Deep learning‐based low‐dose CT simulator for non‐linear reconstruction methods

Sjoerd A M Tunissen, Nikita Moriakov, Mikhail Mikerov, Ewoud J Smit, Ioannis Sechopoulos, Jonas Teuwen

Medical Physics · 2024

DOI: 10.1002/mp.17232 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Sjoerd A M Tunissen und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging, Radboudumc, Nijmegen, The Netherlands. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Image Processing, Computer-Assisted, Radiation Dosage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Health Holland, Canon Medical Systems.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Computer algorithms that simulate lower‐doses computed tomography (CT) images from clinical‐dose images are widely available. However, most operate in the projection domain and assume access to the reconstruction method. Access to commercial reconstruction methods may often not be available in medical research, making image‐domain noise simulation methods useful. However, the introduction of non‐linear reconstruction methods, such as iterative and deep learning‐based reconstruction, makes noise insertion in the image domain intractable, as it is not possible to determine the noise textures analytically.

Purpose

To develop a deep learning‐based image‐domain method to generate low‐dose CT images from clinical‐dose CT (CDCT) images for non‐linear reconstruction methods.

Methods

We propose a fully image domain‐based method, utilizing a series of three convolutional neural networks (CNNs), which, respectively, denoise CDCT images, predict the standard deviation map of the low‐dose image, and generate the noise power spectra (NPS) of local patches throughout the low‐dose image. All three models have U‐net‐based architectures and are partly or fully three‐dimensional. As a use case for this study and with no loss of generality, we use paired low‐dose and clinical‐dose brain CT scans. A dataset of paired scans was retrospectively obtained. All images were acquired with a wide‐area detector clinical system and reconstructed using its standard clinical iterative algorithm. Each pair was registered using rigid registration to correct for motion between acquisitions. The data was randomly partitioned into training ( samples), validation ( samples), and test ( samples) sets. The performance of each of these three CNNs was validated separately. For the denoising CNN, the local standard deviation decrease, and bias were determined. For the standard deviation map CNN, the real and estimated standard deviations were compared locally. Finally, for the NPS CNN, the NPS of the synthetic and real low‐dose noise were compared inside and outside the skull. Two proof‐of‐concept denoising studies were performed to determine if the performance of a CNN‐ or a gradient‐based denoising filter on the synthetic low‐dose data versus real data differed.

Results

The denoising network had a median decrease in noise in the cerebrospinal fluid by a factor of and introduced a median bias of HU. The network for standard deviation map estimation had a median error of HU. The noise power spectrum estimation network was able to capture the anisotropic and shift‐variant nature of the noise structure by showing good agreement between the synthetic and real low‐dose noise and their corresponding power spectra. The two proof of concept denoising studies showed only minimal difference in standard deviation improvement ratio between the synthetic and real low‐dose CT images with the median difference between the two being 0.0 and +0.05 for the CNN‐ and gradient‐based filter, respectively.

Conclusion

The proposed method demonstrated good performance in generating synthetic low‐dose brain CT scans without access to the projection data or to the reconstruction method. This method can generate multiple low‐dose image realizations from one clinical‐dose image, so it is useful for validation, optimization, and repeatability studies of image‐processing algorithms.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningTomography, X-Ray ComputedImage Processing, Computer-AssistedRadiation DosageHumansNonlinear DynamicsBrainSignal-To-Noise Ratio

Stichwörter der Autor:innen

deep learninglow‐dose CTsimulation

Autor:innen und Institutionen

  • Department of Medical Imaging, Radboudumc, Nijmegen, The Netherlands.
  • AI for Oncology Lab, Netherlands Cancer Institute, Amsterdam, The Netherlands.
  • Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands.
  • Technical Medicine Centre, University of Twente, Enschede, The Netherlands.
  • Depatment of Radiology, Memorial Sloan Kettering Cancer Center, New York, USA.

Förderung

  • Health Holland
  • Canon Medical Systems

Zitieren

APA

Tunissen, S. A. M., Moriakov, N., Mikerov, M., Smit, E. J., Sechopoulos, I., & Teuwen, J. (2024). Deep learning‐based low‐dose CT simulator for non‐linear reconstruction methods. Medical Physics. https://doi.org/10.1002/mp.17232

Vancouver

Tunissen SAM, Moriakov N, Mikerov M, Smit EJ, Sechopoulos I, Teuwen J. Deep learning‐based low‐dose CT simulator for non‐linear reconstruction methods. Medical Physics. 2024. doi:10.1002/mp.17232

BibTeX
@article{Tunissen2024,
  author = {Tunissen, Sjoerd A M and Moriakov, Nikita and Mikerov, Mikhail and Smit, Ewoud J and Sechopoulos, Ioannis and Teuwen, Jonas},
  title = {Deep learning‐based low‐dose CT simulator for non‐linear reconstruction methods},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17232}
}
Kostenlos starten →