Comprehensive dosimetric characterization of novel silicon carbide detectors with UHDR electron beams for FLASH radiotherapy

Giuliana Milluzzo, Marzio De Napoli, Fabio Di Martino, Antonino Amato, Damiano Del Sarto, Maria Cristina D'Oca, Maurizio Marrale, Luigi Masturzo, Elisabetta Medina, Chinonso Okpuwe, Jake Harold Pensavalle, Anna Vignati und 2 weitere

Medical Physics · 2024

DOI: 10.1002/mp.17172 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Giuliana Milluzzo und 13 weiteren Autor:innen. Beteiligte Einrichtung: National Institute of Nuclear Physics (INFN), Catania Division, Catania, Italy. Hauptthemen laut MEDLINE: Carbon Compounds, Inorganic, Silicon Compounds, Radiometry, Electrons. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: MUR-PNRR project SAMOTHRACE, Fondazione Pisa, PNRR-PNC project ANTHEM.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The extremely fast delivery of doses with ultra high dose rate (UHDR) beams necessitates the investigation of novel approaches for real‐time dosimetry and beam monitoring. This aspect is fundamental in the perspective of the clinical application of FLASH radiotherapy (FLASH‐RT), as conventional dosimeters tend to saturate at such extreme dose rates.

Purpose

This study aims to experimentally characterize newly developed silicon carbide (SiC) detectors of various active volumes at UHDRs and systematically assesses their response to establish their suitability for dosimetry in FLASH‐RT.

Methods

SiC PiN junction detectors, recently realized and provided by STLab company, with different active areas (ranging from 4.5 to 10 mm 2 ) and thicknesses (10–20 µm), were irradiated using 9 MeV UHDR pulsed electron beams accelerated by the ElectronFLASH linac at the Centro Pisano for FLASH Radiotherapy (CPFR). The linearity of the SiC response as a function of the delivered dose per pulse (DPP), which in turn corresponds to a specific instantaneous dose rate, was studied under various experimental conditions by measuring the produced charge within the SiC active layer with an electrometer. Due to the extremely high peak currents, an external customized electronic RC circuit was built and used in conjunction with the electrometer to avoid saturation.

Results

The study revealed a linear response for the different SiC detectors employed up to 21 Gy/pulse for SiC detectors with 4.5 mm 2 /10 µm active area and thickness. These values correspond to a maximum instantaneous dose rate of 5.5 MGy/s and are indicative of the maximum achievable monitored DPP and instantaneous dose rate of the linac used during the measurements.

Conclusions

The results clearly demonstrate that the developed devices exhibit a dose‐rate independent response even under extreme instantaneous dose rates and dose per pulse values. A systematic study of the SiC response was also performed as a function of the applied voltage bias, demonstrating the reliability of these dosimeters with UHDR also without any applied voltage. This demonstrates the great potential of SiC detectors for accurate dosimetry in the context of FLASH‐RT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Carbon Compounds, InorganicSilicon CompoundsRadiometryElectronsRadiotherapy DosageRadiotherapy, High-Energy

Stichwörter der Autor:innen

FLASH radiotherapyUHDRdosimetrysilicon carbide

Substanzen

silicon carbideCarbon Compounds, InorganicSilicon Compounds

Autor:innen und Institutionen

  • National Institute of Nuclear Physics (INFN), Catania Division, Catania, Italy.
  • Centro Pisano ricerca e implementazione clinica Flash Radiotherapy (CPFR@CISUP), Pisa, Italy.
  • Fisica Sanitaria, Azienda Ospedaliero Universitaria Pisa AOUP, Pisa, Italy.
  • National Institute of Nuclear Physics (INFN), Pisa Division, Pisa, Italy.
  • STLab srl, Catania, Italy.
  • National Institute of Nuclear Physics (INFN), Laboratori Nazionali del Sud, Catania, Italy.

Förderung

  • MUR-PNRR project SAMOTHRACE
  • Fondazione Pisa
  • PNRR-PNC project ANTHEM
  • CSN5 of the INFN with the FRIDA national project

Zitieren

APA

Milluzzo, G., De Napoli, M., Di Martino, F., Amato, A., Del Sarto, D., D'Oca, M. C., Marrale, M., Masturzo, L., Medina, E., Okpuwe, C., Pensavalle, J. H., Vignati, A., Camarda, M., & Romano, F. (2024). Comprehensive dosimetric characterization of novel silicon carbide detectors with UHDR electron beams for FLASH radiotherapy. Medical Physics. https://doi.org/10.1002/mp.17172

Vancouver

Milluzzo G, De Napoli M, Di Martino F, Amato A, Del Sarto D, D'Oca MC, et al. Comprehensive dosimetric characterization of novel silicon carbide detectors with UHDR electron beams for FLASH radiotherapy. Medical Physics. 2024. doi:10.1002/mp.17172

BibTeX
@article{Milluzzo2024,
  author = {Milluzzo, Giuliana and De Napoli, Marzio and Di Martino, Fabio and Amato, Antonino and Del Sarto, Damiano and D'Oca, Maria Cristina and Marrale, Maurizio and Masturzo, Luigi and Medina, Elisabetta and Okpuwe, Chinonso and Pensavalle, Jake Harold and Vignati, Anna and Camarda, Massimo and Romano, Francesco},
  title = {Comprehensive dosimetric characterization of novel silicon carbide detectors with UHDR electron beams for FLASH radiotherapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17172}
}
Kostenlos starten →