga‐PRC enhanced the CNR by 13.2%.

Conclusions

Including 68 Ga‐PRC in the PET reconstruction enhanced the image quality of 68 Ga PET data compared to the standard non‐PRC reconstructions for similar noise levels. Limited patient results also supported this improvement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gallium RadioisotopesImage Processing, Computer-AssistedPhantoms, ImagingPositron-Emission TomographyHumansMonte Carlo MethodMalePositron Emission Tomography Computed TomographyAlgorithms

Stichwörter der Autor:innen

Gallium‐68OSEMQ.Clearpositron emission tomographypositron rangepositron range correction

Substanzen

Gallium RadioisotopesGallium-68

Autor:innen und Institutionen

  • Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
  • School for Oncology and Reproduction (GROW), Maastricht University, Maastricht, The Netherlands.
  • Nuclear Medicine and Molecular Imaging, Department of Imaging & Pathology, KU Leuven, Leuven, Belgium.
  • Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands.
  • Molecular Imaging, GE HealthCare, Waukesha, Wisconsin, USA.
  • National Security Sciences Directorate, Oak Ridge National Laboratory, Oak Ridge, Tennessee, USA.

Zitieren

APA

Gavriilidis, P., Koole, M., Marinus, A., Jansen, F. P., Deller, T. W., Mottaghy, F. M., & Wierts, R. (2024). Performance evaluation of iterative PET reconstruction with resolution recovery incorporating Gallium‐68 positron range correction. Medical Physics, 51(9), 5927-5942. https://doi.org/10.1002/mp.17259

Vancouver

Gavriilidis P, Koole M, Marinus A, Jansen FP, Deller TW, Mottaghy FM, et al. Performance evaluation of iterative PET reconstruction with resolution recovery incorporating Gallium‐68 positron range correction. Medical Physics. 2024;51(9):5927-5942. doi:10.1002/mp.17259

BibTeX
@article{Gavriilidis2024,
  author = {Gavriilidis, Prodromos and Koole, Michel and Marinus, Anouk and Jansen, Floris P and Deller, Timothy W and Mottaghy, Felix M and Wierts, Roel},
  title = {Performance evaluation of iterative PET reconstruction with resolution recovery incorporating Gallium‐68 positron range correction},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {9},
  pages = {5927--5942},
  doi = {10.1002/mp.17259}
}
Kostenlos starten →