Characterization of a segmented printed circuit board (PCB) as a standard for absorbed dose to water from alpha‐emitting radionuclides

Ahtesham Ullah Khan, Jeff Radtke, Larry DeWerd

Medical Physics · 2024

DOI: 10.1002/mp.16940 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Ahtesham Ullah Khan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Water, Monte Carlo Method, Radiometry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Our previous work introduced and evaluated a standard for surface absorbed dose rate per unit radioactivity to water from unsealed alpha‐emitting radionuclides used in targeted radionuclide therapy (TRT). An overall uncertainty over 4.0% at k = 1 was reported for the absorbed dose to air measurements, which was partially attributed to the rotational alignment uncertainty in the geometrical setup.

Purpose

A printed circuit board (PCB) with a segmented guard was constructed to align the extrapolation chamber (EC) and the source plates using a differential capacitance technique. The PCB EC aimed to enhance the repeatability of the ionization current measurements. The PCB EC was evaluated using a thin film 210 Po source. The measured absorbed dose to air cavity was compared with the Monte Carlo (MC) calculations. Using the extrapolation method, the surface absorbed dose rate to water was calculated.

Methods

The PCB EC was constructed with a 4.50 mm diameter collector surrounded by four sectors and a guard electrode. The sectors were isolated for rotational alignment and later connected to the guard for ionization current measurements. A bridge circuit measured differential capacitance between opposing sectors, and a hexapod motion stage rotated the source substrate to minimize the differential capacitance. The EC was evaluated using a 210 Po source with a 3.20 mm diameter and 1.253 Ci radioactivity. MC simulations were performed to calculate the , , and correction factors. Ionization current measurements were performed for air gaps in the 0.3–0.525 mm range and surface absorbed dose rate to water was calculated.

Results

Rotational offsets of up to 3.0° were found and the current repeatability was found to increase with the absorbed dose to air uncertainty calculated to be ∼2.0%. Using the capacitance method, the effective EC diameter was measured to be 4.53 mm. The recombination, polarity, and electrometer corrections were reported to be within 1.00% across all measurement trials. The MC‐calculated correction factors were calculated to be much larger than the recombination and polarity correction factors. The average , , and corrections were calculated to be 1.063, 0.9402, and 2.136, respectively. The MC‐calculated absorbed dose to air was found to overestimate the absorbed dose by over 4.00% when compared with the measured absorbed dose to air. The surface absorbed dose rate to water was calculated to be Gy/s/Bq with an overall uncertainty of 4.07%.

Conclusions

The constructed PCB EC was deemed suitable as an absorbed dose standard. A repeatable rotational alignment was achieved using the differential capacitance technique. The metal electrodes on the PCB made a difference of < 1.00% on the backscatter correction when compared to the EC comprised of polystyrene‐equivalent collector. A 20% difference in the surface absorbed dose rate to water was found between the two ECs, which is attributed to the cavity diameter differences leading to different magnitudes of dose fall‐off along the lateral direction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

WaterMonte Carlo MethodRadiometryAlpha ParticlesRadiation DosageReference StandardsRadioisotopes

Stichwörter der Autor:innen

Monte Carlo validationTRTabsorbed dosealpha‐emitting radionuclidesdosimetryextrapolation chamberprinted circuit boardstandard

Substanzen

WaterRadioisotopes

Autor:innen und Institutionen

  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.

Zitieren

APA

Khan, A. U., Radtke, J., & DeWerd, L. (2024). Characterization of a segmented printed circuit board (PCB) as a standard for absorbed dose to water from alpha‐emitting radionuclides. Medical Physics. https://doi.org/10.1002/mp.16940

Vancouver

Khan AU, Radtke J, DeWerd L. Characterization of a segmented printed circuit board (PCB) as a standard for absorbed dose to water from alpha‐emitting radionuclides. Medical Physics. 2024. doi:10.1002/mp.16940

BibTeX
@article{Khan2024,
  author = {Khan, Ahtesham Ullah and Radtke, Jeff and DeWerd, Larry},
  title = {Characterization of a segmented printed circuit board (PCB) as a standard for absorbed dose to water from alpha‐emitting radionuclides},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16940}
}
Kostenlos starten →